DETERMINATIONS OF ORGAN DOSES AND EFFECTIVE DOSE EQUIVALENTS FROM COMPUTED TOMOGRAPHIC EXAMINATION

被引:73
|
作者
NISHIZAWA, K [1 ]
MARUYAMA, T [1 ]
TAKAYAMA, M [1 ]
OKADA, M [1 ]
HACHIYA, J [1 ]
FURUYA, Y [1 ]
机构
[1] NATL INST RADIOL SCI,DIV PHYS,CHIBA 260,JAPAN
来源
BRITISH JOURNAL OF RADIOLOGY | 1991年 / 64卷 / 757期
关键词
ORGAN DOSE MEASUREMENT; EFFECTIVE DOSE EQUIVALENT; COMPUTED TOMOGRAPHY; MEDICAL EXPOSURE;
D O I
10.1259/0007-1285-64-757-20
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The organ or tissue doses were determined with a phantom measurement for 12 types of CT scanners widely used in Japan. Two types of thermoluminescent dosimeters were used for the dose determinations in a Rando woman phantom. The effective dose equivalents recommended by the International Commission on Radiological Protection were calculated using the measured organ or tissue doses. It was found that the CT scanners currently available give quite different organ or tissue doses. When selecting the optimum technical factors for scanning, therefore, it is important to take into consideration the balance of the image quality and the radiation exposure to patients.
引用
收藏
页码:20 / 28
页数:9
相关论文
共 50 条
  • [41] BETTER ESTIMATING RADIATION EFFECTIVE DOSE FROM CARDIAC COMPUTED TOMOGRAPHY (CT): A COMPREHENSIVE EXAMINATION OF THE RELATIONSHIP BETWEEN EFFECTIVE DOSE AND DOSE-LENGTH PRODUCT
    Trattner, Sigal
    Halliburton, Sandra
    Thompson, Carla
    Chelliah, Anjali
    Xu, Yanping
    Jambawalikar, Sachin
    Peters, M. Robert
    Ghesani, Munir
    Jang, James
    Hoffmann, Udo
    Brenner, David J.
    Einstein, Andrew Jeffrey
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 67 (13) : 1744 - 1744
  • [42] Evaluation of organ doses and specific k effective dose of 64-slice CT thorax examination using an adult anthropomorphic phantom
    Hashim, S.
    Karim, M. K. A.
    Bakar, K. A.
    Sabarudin, A.
    Chin, A. W.
    Saripan, M. I.
    Bradley, D. A.
    RADIATION PHYSICS AND CHEMISTRY, 2016, 126 : 14 - 20
  • [43] EFFECTIVE DOSE EQUIVALENTS FOR PHOTON EXPOSURES FROM PLANE SOURCES ON THE GROUND
    JACOB, P
    PARETZKE, HG
    ROSENBAUM, H
    ZANKL, M
    RADIATION PROTECTION DOSIMETRY, 1986, 14 (04) : 299 - 310
  • [44] Organ and effective doses in newborns and infants undergoing voiding cystourethrograms (VCUG): A comparison of stylized and tomographic phantoms
    Pazik, Frank D.
    Staton, Robert J.
    Williams, Jonathon L.
    Arreola, Manuel M.
    Hintenlang, David E.
    Bolch, Wesley E.
    MEDICAL PHYSICS, 2007, 34 (01) : 294 - 306
  • [45] PROBLEMS IN ESTIMATING EFFECTIVE DOSE EQUIVALENTS FROM THE RESULTS OF BIOASSAYS FOR TRITIUM
    OSBORNE, RV
    JOHNSON, JR
    HEALTH PHYSICS, 1979, 37 (06): : 808 - 808
  • [46] Organ and Effective Dose in Pediatric Patient in 64-Slice Computed Tomography
    Chen, H.
    Lamba, M.
    Alsip, C.
    Lemen, L.
    MEDICAL PHYSICS, 2008, 35 (06) : 2651 - +
  • [47] Effective and Organ Specific Radiation Doses from Videourodynamics in Children
    Hsi, Ryan S.
    Dearn, Jean
    Dean, Melanie
    Zamora, David A.
    Kanal, Kalpana M.
    Harper, Jonathan D.
    Merguerian, Paul A.
    JOURNAL OF UROLOGY, 2013, 190 (04): : 1364 - 1369
  • [48] Evaluation of image quality, organ doses, effective dose, and cancer risk from pediatric brain CT scans
    Kiani, M.
    Chaparian, A.
    EUROPEAN JOURNAL OF RADIOLOGY, 2023, 158
  • [49] ESTIMATION OF THE ORGAN ABSORBED DOSES AND EFFECTIVE DOSE FROM 68Ga-PSMA-11 PET SCAN
    Demirci, Emre
    Toklu, Turkay
    Yeyin, Nami
    Ocak, Meltem
    Alan-Selcuk, Nalan
    Araman, Ahmet
    Kabasakal, Levent
    RADIATION PROTECTION DOSIMETRY, 2018, 182 (04) : 518 - 524
  • [50] Evaluation of Radiation Dose in Coronary Computed Tomography-Angiography Examinations: Using Organ dose and Effective dose
    Shibata, Hideki
    Kondo, Yuya
    Fukada, Shinji
    Asada, Yasuki
    Kozawa, Isao
    RADIATION PHYSICS AND CHEMISTRY, 2019, 158 : 218 - 221