Uncertainty in positioning ion chamber at reference depth for various water phantoms

被引:7
|
作者
Kinoshita, Naoki [1 ,2 ]
Oguchi, Hiroshi [1 ]
Adachi, Toshiki [2 ]
Shioura, Hiroki [3 ]
Kimura, Hirohiko [3 ]
机构
[1] Nagoya Univ, Grad Sch Med, Dept Radiol & Med Lab Sci, Nagoya, Aichi 4618673, Japan
[2] Univ Fukui Hosp, Radiol Ctr, Eiheiji, Fukui 9101193, Japan
[3] Univ Fukui Hosp, Dept Radiol, Eiheiji, Fukui 9101193, Japan
关键词
Positioning uncertainty; Ionization chamber; Water phantom; External photon beam radiotherapy; Calibration;
D O I
10.1016/j.rpor.2018.03.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Uncertainty in the calibration of high-energy radiation sources is dependent on user and equipment type. Aim: We evaluated the uncertainty in the positioning of a cylindrical chamber at a reference depth for reference dosimetry of high-energy photon beams and the resulting uncertainty in the chamber readings for 6- and 10-MV photon beams. The aim was to investigate major contributions to the positioning uncertainty to reduce the uncertainty in calibration for external photon beam radiotherapy. Materials and methods: The following phantoms were used: Dose View 1D, WP1D, 1D SCAN-NER, and QWP-07 as one-dimensional (1D) phantoms for a vertical-beam geometry; GRI-7632 as a phantom for a fixed waterproofing sleeve; and PTW type 41023 and QWP-04 as 1D phantoms for a horizontal-beam geometry. The uncertainties were analyzed as per the Guide to the Expression of Uncertainty in Measurement. Results: The positioning and resultant uncertainties in chamber readings ranged from 0.22 to 0.35 mm and 0.12-0.25%, respectively, among the phantoms (using a coverage factor k = 1 in both cases). The major contributions to positioning uncertainty are: definition of the origin for phantoms among users for the 1D phantoms for a vertical-beam geometry, water level adjustment among users for the phantom for a fixed waterproofing sleeve, phantom window deformation, and non-water material of the window for the 1D phantoms for a horizontal-beam geometry. Conclusion: The positioning and resultant uncertainties in chamber readings exhibited minor differences among the seven phantoms. The major components of these uncertainties differed among the phantom types investigated. (c) 2018 Greater Poland Cancer Centre. Published by Elsevier Sp.zo.o. All rights reserved.
引用
收藏
页码:199 / 206
页数:8
相关论文
共 50 条
  • [31] Absorbed dose to water reference dosimetry using solid phantoms in the context of absorbed-dose protocols
    Seuntjens, J
    Olivares, M
    Evans, M
    Podgorsak, E
    MEDICAL PHYSICS, 2005, 32 (09) : 2945 - 2953
  • [32] GNSS-Acoustic positioning error in the vertical component considering the uncertainty of a reference sound speed profile
    Tomita, Fumiaki
    Kido, Motoyuki
    Iinuma, Takeshi
    Ohta, Yusaku
    MARINE GEOPHYSICAL RESEARCH, 2025, 46 (01)
  • [33] The Octavius1500 2D ion chamber array and its associated phantoms: Dosimetric characterization of a new prototype
    Van Esch, Ann
    Basta, Katarzyna
    Evrard, Marie
    Ghislain, Michel
    Sergent, Francois
    Huyskens, Dominique P.
    MEDICAL PHYSICS, 2014, 41 (09)
  • [34] Evaluation of ion recombination and polarity effect on photon depth dose measurements using mini- and micro-ion chamber
    Fogliata, Antonella
    Bresolin, Andrea
    Gallo, Pasqualina
    La Fauci, Francesco
    Pelizzoli, Marco
    Reggiori, Giacomo
    Cozzi, Luca
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025, 26 (01):
  • [35] WATER WETTING HEATS OF VARIOUS SILICA GELS WITH REFERENCE TO THEIR DEGREE OF HYDRATION
    EGOROV, MM
    KRASSILNIKOV, KG
    SYSSOEV, EA
    DOKLADY AKADEMII NAUK SSSR, 1956, 108 (01): : 103 - 106
  • [36] Ion chamber dosimetry in non-reference conditions for 15 MV narrow photon fields
    Montaño, EG
    Sánchez-Doblado, F
    Capote, R
    Hermoso, E
    Arrans, R
    Andreo, P
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S184 - S184
  • [38] X-Ray induced depth profiling of ion implantations into various semiconductor materials
    Hoenicke, Philipp
    Mueller, Matthias
    Beckhoff, Burkhard
    ULTRA CLEAN PROCESSING OF SEMICONDUCTOR SURFACES XI, 2013, 195 : 274 - 276
  • [39] Decrease of the intensity of cosmic rays in water to a depth of 440 m measured with counters and ionization chamber
    von Gemert, JCA
    Clay, PH
    PROCEEDINGS OF THE KONINKLIJKE NEDERLANDSE AKADEMIE VAN WETENSCHAPPEN, 1938, 41 (6/10): : 694 - 705
  • [40] Experimental Validation of a Prototype Ion Chamber for Accurate Magnetic Resonance Image Guided Radiation Therapy (MRgRT) Output Measurements in Solid Phantoms
    Muir, B.
    Nusrat, H.
    Sarfehnia, A.
    Renaud, J.
    MEDICAL PHYSICS, 2021, 48 (06)