APPROACHING COMPARABILITY AND RESULTS OF PULSATILE FLOW IN-VITRO TESTING OF PROSTHETIC HEART-VALVES

被引:7
|
作者
BARBARO, V [1 ]
BOCCANERA, G [1 ]
DANIELE, C [1 ]
GRIGIONI, M [1 ]
PALOMBO, A [1 ]
机构
[1] UNIV ROMA LA SAPIENZA,FAC ENGN,DEPT ELECTR,ROME,ITALY
关键词
D O I
10.3109/03091909509012416
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The testing of prosthetic heart values under pulsatile conditions is still a subject for debate among researchers and competent standardization bodies. The laboratory of Biomedical Engineering of the Istituto Superiore di Sanita in Rome, has reproduced the current inter-laboratory situation with several test apparatuses, focusing on the definition of significant measurement parameters and procedures to obtain reasonably comparable data. The laboratory is also equipped with a Laser Doppler Anemometer (LDA) and a High-Speed Cinematographic system (HSC). A 29 mm tilting disc valve model, was mounted in the aortic position. Under tightly controlled system conditions the analyses performed on two pulse duplicators (PDs) may be deemed consistent for the valve model tested. Useful results, an the same value specimen, are reported concerning velocity profiles and turbulent shear stress values (TSS). Furthermore valve motion on the Sheffield PD was monitored during the closing phase, and related cinematic data reported. The applied methodologies can provide relevant data to support surgeon decision making.
引用
收藏
页码:115 / 118
页数:4
相关论文
共 50 条
  • [41] THE CURRENT STATUS OF PROSTHETIC HEART-VALVES
    YOGANATHAN, AP
    HARRISON, EC
    FRANCH, RH
    ACS SYMPOSIUM SERIES, 1984, 256 : 111 - 150
  • [42] PROSTHETIC HEART-VALVES - SCYLLA AND CHARYBDIS
    NORMAN, JC
    DENNIS, C
    WIETING, D
    BRAUNSWALD, NS
    CARDIOVASCULAR DISEASES, 1978, 5 (02): : 91 - 101
  • [43] NONINVASIVE ASSESSMENT OF PROSTHETIC HEART-VALVES
    CHAN, KL
    CANADIAN JOURNAL OF CARDIOLOGY, 1988, 4 (08) : 417 - 421
  • [44] PROSTHETIC HEART-VALVES - DESIGN CONSIDERATIONS
    TASCON, M
    ANNALS OF THORACIC SURGERY, 1989, 48 (03): : S16 - S17
  • [45] PROSTHETIC HEART-VALVES, ANTICOAGULANTS AND PREGNANCY
    IBARRAPEREZ, C
    NIZRAMOS, J
    GARCIAALONSO, A
    ECHAVARRIALUNA, J
    CHEST, 1985, 88 : S11 - S11
  • [46] PREGNANCY IN PATIENTS WITH PROSTHETIC HEART-VALVES
    ROBBINS, G
    KERNOFF, PBA
    BRITISH MEDICAL JOURNAL, 1983, 287 (6385): : 134 - 134
  • [47] BACTEREMIA IN PATIENTS WITH PROSTHETIC HEART-VALVES
    TORSETH, J
    ASHIKAGA, T
    BEATY, HN
    CLINICAL RESEARCH, 1989, 37 (04): : A911 - A911
  • [48] DYNAMIC CHARACTERISTICS OF PROSTHETIC HEART-VALVES
    SHE, JW
    LI, MY
    HUAN, LJ
    YU, YZ
    MEDICAL ENGINEERING & PHYSICS, 1995, 17 (04) : 273 - 281
  • [49] Multidetector CT imaging of mechanical prosthetic heart valves: quantification of artifacts with a pulsatile in-vitro model
    Petr Symersky
    Ricardo P. J. Budde
    Paul Westers
    Bas A. J. M. de Mol
    Mathias Prokop
    European Radiology, 2011, 21 : 2103 - 2110
  • [50] Multidetector CT imaging of mechanical prosthetic heart valves: quantification of artifacts with a pulsatile in-vitro model
    Symersky, Petr
    Budde, Ricardo P. J.
    Westers, Paul
    de Mol, Bas A. J. M.
    Prokop, Mathias
    EUROPEAN RADIOLOGY, 2011, 21 (10) : 2103 - 2110