Validation of Bulk Analysis with Simulated Swipe Samples Containing Ultra-Trace Amounts of Uranium and Plutonium Using MC-ICP-MS

被引:5
|
作者
Lim, Sang Ho [1 ,2 ]
Han, Sun-Ho [1 ]
Park, Jong-Ho [1 ,2 ]
Park, Ranhee [1 ]
Lee, Min Young [1 ]
Park, Jinkyu [1 ]
Lee, Chi-Gyu [1 ]
Song, Kyuseok [1 ]
机构
[1] Korea Atom Energy Res Inst, Nucl Chem Res Div, Daedeok Daero 989-111, Daejeon 34057, South Korea
[2] Univ Sci & Technol, Dept Radiochem & Nucl Nonproliferat, 217 Gajeong-ro, Daejeon 34113, South Korea
关键词
Bulk analysis; Environmental samples; Chemical separation; IDMS; MC-ICP-MS;
D O I
10.5478/MSL.2015.6.3.75
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Suitable analytical procedures for the bulk analysis of ultra-trace amounts of uranium and plutonium have been developed using multi-collector inductively coupled mass spectrometry (MC-ICP-MS). The quantification and determination of the isotopic ratios of uranium and plutonium in three simulated swipe samples, a swipe blank, and a process blank were performed to validate the analytical performance. The analytical results for the simulated swipe samples were in good agreement with the certified values, based on the measurement quality goals for the analysis of bulk environmental samples recommended by the International Atomic Energy Agency (IAEA)
引用
收藏
页码:75 / 79
页数:5
相关论文
共 31 条
  • [1] Optimization of uranium and plutonium separations using TEVA and UTEVA cartridges for MC-ICP-MS analysis of environmental swipe samples
    Metzger, Shalina C.
    Rogers, Kayron T.
    Bostick, Debra A.
    McBay, Eddy H.
    Ticknor, Brian W.
    Manard, Benjamin T.
    Hexel, Cole R.
    TALANTA, 2019, 198 : 257 - 262
  • [2] Improvement in the chemical separation and determination of uncertainties for bulk analysis of Pu isotopes at ultra-trace levels by using MC-ICP-MS
    Lim, Sang Ho
    Park, Jinkyu
    Han, Sun-Ho
    Park, Ranhee
    Lee, Chi-Gyu
    Park, Jong-Ho
    Song, Kyuseok
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2016, 307 (03) : 1853 - 1859
  • [3] Improvement in the chemical separation and determination of uncertainties for bulk analysis of Pu isotopes at ultra-trace levels by using MC-ICP-MS
    Sang Ho Lim
    Jinkyu Park
    Sun-Ho Han
    Ranhee Park
    Chi-Gyu Lee
    Jong-Ho Park
    Kyuseok Song
    Journal of Radioanalytical and Nuclear Chemistry, 2016, 307 : 1853 - 1859
  • [4] Determination of Plutonium and Uranium Radionuclides in Glacier Ice Samples by MC-ICP-MS
    Rollin, Stefan
    Sahli, Hans
    Gnagi, Lars
    Alvarado, Jose A. Corcho
    CHIMIA, 2020, 74 (12) : 989 - 994
  • [5] Systematic Evaluation of Uranium Impurities in Process Blanks for Ultra-Trace Analysis on Environmental Safeguards Samples using MC-ICP-MS and Isotope Dilution Mass Spectrometry
    Lim, Sang Ho
    Han, Sun-Ho
    Park, Jong-Ho
    Park, Ranhee
    Lee, Chi-Gyu
    Song, Kyuseok
    ASIAN JOURNAL OF CHEMISTRY, 2014, 26 (13) : 4070 - 4072
  • [6] High-precision measurement of Cd isotopes in ultra-trace Cd samples using double spike-standard addition MC-ICP-MS
    Chang, Hui
    Zhu, Jian-Ming
    Wang, Xiangli
    Gao, Ting
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2023, 38 (04) : 950 - 962
  • [7] Uranium isotope analysis by MC-ICP-MS in sub-ng sized samples
    Boulyga, Sergei F.
    Koepf, Andreas
    Konegger-Kappel, Stefanie
    Macsik, Zsuzsanna
    Stadelmann, Guillaume
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2016, 31 (11) : 2272 - 2284
  • [8] Improvement of Measurement Precisions for Uranium Isotopes at Ultra Trace Levels by Modification of the Sample Introduction System in MC-ICP-MS
    Park, Ranhee
    Lim, Sang Ho
    Han, Sun-Ho
    Lee, Min Young
    Park, Jinkyu
    Lee, Chi-Gyu
    Song, Kyuseok
    MASS SPECTROMETRY LETTERS, 2016, 7 (02) : 50 - 54
  • [9] High precision potassium isotope analysis of geological samples using MC-ICP-MS
    Chen Y.
    Sun H.
    Gu H.
    Huang C.
    Cao Z.
    Wang F.
    Ge C.
    Dizhi Xuebao/Acta Geologica Sinica, 2023, 97 (04): : 1360 - 1370
  • [10] A new era of plutonium assay and isotope ratio measurements at ultra-low levels in IAEA safeguards environmental swipe samples by isotope dilution MC-ICP-MS using a 244Pu spike
    Konegger-Kappel, Stefanie
    Koepf, Andreas
    Katona, Robert
    Kilburn, Matthew R.
    Boulyga, Sergei F.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2023, 332 (07) : 2887 - 2904