ANTERIOR SCREW FIXATION OF TYPE-II ODONTOID FRACTURES - A BIOMECHANICAL STUDY

被引:51
|
作者
MCBRIDE, AD
MUKHERJEE, DP
KRUSE, RN
ALBRIGHT, JA
机构
[1] Department of Orthopaedic Surgery, Louisiana State University Medical Center, Shreveport, LA
关键词
BIOMECHANICS; INTERNAL SCREW FIXATION; ODONTOID FRACTURE;
D O I
10.1097/00007632-199509000-00001
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. This study examined the torsional and shear stiffness of Type II odontoid fractures that were stabilized with two 3.5-mm cannulated AO screws or a single 4.5-mm cannulated Herbert screw. Objective. To determine the stability of a commonly used two-screw method versus a single-screw fixation method. Insertion of a single screw could make the procedure technically easier. Summary of Background Data. Biomechanical studies have shown the 4.5-mm Herbert screw to generate greater compressive forces than 3.5-mm cannulated screws, which are commonly used to anteriorly stabilize Type II odontoid fractures. No previous biomechanical studies have compared the shear or torsional stiffness of Type II odontoid fractures stabilized with these screws. Methods. Twelve cadaver C2 vertebrae were harvested and an osteotomy was performed to simulate a Type II fracture pattern. Six were stabilized with two 3.5-mm cannulated screws, and the remaining were stabilized with a single 4.5-mm cannulated Herbert screw. Each specimen then was tested in torsion through +/-0.75 degrees, +/-1.25 degrees, and +/-1.75 degrees for 10 cycles each. Shear forces then were applied from an anterior to posterior direction to a maximum load of 44.48 N. Results. The mean torsional stiffness for the Herbert screw was 1196 N . m/deg, which was significantly greater than the mean stiffness of the 3.5-mm screw fixation, which measured 434 N . m/deg. The mean shear stiffness for the Herbert screw fixation was 106.9 kN/m, compared to 86.1 kN/m for the 3.5-mm cannulated screw. This was not statistically significant. Conclusions. The biomechanical properties of the 4.5-mm cannulated Herbert screw suggest it may lend ;itself for use in fixation of Type II odontoid fractures. With only a single screw to insert, this technically demanding procedure may be less daunting.
引用
收藏
页码:1855 / 1859
页数:5
相关论文
共 50 条
  • [31] Treatment of odontoid fractures with single anterior screw fixation
    Song, Kyung-Jin
    Lee, Kwang-Bok
    Kim, Ki-Nam
    JOURNAL OF CLINICAL NEUROSCIENCE, 2007, 14 (09) : 824 - 830
  • [32] Surgical treatment of acute type II and rostral type III odontoid fractures managed by anterior screw fixation
    Fountas, KN
    Machinis, TG
    Kapsalaki, EZ
    Dimopoulos, VG
    Feltes, CH
    SOUTHERN MEDICAL JOURNAL, 2005, 98 (09) : 896 - 901
  • [33] Comparison of Percutaneous and Open Anterior Screw Fixation in the Treatment of Type II and Rostral Type III Odontoid Fractures
    Wang, Jian
    Zhou, Yue
    Zhang, Zheng Feng
    Li, Chang Qing
    Zheng, Wen Jie
    Liu, Jie
    SPINE, 2011, 36 (18) : 1459 - 1463
  • [34] ONE HERBERT DOUBLE-THREADED COMPRESSION SCREW FIXATION OF DISPLACED TYPE-II ODONTOID FRACTURES
    CHANG, KW
    LIU, YW
    CHENG, PG
    CHANG, LB
    SUEN, KL
    CHUNG, WL
    CHEN, UL
    LIANG, PL
    JOURNAL OF SPINAL DISORDERS, 1994, 7 (01): : 62 - 69
  • [35] TREATMENT OF REMOTE TYPE II AXIS FRACTURES IN THE ELDERLY: FEASIBILITY OF ANTERIOR ODONTOID SCREW FIXATION COMMENTS
    Wang, Michael Y.
    Apfelbaum, Ronald I.
    Resnick, Daniel K.
    Sonntag, Volker K. H.
    Traynelis, Vincent C.
    NEUROSURGERY, 2008, 63 (06) : 1150 - 1151
  • [36] Anterior screw fixation in type II odontoid fractures: Is there a difference in outcome between age groups? Comments
    Sonntag, VKH
    Benzel, EC
    Haid, RW
    McCormick, PC
    NEUROSURGERY, 2003, 52 (05) : 1092 - 1094
  • [37] Cement-augmented anterior screw fixation of Type II odontoid fractures in elderly patients with osteoporosis
    Kohlhof, Hendrik
    Seidel, Ulrich
    Hoppe, Sven
    Keel, Marius J.
    Benneker, Lorin M.
    SPINE JOURNAL, 2013, 13 (12): : 1858 - 1863
  • [38] Anterior screw fixation for type II odontoid process fractures: A single-center experience with the double Herbert screw fixation technique (Retrospective cohort study)
    Pongmanee, Suthipas
    Kaensuk, Sitthikorn
    Sarasombath, Peem
    Rojdumrongrattana, Borvornsake
    Kritworakarn, Noparoot
    Liawrungrueang, Wongthawat
    ANNALS OF MEDICINE AND SURGERY, 2022, 74
  • [39] Placement of a single anterior odontoid screw for type II odontoid fractures in a setting with limited resources
    Ahmed Hamdy Ashry
    Hashem Mohamed Aboul-Ela
    The Egyptian Journal of Neurology, Psychiatry and Neurosurgery, 55
  • [40] Placement of a single anterior odontoid screw for type II odontoid fractures in a setting with limited resources
    Ashry, Ahmed Hamdy
    Aboul-Ela, Hashem Mohamed
    EGYPTIAN JOURNAL OF NEUROLOGY PSYCHIATRY AND NEUROSURGERY, 2019, 55 (01):