INITIAL CLINICAL-EXPERIENCE WITH THE AHMED GLAUCOMA VALVE IMPLANT

被引:277
|
作者
COLEMAN, AL
HILL, R
WILSON, MR
CHOPLIN, N
KOTASNEUMANN, R
TAM, M
BACHARACH, J
PANEK, WC
机构
[1] UNIV CALIF IRVINE,SCH MED,DEPT OPHTHALMOL,IRVINE,CA 92717
[2] USN,MED CTR,SAN DIEGO,CA 92152
[3] CHAIM SHEBA MED CTR,GOLDSCHLEGER EYE INST,IL-52621 TEL HASHOMER,ISRAEL
[4] CALIF PACIFIC MED CTR,SAN FRANCISCO,CA
[5] VET ADM OUTPATIENT CLIN,SACRAMENTO,CA
关键词
D O I
10.1016/S0002-9394(14)73755-9
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: We studied the Ahmed Glaucoma Valve implant, an aqueous shunting device that has a unidirectional valve mechanism designed to prevent postoperative hypotony in eyes with in tractable glaucoma. METHODS: In this multicenter, prospective clinical trial, we studied 60 eyes (60 patients) with increased intraocular pressure or glaucoma that had not responded to medical treatment, laser photocoagulation, or previous glaucoma surgery, in which the Ahmed Glaucoma Valve implant was placed to decrease intraocular pressure. RESULTS: Success was defined as intraocular pressure less than 22 mm Hg and greater than 4 mm Hg for two months or longer, intraocular pressure that was lowered by at least 20% from preoperative values (in eyes with preoperative intraocular pressures less than 22 mm Hg), and no additional glaucoma surgery or visually devastating complications. Cumulative probability of success at 12 months was 78%. Eight (13%) of 60 eyes had intraocular pressure less than 5 mm Hg the first postoperative day, Two other eyes had shallow anterior chambers, which required anterior chamber reformation. The major complications associated with the use of the valve were serous choroidal detachments in 13 eyes (22%), blockage of the tube in six eyes (10%), malposition of the tube in four eyes (7%), a suprachoroidal hemorrhage in one eye (2%), and corneal graft rejections in three (19%) of 16 eyes with corneal grafts. CONCLUSIONS: Although the 12-month success with the Ahmed Glaucoma Valve implant is similar to that reported for other drainage devices, the complications associated with overfiltration in the immediate postoperative period appear to be less frequent than with other valved drainage devices, Randomized, prospective studies to compare the Ahmed Glaucoma Valve implant with other drainage devices are needed to make clinical comparisons of the different devices.
引用
收藏
页码:23 / 31
页数:9
相关论文
共 50 条
  • [1] INITIAL CLINICAL-EXPERIENCE WITH THE AHMED GLAUCOMA VALVE IMPLANT-CORRECTION
    COLEMAN, AL
    WILSON, MR
    TAM, M
    HILL, R
    CHOPLIN, N
    KOTASNEUMANN, R
    BACHARACH, J
    PANEK, WC
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1995, 120 (05) : 684 - 684
  • [2] Initial clinical experience with the Ahmed glaucoma valve implant in the management of neovascular glaucoma
    Patel, S
    Ge, N
    Varma, R
    Chen, P
    Heuer, D
    Lee, P
    Dacey, M
    Philips, A
    Minckler, D
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 1173 - 1173
  • [3] Initial clinical experience with the Ahmed Glaucoma Valve implant in pediatric patients
    Coleman, AL
    Smyth, RJ
    Wilson, MR
    Tam, M
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1997, 115 (02) : 186 - 191
  • [4] Initial clinical experience with the Ahmed glaucoma valve implant in the management of complicated glaucoma.
    Ge, N
    Patel, S
    Heuer, D
    Minckler, D
    Varma, R
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 1178 - 1178
  • [5] Intermediate-term clinical experience with the Ahmed Glaucoma Valve implant
    Huang, MC
    Netland, PA
    Coleman, AL
    Siegner, SW
    Moster, MR
    Hill, RA
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1999, 127 (01) : 27 - 33
  • [6] INITIAL CLINICAL EXPERIENCE WITH AHMED VALVE IMPLANTATION IN REFRACTORY PEDIATRIC GLAUCOMA
    Novak-Laus, Katia
    Herman, Jelena Skunca
    Prskalo, Marija Simic
    Jurisic, Darija
    Mondic, Zdravko
    [J]. ACTA CLINICA CROATICA, 2016, 55 (04) : 555 - 559
  • [7] Ahmed glaucoma valve implant: Experience in East Africa
    Kiage, D. O.
    Gradin, D.
    Gichuhi, S.
    Damji, K. F.
    [J]. MIDDLE EAST AFRICAN JOURNAL OF OPHTHALMOLOGY, 2009, 16 (03) : 151 - 155
  • [8] A clinical study of the Ahmed glaucoma valve implant in advanced glaucoma
    Ayyala, RS
    Zurakowski, D
    Smith, JA
    Monshizadeh, R
    Netland, PA
    Richards, DW
    Layden, WE
    [J]. OPHTHALMOLOGY, 1998, 105 (10) : 1968 - 1976
  • [9] Clinical experience with the Ahmed glaucoma valve implant in eyes with prior or concurrent penetrating keratoplasties
    Coleman, AL
    Mondino, BJ
    Wilson, MR
    Casey, R
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1997, 123 (01) : 54 - 61
  • [10] Initial experience of the Ahmed valved implant in the management of refractory glaucoma
    Aziz, S.
    McConnachie, A.
    Montgomery, D. M. I.
    [J]. EYE, 2007, 21 (02) : 278 - 279