24 YEARS IMPLANT DURATION OF THE AORTIC STARR-EDWARDS SILASTIC BALL PROSTHESIS - A VALVE OF THE PAST

被引:5
|
作者
MACHLER, HE
SCHMIDT, CH
NEUNER, P
IBERER, F
ANELLIMONTI, M
DACAR, D
RIGLER, B
KRAFTKINZ, J
机构
[1] Department of Surgery, University of Graz
[2] Institute of Pathology, University of Graz
关键词
AORTIC STARR-EDWARDS BALL PROSTHESIS; LONGEVITY; RELIABILITY;
D O I
10.1016/1010-7940(93)90031-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Six years after the first aortic valve replacement with the Starr-Edwards Silastic ball prosthesis at Oregon Health Sciences University, the model 1260 was implanted in a 44-year-old patient at our department. During 24 years no signs of dysfunction, thromboembolism, thrombosis, periprosthetic leaks or hemolysis were observed. Material test showed no signs of fatigue. The dimension of the Silastic poppet was found to be one millimeter less than the original specifications of these poppets and there were some lipid infiltrations. Functionally the poppet was found to be nonvariant. Histologic findings detected focal hyalinization and giant cells without signs of acute inflammation or ulceration. With regard to its reliability and durability, the Starr-Edwards valve prosthesis should not be viewed with disfavor.
引用
收藏
页码:114 / 116
页数:3
相关论文
共 50 条
  • [1] Replacement of an aortic Starr-Edwards ball valve prosthesis 28 years after implantation
    Tayama, K
    Akashi, H
    Tayama, E
    Fukunaga, S
    Kawara, T
    Kosuga, K
    Aoyagi, S
    [J]. CIRCULATION JOURNAL, 2004, 68 (05) : 507 - 508
  • [2] Fatal embolization of ball portion of Starr-Edwards aortic valve prosthesis
    Reddy, KK
    Anders, KH
    Sathyavagiswaran, L
    [J]. JOURNAL OF FORENSIC SCIENCES, 1998, 43 (01) : 225 - 227
  • [3] BEYOND 24 YEARS WITH THE AORTIC STARR-EDWARD SILASTIC BALL VALVE - A VALVE OF THE PAST - NO
    STARR, A
    GRUNKEMEIER, GL
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1993, 7 (03) : 113 - 114
  • [4] BALL FRACTURE AND EXTRUSION IN STARR-EDWARDS AORTIC VALVE PROSTHESIS WITH DISSEMINATION OF BALL MATERIAL
    HAMEED, K
    ASHFAQ, S
    WAUGH, DOW
    [J]. ARCHIVES OF PATHOLOGY, 1968, 86 (05): : 520 - &
  • [5] Performance profile of the Starr-Edwards aortic cloth covered valve, track valve, and silastic ball valve
    Lund, O
    Pilegaard, HK
    Ilkjaer, LB
    Nielsen, SL
    Arildsen, H
    Albrechtsen, OK
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1999, 16 (04) : 403 - 413
  • [6] 49 years of normal functioning Starr-Edwards aortic valve prosthesis
    Sabouni, Mouhamed Amr
    Baumeister, Robin H.
    Traverse, Paul
    [J]. OXFORD MEDICAL CASE REPORTS, 2020, (02): : 39 - 42
  • [7] AORTIC VALVE REPLACEMENT WITH STARR-EDWARDS BALL-VALVE PROSTHESIS - INDICATIONS AND RESULTS
    LEWIS, RP
    HERR, RH
    STARR, A
    GRISWOLD, HE
    [J]. AMERICAN HEART JOURNAL, 1966, 71 (04) : 549 - &
  • [8] FATAL OBSTRUCTION OF AORTIC ROOT BY STARR-EDWARDS AORTIC BALL-VALVE PROSTHESIS
    NAJAFI, H
    DEWALL, RA
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1966, 51 (02): : 180 - &
  • [9] Ascending aortic aneurysm in a patient with an aortic Starr-Edwards ball valve prosthesis implanted 39 years previously
    Nishigawa K.
    Totsugawa T.
    Yoshitaka H.
    Tsushima Y.
    Kuinose M.
    Chikazawa G.
    [J]. General Thoracic and Cardiovascular Surgery, 2010, 58 (3) : 138 - 140
  • [10] HEALING AROUND STARR-EDWARDS AORTIC VALVE PROSTHESIS
    VASCONEZ, LO
    SHANKLIN, DR
    WHEAT, MW
    [J]. CIRCULATION, 1967, 36 (4S2) : I257 - &