DUAL TIME TO DIGITAL CONVERTER FOR DELAY-LINE READOUT OF POSITION-SENSITIVE GAS-FILLED DETECTORS

被引:8
|
作者
BERRY, A
机构
[1] European Synchrotron Radiation Facility, Grenoble
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 1993年 / 64卷 / 05期
关键词
D O I
10.1063/1.1144120
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new dual channel time digitizer has been designed. It is based on the coarse counter and interpolation method. A new technique of interpolation has been employed which gives a FWHM resolution of 50 ps and a differential nonlinearity of +/-2%, at a master oscillator frequency of 200 MHz. The module has a dynamic measurement range of 10 bits with a fixed dead time of 30 ns. It is a single width C size VXI format. Presented is a description of the new interpolation method and typical performance results.
引用
收藏
页码:1222 / 1228
页数:7
相关论文
共 50 条
  • [1] A FAST TIME-TO-DIGITAL CONVERTER FOR POSITION-SENSITIVE RADIATION DETECTORS WITH DELAY-LINE READOUTS
    HARDER, JA
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 265 (03): : 500 - 510
  • [2] A POSITION-SENSITIVE MICROCHANNELPLATE DETECTOR USING A DELAY-LINE READOUT ANODE
    KELLER, H
    KLINGELHOFER, G
    KANKELEIT, E
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 258 (02): : 221 - 224
  • [3] A FAST READOUT SYSTEM FOR 2D GAS-FILLED DETECTORS BASED ON THE DELAY-LINE METHOD
    BERRY, A
    GABRIEL, A
    KOCSIS, M
    RIEKEL, C
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 348 (2-3): : 334 - 337
  • [4] A beginners' guide to gas-filled proportional detectors with delay line readout
    Petrascu, AM
    Koch, MHJ
    Gabriel, A
    [J]. JOURNAL OF MACROMOLECULAR SCIENCE-PHYSICS, 1998, B37 (04): : 463 - 483
  • [5] LOW-DEAD-TIME DELAY-LINE READOUT FOR POSITION-SENSITIVE PROPORTIONAL-COUNTERS
    ANDERSON, AN
    BARRETT, RJ
    WADLINGER, EA
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1974, 119 (02): : 365 - 366
  • [6] DELAY-LINE READOUT OF MICROCHANNEL PLATES IN A PROTOTYPE POSITION-SENSITIVE PHOTOMULTIPLIER TUBE
    WILLIAMS, MB
    SOBOTTKA, SE
    SHEPHERD, JA
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 302 (01): : 105 - 112
  • [7] Digitized signal acquisition of gas-filled, multiwire, neutron detector with delay-line readout
    Orban, J.
    Rosta, L.
    Nagy, A.
    [J]. INTERNATIONAL WORKSHOP ON NEUTRON OPTICS AND DETECTORS (NOP&D 2013), 2014, 528
  • [8] Readout electronics for position-sensitive gas filled detectors at the European Synchrotron Radiation Facility
    Hervé, C
    Le Caër, T
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (01): : 226 - 231
  • [9] A data acquisition system for two-dimensional position sensitive micropattern gas detectors with delay-line readout
    Hanu, A. R.
    Prestwich, W. V.
    Byun, S. H.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 780 : 33 - 39
  • [10] DIRECT TIME DIGITIZING FOR DETECTORS WITH DELAY-LINE READOUT
    BOULIN, C
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 201 (01): : 237 - 239