DYNAMIC FIELD SHAPING FOR STEREOTAXIC RADIOSURGERY - A MODELING STUDY

被引:62
|
作者
NEDZI, LA
KOOY, HM
ALEXANDER, E
SVENSSON, GK
LOEFFLER, JS
机构
[1] HARVARD UNIV,CHILDRENS HOSP,SCH MED,BOSTON,MA 02115
[2] HARVARD UNIV,BRIGHAM & WOMENS HOSP,SCH MED,STEREOTACT RADIOSURG PROGRAM,BOSTON,MA 02115
关键词
STEREOTAXIC RADIOSURGERY; MULTI-LEAF COLLIMATOR; CONFORMAL THERAPY; TREATMENT PLANNING;
D O I
10.1016/0360-3016(93)90316-N
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: This work assesses the relative field shaping advantages of dynamic field shaping devices for stereotactic radiosurgery using a linear accelerator. Methods and Materials: We selected 43 intracranial tumors (2.0-4.2 cm maximum dimension, 1.5-25.5 cc tumor volume) out of the first 64 intracranial tumors treated with radiosurgery at the Joint Center for Radiation Therapy. We modeled five field shaping devices, each including a fixed auxilliary circular collimator: (a) fixed circular collimator alone; (b) two independent parallel jaws; (c) four independent rectangular jaws; (d) four independent rotatable jaws; and (e) ''ideal'' multileaf collimator. We adjusted the model parameters until the minimum target isodose was 80% of the dose delivered to isocenter. We defined the treatment volume ratio as the target volume divided by the treatment volume (volume receiving at least the minimum target dose). We used the treatment volume ratio to compare the five models and the actual patient treatments. Results: For 34 tumors originally treated with one isocenter, the median Treatment Volume Ratio was higher for all of the device models except the fixed circular collimator compared to the actual patient treatments. For the nine tumors originally treated with multiple isocenters, the median Treatment Volume Ratio for the actual multiple isocenter treatments was similar to that for two parallel jaws, four rectangular jaws and four rotatable jaws. Only the median ''ideal'' collimator treatment volume ratio was higher for these nine tumors. Conclusion: Simple field shaping devices have approximately 50% of the conformal advantage of an ''ideal'' multileaf collimator. Approximately 50% of typical radiosurgical tumors between 2 and 4 cm have field shaping advantages which exceed the geometrical uncertainties inherent in linear accelerator radiosurgery treatments. The three models, two parallel, four rectangular, or four rotatable independent jaws would improve current linear accelerator technology by providing homogeneous doses with equivalent field shaping for most tumors originally treated with inhomogeneous multiple isocenter plans (6/9 tumors in the current series).
引用
收藏
页码:859 / 869
页数:11
相关论文
共 50 条
  • [41] TREATMENT OF ACOUSTIC NEURINOMAS BY STEREOTAXIC RADIOSURGERY
    NOREN, G
    ACTA NEUROCHIRURGICA, 1977, 37 (3-4) : 311 - 311
  • [42] STEREOTAXIC RADIOSURGERY FOR CEREBRAL METASTATIC MELANOMA
    SOMAZA, S
    KONDZIOLKA, D
    LUNSFORD, LD
    KIRKWOOD, JM
    FLICKINGER, JC
    JOURNAL OF NEUROSURGERY, 1993, 79 (05) : 661 - 666
  • [43] DOSIMETRY AND DOSE PLANNING IN STEREOTAXIC RADIOSURGERY
    WALTON, L
    BOMFORD, CK
    CLINICAL RADIOLOGY, 1987, 38 (05) : 554 - 554
  • [44] STEREOTAXIC RADIOSURGERY - A REVIEW AND COMPARISON OF METHODS
    PHILLIPS, MH
    STELZER, KJ
    GRIFFIN, TW
    MAYBERG, MR
    WINN, HR
    JOURNAL OF CLINICAL ONCOLOGY, 1994, 12 (05) : 1085 - 1099
  • [45] FLUORESCENT SCREEN VIDEO IMAGING DOR CHECKING LINAC BEAMS IN DYNAMIC STEREOTAXIC RADIOSURGERY
    LIGHTSTONE, AW
    RAMANI, R
    OBRIEN, PF
    GILLIES, BA
    MEDICAL PHYSICS, 1994, 21 (12) : 1991 - 1996
  • [46] Dynamic modeling and input shaping for MEMS
    Popa, DO
    Wen, JT
    Stephanou, H
    Skidmore, G
    Ellis, M
    NSTI NANOTECH 2004, VOL 2, TECHNICAL PROCEEDINGS, 2004, : 315 - 318
  • [47] STEREOTAXIC RADIOSURGERY FOR ARTERIOVENOUS-MALFORMATIONS OF THE BRAIN
    LUNSFORD, LD
    KONDZIOLKA, D
    FLICKINGER, JC
    BISSONETTE, DJ
    JUNGREIS, CA
    MAITZ, AH
    HORTON, JA
    COFFEY, RJ
    JOURNAL OF NEUROSURGERY, 1991, 75 (04) : 512 - 524
  • [48] THYROID DOSES DUE TO STEREOTAXIC RADIOSURGERY OF THE BRAIN
    SHIN, K
    SIBATA, C
    DEALMEIDA, CE
    DESOUZA, C
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 : 261 - 261
  • [49] TREATMENT PLANNING FOR STEREOTAXIC RADIOSURGERY OF INTRACRANIAL LESIONS
    KOOY, HM
    NEDZI, LA
    LOEFFLER, JS
    ALEXANDER, E
    CHENG, CW
    MANNARINO, EG
    HOLUPKA, EJ
    SIDDON, RL
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (03): : 683 - 693
  • [50] STEREOTAXIC RADIOSURGERY UTILIZING A LINEAR-ACCELERATOR
    COLOMBO, F
    BENEDETTI, A
    POZZA, F
    ZANARDO, A
    AVANZO, RC
    CHIEREGO, G
    MARCHETTI, C
    APPLIED NEUROPHYSIOLOGY, 1985, 48 (1-6) : 133 - 145