ANTERIOR SCREW FIXATION OF TYPE-II ODONTOID FRACTURES - A BIOMECHANICAL STUDY

被引:51
|
作者
MCBRIDE, AD
MUKHERJEE, DP
KRUSE, RN
ALBRIGHT, JA
机构
[1] Department of Orthopaedic Surgery, Louisiana State University Medical Center, Shreveport, LA
关键词
BIOMECHANICS; INTERNAL SCREW FIXATION; ODONTOID FRACTURE;
D O I
10.1097/00007632-199509000-00001
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. This study examined the torsional and shear stiffness of Type II odontoid fractures that were stabilized with two 3.5-mm cannulated AO screws or a single 4.5-mm cannulated Herbert screw. Objective. To determine the stability of a commonly used two-screw method versus a single-screw fixation method. Insertion of a single screw could make the procedure technically easier. Summary of Background Data. Biomechanical studies have shown the 4.5-mm Herbert screw to generate greater compressive forces than 3.5-mm cannulated screws, which are commonly used to anteriorly stabilize Type II odontoid fractures. No previous biomechanical studies have compared the shear or torsional stiffness of Type II odontoid fractures stabilized with these screws. Methods. Twelve cadaver C2 vertebrae were harvested and an osteotomy was performed to simulate a Type II fracture pattern. Six were stabilized with two 3.5-mm cannulated screws, and the remaining were stabilized with a single 4.5-mm cannulated Herbert screw. Each specimen then was tested in torsion through +/-0.75 degrees, +/-1.25 degrees, and +/-1.75 degrees for 10 cycles each. Shear forces then were applied from an anterior to posterior direction to a maximum load of 44.48 N. Results. The mean torsional stiffness for the Herbert screw was 1196 N . m/deg, which was significantly greater than the mean stiffness of the 3.5-mm screw fixation, which measured 434 N . m/deg. The mean shear stiffness for the Herbert screw fixation was 106.9 kN/m, compared to 86.1 kN/m for the 3.5-mm cannulated screw. This was not statistically significant. Conclusions. The biomechanical properties of the 4.5-mm cannulated Herbert screw suggest it may lend ;itself for use in fixation of Type II odontoid fractures. With only a single screw to insert, this technically demanding procedure may be less daunting.
引用
收藏
页码:1855 / 1859
页数:5
相关论文
共 50 条
  • [2] ANTERIOR SCREW FIXATION OF POSTERIORLY DISPLACED TYPE-II ODONTOID FRACTURES
    GEISLER, FH
    CHENG, C
    POKA, A
    BRUMBACK, RJ
    NEUROSURGERY, 1989, 25 (01) : 30 - 38
  • [3] Anterior Screw Fixation of Type II Odontoid Fractures in the Elderly
    Collins, Iona
    Min, Woo-Kie
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2008, 65 (05): : 1083 - 1087
  • [4] Clinical Results of Anterior Odontoid Screw Fixation for Type II Odontoid Fractures
    Keskin, Fatih
    Gomleksiz, Cengiz
    Sasani, Mehdi
    Oktenoglu, Tunc
    Suzer, Tuncer
    Ozer, Ali Fahir
    TURKISH NEUROSURGERY, 2014, 24 (05) : 750 - 756
  • [5] Avoiding pitfalls in anterior screw fixation for Type II odontoid fractures
    Mazur, Marcus D.
    Mumert, Michael L.
    Bisson, Erica F.
    Schmidt, Meic H.
    NEUROSURGICAL FOCUS, 2011, 31 (04)
  • [6] Odontoid Screw Fixation Repair of Type II Odontoid Fractures
    Kalemci, Orhan
    Yucesoy, Kemal
    Yuksel, Kasim Zafer
    Kizmazoglu, Ceren
    Guner, Metin
    Acar, Umit
    JOURNAL OF NEUROLOGICAL SCIENCES-TURKISH, 2011, 28 (03): : 355 - 361
  • [7] ANTERIOR SCREW FIXATION OF ODONTOID FRACTURES
    MOOSSY, JJ
    HANLEY, EN
    NEUROSURGERY, 1989, 25 (05) : 847 - 847
  • [8] Anterior screw fixation of odontoid fractures
    Morandi, X
    Hanna, A
    Hamlat, A
    Brassier, G
    SURGICAL NEUROLOGY, 1999, 51 (03): : 236 - 240
  • [9] Fixation of odontoid fractures by an anterior screw
    Henry, AD
    Bohly, J
    Grosse, A
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (03): : 472 - 477
  • [10] Anterior screw fixation for odontoid fractures
    Müller, EJ
    Wick, M
    Russe, OJ
    Palta, M
    Muhr, G
    UNFALLCHIRURG, 2000, 103 (01): : 38 - 43