OPTIMIZED BIOSENSOR FOR WHOLE-BLOOD MEASUREMENTS USING A NEW CELLULOSE-BASED MEMBRANE

被引:20
|
作者
EISELE, S
AMMON, HPT
KINDERVATER, R
GROBE, A
GOPEL, W
机构
[1] FRAUNHOFER INST SURFACES & BIOENGN,W-7000 STUTTGART,GERMANY
[2] UNIV TUBINGEN,INST PHYS & THEORET CHEM,W-7400 TUBINGEN 1,GERMANY
来源
BIOSENSORS & BIOELECTRONICS | 1994年 / 9卷 / 02期
关键词
AMPEROMETRIC; GLUCOSE SENSOR; GOD; LONG-TERM STABILITY; LINEARITY RANGE; CUPROPHAN(R); BACTERIAL CELLULOSE; DILUTED BLOOD; WHOLE BLOOD;
D O I
10.1016/0956-5663(94)80102-9
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
We used bacterial cellulose membranes as outer membranes for glucose sensors based upon amperometric detection of glucose oxidase reaction products. We found a long-term stability in 1:10 diluted blood of about 200 h. In comparison a Cuprophan(R) membrane was stable for only 30 h under the same conditions. Comparing the two membranes in undiluted human blood, a similar trend was found in the long-term stability: Cuprophan(R) was stable for only 3-4 h and the glucose sensor with the bacterial cellulose membrane was stable for more than 24 h. In addition, the measuring range of the glucose sensor could be extended up to 170 mM glucose by treating the bacterial cellulose membrane with a polyamide.
引用
收藏
页码:119 / 124
页数:6
相关论文
共 50 条
  • [1] PROBLEMS BY USING WHOLE-BLOOD IN HEPARIN MEASUREMENTS
    TALSTAD, I
    THROMBOSIS AND HAEMOSTASIS, 1982, 47 (02) : 177 - 181
  • [2] UREA-SENSING MEMBRANE ELECTRODE FOR WHOLE-BLOOD MEASUREMENTS
    PAPASTATHOPOULOS, DS
    RECHNITZ, GA
    ANALYTICA CHIMICA ACTA, 1975, 79 (OCT) : 17 - 26
  • [3] WHOLE-BLOOD - PLASMA SEPARATION USING MEMBRANE FILTRATION
    WILTBANK, TB
    SOLOMAN, BA
    LYSAGHT, MJ
    CASTINO, F
    FRIEDMAN, LI
    BLOOD, 1977, 50 (05) : 312 - 312
  • [4] WHOLE-BLOOD MEASUREMENTS WITH THERMAL MICROBIOSENSORS
    DANIELSSON, B
    XIE, B
    HEDBERG, U
    SENSORS AND ACTUATORS B-CHEMICAL, 1993, 14 (1-3) : 758 - 759
  • [6] AMPEROMETRIC GLUCOSE BIOSENSOR FOR AN UNDILUTED WHOLE-BLOOD ANALYSIS
    MATUSZEWSKI, W
    TROJANOWICZ, M
    LEWENSTAM, A
    MOSZCZYNSKA, A
    LANGEMOROZ, E
    ANALYTICAL SCIENCES, 1994, 10 (03) : 423 - 428
  • [7] A Peptide-Based Electrochemical Biosensor for Facile Measurement of Whole-Blood Heparin
    Meng, Fanyu
    Liang, Weiguo
    Sun, Haixuan
    Wu, Ligang
    Gong, Xiao
    Miao, Peng
    CHEMELECTROCHEM, 2017, 4 (03): : 472 - 475
  • [8] NEW MICROMETHOD FOR LYMPHOCYTE STIMULATION USING WHOLE-BLOOD
    ESKOLA, J
    SOPPI, E
    VILJANEN, M
    RUUSKANEN, O
    IMMUNOLOGICAL COMMUNICATIONS, 1975, 4 (04): : 297 - 307
  • [9] Review on cellulose-based membrane materials
    Gao C.-T.
    Cai L.
    Dong F.-P.
    Xie H.-B.
    Qin S.-H.
    Zheng Q.
    Gao Xiao Hua Xue Gong Cheng Xue Bao/Journal of Chemical Engineering of Chinese Universities, 2019, 33 (05): : 1037 - 1047
  • [10] CLINICAL USE OF WHOLE-BLOOD VISCOSITY MEASUREMENTS
    LOWE, GDO
    BRITISH JOURNAL OF HAEMATOLOGY, 1981, 49 (01) : 145 - 146