ANTERIOR SCREW FIXATION OF TYPE-II ODONTOID FRACTURES - A BIOMECHANICAL STUDY

被引:51
|
作者
MCBRIDE, AD
MUKHERJEE, DP
KRUSE, RN
ALBRIGHT, JA
机构
[1] Department of Orthopaedic Surgery, Louisiana State University Medical Center, Shreveport, LA
关键词
BIOMECHANICS; INTERNAL SCREW FIXATION; ODONTOID FRACTURE;
D O I
10.1097/00007632-199509000-00001
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. This study examined the torsional and shear stiffness of Type II odontoid fractures that were stabilized with two 3.5-mm cannulated AO screws or a single 4.5-mm cannulated Herbert screw. Objective. To determine the stability of a commonly used two-screw method versus a single-screw fixation method. Insertion of a single screw could make the procedure technically easier. Summary of Background Data. Biomechanical studies have shown the 4.5-mm Herbert screw to generate greater compressive forces than 3.5-mm cannulated screws, which are commonly used to anteriorly stabilize Type II odontoid fractures. No previous biomechanical studies have compared the shear or torsional stiffness of Type II odontoid fractures stabilized with these screws. Methods. Twelve cadaver C2 vertebrae were harvested and an osteotomy was performed to simulate a Type II fracture pattern. Six were stabilized with two 3.5-mm cannulated screws, and the remaining were stabilized with a single 4.5-mm cannulated Herbert screw. Each specimen then was tested in torsion through +/-0.75 degrees, +/-1.25 degrees, and +/-1.75 degrees for 10 cycles each. Shear forces then were applied from an anterior to posterior direction to a maximum load of 44.48 N. Results. The mean torsional stiffness for the Herbert screw was 1196 N . m/deg, which was significantly greater than the mean stiffness of the 3.5-mm screw fixation, which measured 434 N . m/deg. The mean shear stiffness for the Herbert screw fixation was 106.9 kN/m, compared to 86.1 kN/m for the 3.5-mm cannulated screw. This was not statistically significant. Conclusions. The biomechanical properties of the 4.5-mm cannulated Herbert screw suggest it may lend ;itself for use in fixation of Type II odontoid fractures. With only a single screw to insert, this technically demanding procedure may be less daunting.
引用
收藏
页码:1855 / 1859
页数:5
相关论文
共 50 条
  • [21] Management of odontoid fractures with percutaneous anterior odontoid screw fixation
    Yong-Long Chi
    Xiang-Yang Wang
    Hua-Zi Xu
    Yan Lin
    Qi-Shan Huang
    Fang-Min Mao
    Wen-Fei Ni
    Sheng Wang
    Li-Yang Dai
    European Spine Journal, 2007, 16 : 1157 - 1164
  • [22] Management of odontoid fractures with percutaneous anterior odontoid screw fixation
    Chi, Yong-Long
    Wang, Xiang-Yang
    Xu, Hua-Zi
    Lin, Yan
    Huang, Qi-Shan
    Mao, Fang-Min
    Ni, Wen-Fei
    Wang, Sheng
    Dai, Li-Yang
    EUROPEAN SPINE JOURNAL, 2007, 16 (08) : 1157 - 1164
  • [23] TREATMENT OF REMOTE TYPE II AXIS FRACTURES IN THE ELDERLY: FEASIBILITY OF ANTERIOR ODONTOID SCREW FIXATION
    Agrillo, Antonino
    Russo, Natale
    Marotta, Nicola
    Delfini, Roberto
    NEUROSURGERY, 2008, 63 (06) : 1145 - 1150
  • [24] Anterior screw fixation in type II odontoid fractures:: Is there a difference in outcome between age groups?
    Börm, W
    Kast, E
    Richter, HP
    Mohr, K
    NEUROSURGERY, 2003, 52 (05) : 1089 - 1092
  • [25] Analysis of failure following anterior screw fixation of Type II odontoid fractures in geriatric patients
    Osti, Michael
    Philipp, Helmut
    Meusburger, Berthold
    Benedetto, Karl Peter
    EUROPEAN SPINE JOURNAL, 2011, 20 (11) : 1915 - 1920
  • [26] Analysis of failure following anterior screw fixation of Type II odontoid fractures in geriatric patients
    Michael Osti
    Helmut Philipp
    Berthold Meusburger
    Karl Peter Benedetto
    European Spine Journal, 2011, 20 : 1915 - 1920
  • [27] Preliminary Biomechanical Proof of Concept for a Hybrid Locking Plate/Variable Pitch Screw Construct for Anterior Fixation of Type II Odontoid Fractures
    Daniels, Alan H.
    Magee, William
    Badra, Mohammed
    Bay, Brian
    Hettwer, Werner
    Hart, Robert A.
    SPINE, 2012, 37 (19) : E1159 - E1164
  • [28] Acute odontoid fractures stabilization with anterior screw fixation
    Fountas, KN
    Kapsalaki, EZ
    Johnston, KW
    Smisson, HF
    Robinson, JS
    XXXI WORLD CONGRESS OF THE INTERNATIONAL COLLEGE OF SURGEONS, 1998, : 249 - 252
  • [29] A COMPARATIVE-STUDY OF FIXATION TECHNIQUES FOR TYPE-II FRACTURES OF THE ODONTOID PROCESS
    GRAZIANO, G
    JAGGERS, C
    LEE, M
    LYNCH, W
    SPINE, 1993, 18 (16) : 2383 - 2387
  • [30] FRACTURES OF THE ODONTOID PROCESS - TREATMENT WITH ANTERIOR SCREW FIXATION
    AEBI, M
    ETTER, C
    COSCIA, M
    SPINE, 1989, 14 (10) : 1065 - 1070