A new device to noninvasively estimate the intraocular pressure produced during ocular compression

被引:4
|
作者
Korenfeld, Michael S. [1 ,2 ]
Dueker, David K. [3 ]
机构
[1] Comprehens Eye Care Ltd, 901 East 3rd St, Washington, MO 63090 USA
[2] Washington Univ, Dept Ophthalmol & Visual Sci, Washington, MO USA
[3] Hamad Med Corp, Doha, Qatar
来源
CLINICAL OPHTHALMOLOGY | 2016年 / 10卷
关键词
glaucoma; intraocular pressure; tonometer; ocular compression;
D O I
10.2147/OPTH.S92954
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To describe a noninvasive instrument that estimates intraocular pressure during episodes of external globe compression and to demonstrate the accuracy and reliability of this device by comparing it to the intraocular pressures simultaneously and manometrically measured in cannulated eyes. Methods: A thin fluid-filled bladder was constructed from flexible and inelastic plastic sheeting and was connected to a pressure transducer with high pressure tubing. The output of the pressure transducer was sent to an amplifier and recorded. This device was validated by measuring induced pressure in the fluid-filled bladder while digital pressure was applied to one surface, and the other surface was placed directly against a human cadaver eye or in vivo pig eye. The human cadaver and in vivo pig eyes were each cannulated to provide a manometric intraocular pressure control. Results: The measurements obtained with the newly described device were within similar to 5% of simultaneously measured manometric intraocular pressures in both a human cadaver and in vivo pig eye model for a pressure range of similar to 15-100 mmHg. Conclusion: This novel noninvasive device is useful for estimating the intraocular pressure transients induced during any form of external globe compression; this is a clinical setting where no other devices can be used to estimate intraocular pressure.
引用
收藏
页码:71 / 77
页数:7
相关论文
共 50 条
  • [11] Intraocular pressure change from ocular compression: a study of aqueous outflow facility
    Lam, Andrew K. C.
    Xu, Fang Yu
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (07)
  • [12] Increased Intraocular Pressure During Hemodialysis: Ocular Dialysis Disequilibrium
    Maja, Ayodele K.
    Lewis, Chad Y.
    Steffen, Eric
    Zegans, Michael E.
    Graber, Martha L.
    KIDNEY MEDICINE, 2022, 4 (09)
  • [13] Intraocular pressure and ocular pulse amplitude variations during the Valsalva maneuver
    Aykan, Umit
    Erdurmus, Mesut
    Yilmaz, Burhan
    Bilge, Ahmet Hamdi
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (08) : 1183 - 1186
  • [14] Intraocular pressure and ocular pulse amplitude variations during the Valsalva maneuver
    Umit Aykan
    Mesut Erdurmus
    Burhan Yilmaz
    Ahmet Hamdi Bilge
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2010, 248 : 1183 - 1186
  • [15] The impact of ocular hemodynamics and intracranial pressure on intraocular pressure during acute gravitational changes
    Nelson, Emily S.
    Mulugeta, Lealem
    Feola, Andrew
    Raykin, Julia
    Myers, Jerry G.
    Samuels, Brian C.
    Ethier, C. Ross
    JOURNAL OF APPLIED PHYSIOLOGY, 2017, 123 (02) : 352 - 363
  • [16] Ocular blood flow in the rat during acute elevation of intraocular pressure
    Wang, L
    Fortune, B
    Cull, G
    Bui, B
    Cioffi, GA
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [17] Reduction of intraocular pressure and ocular pulse amplitude during general anesthesia
    Zuche, H. H.
    Morinello, E.
    Viestenz, A.
    Fiorentzis, M.
    Seitz, B.
    Volk, T.
    Viestenz, A.
    OPHTHALMOLOGE, 2015, 112 (09): : 764 - 769
  • [18] Development and validation of a new intraocular pressure estimate for patients with soft corneas
    Chen, Kai-Jung
    Eliasy, Ashkan
    Vinciguerra, Riccardo
    Abass, Ahmed
    Lopes, Bernardo T.
    Vinciguerra, Paolo
    Ambrosio, Renato, Jr.
    Roberts, Cynthia J.
    Elsheikh, Ahmed
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (09): : 1316 - 1323
  • [19] THE TONOMETRIC SENSOR, A NEW DEVICE FOR THE MEASUREMENT OF INTRAOCULAR-PRESSURE
    VOORTHUYZEN, JA
    DENBESTEN, C
    BERGVELD, P
    SENSORS AND ACTUATORS, 1989, 17 (3-4): : 547 - 552
  • [20] OCULAR VOLUME AND PRESSURE CHANGE DURING COMMON CAROTID COMPRESSION IN HUMAN
    BEST, M
    GALIN, MA
    INVESTIGATIVE OPHTHALMOLOGY, 1969, 8 (04): : 459 - &