DETERMINATION OF PLASMA ADENOSINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH COLUMN-SWITCHING AND FLUOROMETRIC DETECTION

被引:15
|
作者
NITHIPATIKOM, K
MIZUMURA, T
GROSS, GJ
机构
[1] Department of Pharmacology and Toxicology, Medical College of Wisconsin, Milwaukee
关键词
D O I
10.1006/abio.1994.1585
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A new method for the rapid determination of plasma adenosine concentrations was developed by using high-performance liquid chromatography with a column switching technique and fluorometric detection. Several ''stop solutions'' were used to prevent the enzymatic degradation and cellular uptake and release of adenosine in blood samples. Red blood cells and certain denatured proteins were separated by centrifugation. Subsequently, the supernatant was transferred directly into autosample vials and adenosine was reacted with chloroacetaldehyde to form a strong fluorescent, 1-N-6-ethenoadenosine. The adenosine derivative was injected directly and separated on a shielded hydrophobic phase column coupled with a C-18 reverse-phase column using a column switching valve. Macromolecules and other interfering substances were excluded by the shielded hydrophobic phase column and bypassed to waste. Then, the adenosine derivative and other retained compounds were switched onto the reverse-phase column for further separation and subsequently to the fluorescence detector. The system reduces the analysis time and contamination of the column and hence allows a shorter cleanup time and a longer column lifetime. Adenosine as low as 30 fmol (signal-to-noise ratio, S/N = 3) can be detected by this method. The percentage of recovery of adenosine in plasma treated with adenosine deaminase was above 90%. This method is very rapid (without tedious sample preparation) and sensitive for determining adenosine in canine blood and should prove to be useful in analyzing the effects of ischemia and reperfusion on arterial and coronary venous adenosine concentrations in blood or perfusate samples released from the ischemic or hypoxic myocardium. (C) 1994 Academic Press, Inc.
引用
收藏
页码:280 / 284
页数:5
相关论文
共 50 条
  • [41] Direct Determination of Benzamides in Serum by Column-Switching High-Performance Liquid Chromatography
    Ryoko Chiba
    Ayako Ogasawara
    Teppei Kubo
    Hiroyuki Yamazaki
    Masuo Umino
    Yoichi Ishizuka
    Analytical Sciences, 2003, 19 : 785 - 789
  • [42] Direct determination of benzamides in serum by column-switching high-performance liquid chromatography
    Chiba, R
    Ogasawara, A
    Kubo, T
    Yamazaki, H
    Umino, M
    Ishizuka, Y
    ANALYTICAL SCIENCES, 2003, 19 (05) : 785 - 789
  • [43] DIRECT INJECTION-ANALYSIS OF MELPHALAN IN PLASMA USING COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    KATO, Y
    KANEKO, H
    MATSUSHITA, T
    INAMORI, K
    EGI, S
    TOGAWA, A
    YOKOYAMA, T
    MOHRI, K
    THERAPEUTIC DRUG MONITORING, 1992, 14 (01) : 66 - 71
  • [44] ROUTINE DETERMINATION OF HYDROXYPHENYTOIN IN URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY USING AN AUTOMATIC COLUMN-SWITCHING TECHNIQUE
    JURGENS, U
    JOURNAL OF CHROMATOGRAPHY, 1983, 275 (02): : 335 - 343
  • [45] DETERMINATION OF RESIDUAL ACRYLAMIDE MONOMER IN SOLUTION AND EMULSION POLYMERS BY COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    TSENG, AM
    JOURNAL OF CHROMATOGRAPHY, 1990, 519 (02): : 363 - 368
  • [46] DETERMINATION OF METOPROLOL IN HUMAN-PLASMA BY COLUMN SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    LECAILLON, JB
    SOUPPART, C
    ABADIE, F
    CHROMATOGRAPHIA, 1982, 16 : 158 - 161
  • [47] Column-switching high-performance liquid chromatographic determination of clarithromycin in human plasma with electrochemical detection
    Choi, SJ
    Kim, SB
    Lee, HY
    Na, DH
    Yoon, YS
    Lee, SS
    Kim, JH
    Lee, KC
    Lee, HS
    TALANTA, 2001, 54 (02) : 377 - 382
  • [48] Simultaneous determination of baicalin, rhein and berberine in rat plasma by column-switching high-performance liquid chromatography
    Li Yi
    Gao Jian-Ping
    Xu Xu
    Dai Lixin
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2006, 838 (01): : 50 - 55
  • [49] AUTOMATED-DETERMINATION OF FREE PHENYTOIN IN HUMAN PLASMA WITH ONLINE EQUILIBRIUM DIALYSIS AND COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    ANDRESEN, AT
    RASMUSSEN, KE
    RUGSTAD, HE
    JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1993, 621 (02): : 189 - 198
  • [50] MEASUREMENT OF MALONDIALDEHYDE IN PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH FLUOROMETRIC DETECTION