High-precision measurement of Cd isotopes in ultra-trace Cd samples using double spike-standard addition MC-ICP-MS

被引:0
|
作者
Chang, Hui [1 ]
Zhu, Jian-Ming [1 ,3 ]
Wang, Xiangli [2 ]
Gao, Ting [3 ]
机构
[1] China Univ Geosci Beijing, State Key Lab Geol Proc & Mineral Resources, Beijing 100083, Peoples R China
[2] Chinese Acad Sci, Inst Geol & Geophys, Key Lab Cenozo Geol & Environm, Beijing 100029, Peoples R China
[3] Chinese Acad Sci, Inst Geochem, State Key Lab Environm Geochem, Guiyang 550081, Peoples R China
基金
中国国家自然科学基金; 国家重点研发计划;
关键词
GEOLOGICAL REFERENCE MATERIALS; CADMIUM; FRACTIONATION; LIVER; SOIL; GEOCHEMISTRY; PERSPECTIVES; BIVALVES; SEAWATER; OCEAN;
D O I
10.1039/d3ja00047h
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The double spike-standard addition (DSSA) is a powerful method proposed for isotope measurement of ultra-trace elements in samples. Here, we test the method's practicality and robustness for obtaining high-precision delta Cd-114/110 data in samples with complex matrices and ultra-trace cadmium (Cd) concentrations as low as 0.004 mu g g(-1). Through a DS data reduction routine and isotope binary mixing model, the delta Cd-114/110 values obtained for numeral geological and biological reference materials (GRMs/BRMs) agree well with certified or previous measurements, even for similar to 2.1 ng Cd of BCR-2. The overall precision of both single- and multi-standard addition depends on the sample fraction (f(spl)) owing to the error propagation. And 0.041 +/- 0.022 parts per thousand (2SD, n = 46) of precision for delta Cd-114/110 can be achieved when the sample fraction (f(spl)) is >= 20%, comparable to 0.056 +/- 0.039 parts per thousand (2SD, n = 23) obtained by the traditional DS method. However, it becomes greater than 0.110 parts per thousand when f(spl) is <20%, indicating that 20% (1/5) -50% (1/2) of f(spl) is the optimal mixing range to obtain high-precision data if the minimal sample sizes are required. These further confirm that accurate and precise Cd isotope ratios can be determined by DSSA. Correspondingly, the purification scheme can be simplified to a single column due to the added standard solution boosting analyte and diluting ratios of matrix/Cd. Animal organs, with 0.004-0.106 mu g g(-1) Cd, yield large variations of delta Cd-114/110 (-0.054 +/- 0.030 parts per thousand (2SD, n = 4) and 0.681 +/- 0.022 parts per thousand (2SD, n = 4) for ovine liver and kidney, respectively), suggesting that Cd isotopes can be fractionated significantly during biological metabolic processes and may be a potential use in medical diagnosis. Robust measurement of Cd isotope composition in ultra-trace Cd samples with complex matrices by DSSA broadens the scope of measurable samples by the traditional DS method, thus potentially opening a range of new opportunities in life, agricultural, environmental, and earth/planetary sciences.
引用
收藏
页码:950 / 962
页数:13
相关论文
共 50 条
  • [1] High-precision measurement of Cd isotopes in ultra-trace Cd samples using double spike-standard addition MC-ICP-MS (vol 38, pg 950, 2023)
    Chang, Hui
    Zhu, Jian-Ming
    Wang, Xiangli
    Gao, Ting
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2024, 39 (04) : 1175 - 1175
  • [2] High-precision MC-ICP-MS measurements of Cd isotopes using a novel double spike method without Sn isobaric interference
    Li, Jin
    Tang, Suo-Han
    Zhu, Xiangkun
    Ma, Jian-Xiong
    Zhu, Zhiyong
    Yan, Bin
    [J]. Journal of Analytical Atomic Spectrometry, 2024, 39 (12) : 3106 - 3115
  • [3] Validating the high-precision measurement of Mo isotopes at the 5 ng level using double spike MC-ICP-MS
    Zhu, Hong-Gang
    Zhu, Jian-Ming
    Tan, Decan
    Lin, Xuan
    Lu, Kaite
    Yang, Wen
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2022, 37 (05) : 1063 - 1075
  • [4] High-sensitivity determination of Cd isotopes in low-Cd geological samples by double spike MC-ICP-MS
    Tan, Decan
    Zhu, Jian-Ming
    Wang, Xiangli
    Han, Guilin
    Lu, Zhuo
    Xu, Wenpo
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2020, 35 (04) : 713 - 727
  • [5] A novel purification method for high precision measurement of Ni isotopes by double spike MC-ICP-MS
    Wu, Guangliang
    Zhu, Jian-Ming
    Wang, Xiangli
    Han, Guilin
    Tan, Decan
    Wang, Shui-Jiong
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2019, 34 (08) : 1639 - 1651
  • [6] High-Precision Cd Isotope Measurements of Soil and Rock Reference Materials by MC-ICP-MS with Double Spike Correction
    Liu, Meng-Shu
    Zhang, Qun
    Zhang, Yingnan
    Zhang, Zhaofeng
    Huang, Fang
    Yu, Hui-Min
    [J]. GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2020, 44 (01) : 169 - 182
  • [7] High-precision measurement of Ag isotopes for silicate rocks by MC-ICP-MS
    Fang, Yuan
    Wen, Qiu-Yu
    Xiao, Zi-Cong
    Yu, Hui-Min
    Huang, Fang
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2024, 39 (07) : 1750 - 1758
  • [8] High-precision measurement of Ag isotopes for silicate rocks by MC-ICP-MS
    Fang, Yuan
    Wen, Qiu-Yu
    Xiao, Zi-Cong
    Yu, Hui-Min
    Huang, Fang
    [J]. Journal of Analytical Atomic Spectrometry, 1600, 39 (07): : 1750 - 1758
  • [9] An efficient Cd two-stage column system for high-precision determination of Cd isotopic compositions by double spike MC-ICP-MS
    Zhang, Zheyuan
    Li, Tao
    Li, Baichan
    Lin, Yibo
    Li, Gaojun
    Chen, Tianyu
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2024, 39 (04) : 1142 - 1151
  • [10] An improved method of Cr purification for high precision measurement of Cr isotopes by double spike MC-ICP-MS
    Zhu, Jian-Ming
    Wu, Guangliang
    Wang, Xiangli
    Han, Guilin
    Zhang, Lixin
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2018, 33 (05) : 809 - 821