摘要
Background. A thoracoscopic technique was developed for the placement of commercially available implantable cardioverter-defibrillator (ICD) patch leads in sheep. Methods and Results. Small ICD patch leads (13.5 cm2, A-67) were placed thoracoscopically in sheep (n=5) that had survived coronary artery ligation from a previous experiment. The technique used three small incisions in the left chest. After lysis of adhesions, the ICD patch lead was introduced through a mediastinoscope. The ICD patch lead was secured in the extrapericardial position with surgical clips placed in the four corners of the ICD patch lead. After 2 weeks, a median sternotomy was performed, and ICD patch leads were reexamined for positioning. Extensive fibrosis was noted to adhere the ICD patch lead to the pericardium. The surgical clips were found intact in all animals without noticeable migration of patch lead position. There was no mortality related to ICD patch lead placement, and estimated blood loss was less than 30 mL without use of cautery. Conclusions. Commercially available ICD patch leads may be reliably and safely placed with minimal patch migration in sheep using thoracoscopic techniques.
引用
收藏
页码:447 / 451
页数:5
相关论文
共 50 条
  • [21] Occurrence and extraction of implantable cardioverter-defibrillator leads with conductor externalization
    Maciag, Aleksander
    Syska, Pawel
    Sterlinski, Maciej
    Kolodzinska, Agnieszka
    Oreziak, Artur
    Kusmierski, Krzysztof
    Marciniak-Emmons, Marta
    Przybylski, Andrzej
    Szumowski, Lukasz
    Opolski, Grzegorz
    Szwed, Hanna
    CARDIOLOGY JOURNAL, 2018, 25 (02) : 221 - 228
  • [22] Failure of implantable cardioverter-defibrillator leads: A matter of lead size?
    Rordorf, Roberto
    Poggio, Luca
    Savastano, Simone
    Vicentini, Alessandro
    Petracci, Barbara
    Chieffo, Enrico
    Klersy, Catherine
    Landolina, Maurizio
    HEART RHYTHM, 2013, 10 (02) : 184 - 190
  • [23] Psychological Effects of Implantable Cardioverter-Defibrillator Leads under Advisory
    Heatherly, Steven J.
    Simmons, Tony
    Fitzgerald, David M.
    Mitchell, Mark
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2011, 34 (06): : 694 - 699
  • [24] TRIBOLOGY OF INSULATION MATERIALS WITHIN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR LEADS
    Wilson, Charles
    Himes, Adam
    PROCEEDINGS OF THE ASME/STLE INTERNATIONAL JOINT TRIBOLOGY CONFERENCE, IJTC 2012, 2013, : 33 - 34
  • [25] Comparison of the performance of implantable cardioverter-defibrillator leads among manufacturers
    Yusuke Kondo
    Masahiro Nakano
    Takatsugu Kajiyama
    Miyo Nakano
    Tomohiko Hayashi
    Ryo Ito
    Haruhiro Takahira
    Yoshio Kobayashi
    Journal of Interventional Cardiac Electrophysiology, 2020, 58 : 133 - 139
  • [26] Comparison of the performance of implantable cardioverter-defibrillator leads among manufacturers
    Kondo, Yusuke
    Nakano, Masahiro
    Kajiyama, Takatsugu
    Nakano, Miyo
    Hayashi, Tomohiko
    Ito, Ryo
    Takahira, Haruhiro
    Kobayashi, Yoshio
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2020, 58 (02) : 133 - 139
  • [27] Evaluation of Chest Pain after Implantable Cardioverter-Defibrillator Placement
    Gahremanpour, Amir
    Saeed, Mohammad
    Birnbaum, Yochai
    TEXAS HEART INSTITUTE JOURNAL, 2017, 44 (03) : 226 - 226
  • [28] Subglandular Placement of an Implantable Cardioverter-Defibrillator for an Improved Cosmetic Outcome
    Wright, Cassidy D.
    Roehl, Kendall R.
    Huang, Shoei K. Stephen
    Mahabir, Raman C.
    ANNALS OF PLASTIC SURGERY, 2013, 71 (05) : 621 - 623
  • [29] Transvenous extraction of implantable cardioverter-defibrillator leads under advisory-A comparison of Riata, Sprint Fidelis, and non-recalled implantable cardioverter-defibrillator leads
    Brunner, Michael P.
    Cronin, Edmond M.
    Jacob, Jessen
    Duarte, Valeria E.
    Tarakji, Khaldoun G.
    Martin, David O.
    Callahan, Thomas
    Borek, P. Peter
    Cantillon, Daniel J.
    Niebauer, Mark J.
    Saliba, Walid I.
    Kanj, Mohamed
    Wazni, Oussama
    Baranowski, Bryan
    Wilkoff, Bruce L.
    HEART RHYTHM, 2013, 10 (10) : 1444 - 1450
  • [30] Benefits of the implantable cardioverter-defibrillator
    Epstein, Andrew E.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 52 (14) : 1122 - 1127