Uncertainty in positioning ion chamber at reference depth for various water phantoms

被引:7
|
作者
Kinoshita, Naoki [1 ,2 ]
Oguchi, Hiroshi [1 ]
Adachi, Toshiki [2 ]
Shioura, Hiroki [3 ]
Kimura, Hirohiko [3 ]
机构
[1] Nagoya Univ, Grad Sch Med, Dept Radiol & Med Lab Sci, Nagoya, Aichi 4618673, Japan
[2] Univ Fukui Hosp, Radiol Ctr, Eiheiji, Fukui 9101193, Japan
[3] Univ Fukui Hosp, Dept Radiol, Eiheiji, Fukui 9101193, Japan
关键词
Positioning uncertainty; Ionization chamber; Water phantom; External photon beam radiotherapy; Calibration;
D O I
10.1016/j.rpor.2018.03.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Uncertainty in the calibration of high-energy radiation sources is dependent on user and equipment type. Aim: We evaluated the uncertainty in the positioning of a cylindrical chamber at a reference depth for reference dosimetry of high-energy photon beams and the resulting uncertainty in the chamber readings for 6- and 10-MV photon beams. The aim was to investigate major contributions to the positioning uncertainty to reduce the uncertainty in calibration for external photon beam radiotherapy. Materials and methods: The following phantoms were used: Dose View 1D, WP1D, 1D SCAN-NER, and QWP-07 as one-dimensional (1D) phantoms for a vertical-beam geometry; GRI-7632 as a phantom for a fixed waterproofing sleeve; and PTW type 41023 and QWP-04 as 1D phantoms for a horizontal-beam geometry. The uncertainties were analyzed as per the Guide to the Expression of Uncertainty in Measurement. Results: The positioning and resultant uncertainties in chamber readings ranged from 0.22 to 0.35 mm and 0.12-0.25%, respectively, among the phantoms (using a coverage factor k = 1 in both cases). The major contributions to positioning uncertainty are: definition of the origin for phantoms among users for the 1D phantoms for a vertical-beam geometry, water level adjustment among users for the phantom for a fixed waterproofing sleeve, phantom window deformation, and non-water material of the window for the 1D phantoms for a horizontal-beam geometry. Conclusion: The positioning and resultant uncertainties in chamber readings exhibited minor differences among the seven phantoms. The major components of these uncertainties differed among the phantom types investigated. (c) 2018 Greater Poland Cancer Centre. Published by Elsevier Sp.zo.o. All rights reserved.
引用
收藏
页码:199 / 206
页数:8
相关论文
共 50 条
  • [21] Indirect check of the stability of the reference ion chamber used for accelerator output calibration
    Kang, Sei-Kwon
    Yoon, Jai-Woong
    Park, Soah
    Hwang, Taejin
    Cheong, Kwang-Ho
    Han, Tae Jin
    Kim, Haeyoung
    Lee, Me-Yeon
    Kim, Kyoung Ju
    Bae, Hoonsik
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 65 (09) : 1475 - 1480
  • [22] The impact of ion chamber components on k B,Q for reference dosimetry in MRgRT
    Campos, Julia Navarro
    de Pooter, Jacco
    de Prez, Leon
    Jansen, Bartel
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (14):
  • [23] Depth profile of an ion-implanted sample for various dose rates
    Sung, IB
    Kim, JK
    Yoo, YS
    Jeong, BY
    Kang, ST
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2005, 46 (02) : 478 - 481
  • [24] Measurement of ionization chamber angular response and associated uncertainty budget analysis in MRLinac reference dosimetry
    Iakovenko, Viktor
    Keller, Brian
    Sahgal, Arjun
    Sarfehnia, Arman
    MEDICAL PHYSICS, 2019, 46 (11) : 5389 - 5389
  • [25] Study on Change of Topography in Closed Water Area with Positioning Data and Depth Data
    Chen, Narisu
    Xiang, Qing
    Yamamoto, Kazuhiro
    Watanabe, Masaji
    ADVANCED SCIENCE LETTERS, 2017, 23 (04) : 3626 - 3632
  • [26] Stable tissue-simulating phantoms with various water and lipid contents for diffuse optical spectroscopy
    Ohmae, Etsuko
    Yoshizawa, Nobuko
    Yoshimoto, Kenji
    Hayashi, Maho
    Wada, Hiroko
    Mimura, Tetsuya
    Suzuki, Hiroaki
    Homma, Shu
    Suzuki, Norihiro
    Ogura, Hiroyuki
    Nasu, Hatsuko
    Sakahara, Harumi
    Yamashita, Yutaka
    Ueda, Yukio
    BIOMEDICAL OPTICS EXPRESS, 2018, 9 (11): : 5792 - 5807
  • [27] ELECTRON-BEAM ABSORPTION IN SOLID AND IN WATER PHANTOMS - DEPTH SCALING AND ENERGY-RANGE RELATIONS
    GROSSWENDT, B
    ROOS, M
    PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (04): : 509 - 518
  • [28] Dosimetric comparison of water phantoms, ion chambers, and data acquisition modes for LINAC characterization
    Cruz, Wilbert
    Narayanasamy, Ganesh
    Papanikolaou, Niko
    Stathakis, Sotirios
    RADIATION MEASUREMENTS, 2015, 82 : 108 - 114
  • [29] Detour factors in water and plastic phantoms and their use for range and depth scaling in electron-beam dosimetry
    FernandezVarea, JM
    Andreo, P
    Tabata, T
    PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (07): : 1119 - 1139
  • [30] Evaluation of the exradin A19 ion chamber for reference dosimetry in megavoltage photon beams
    McEwen, M.
    MEDICAL PHYSICS, 2007, 34 (06) : 2451 - 2451