APPROACHING COMPARABILITY AND RESULTS OF PULSATILE FLOW IN-VITRO TESTING OF PROSTHETIC HEART-VALVES

被引:7
|
作者
BARBARO, V [1 ]
BOCCANERA, G [1 ]
DANIELE, C [1 ]
GRIGIONI, M [1 ]
PALOMBO, A [1 ]
机构
[1] UNIV ROMA LA SAPIENZA,FAC ENGN,DEPT ELECTR,ROME,ITALY
关键词
D O I
10.3109/03091909509012416
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The testing of prosthetic heart values under pulsatile conditions is still a subject for debate among researchers and competent standardization bodies. The laboratory of Biomedical Engineering of the Istituto Superiore di Sanita in Rome, has reproduced the current inter-laboratory situation with several test apparatuses, focusing on the definition of significant measurement parameters and procedures to obtain reasonably comparable data. The laboratory is also equipped with a Laser Doppler Anemometer (LDA) and a High-Speed Cinematographic system (HSC). A 29 mm tilting disc valve model, was mounted in the aortic position. Under tightly controlled system conditions the analyses performed on two pulse duplicators (PDs) may be deemed consistent for the valve model tested. Useful results, an the same value specimen, are reported concerning velocity profiles and turbulent shear stress values (TSS). Furthermore valve motion on the Sheffield PD was monitored during the closing phase, and related cinematic data reported. The applied methodologies can provide relevant data to support surgeon decision making.
引用
收藏
页码:115 / 118
页数:4
相关论文
共 50 条
  • [21] PREGNANCY AND PROSTHETIC HEART-VALVES
    ANTRETTER, H
    BONATTI, J
    LANCET, 1994, 344 (8937): : 1644 - 1644
  • [22] PROSTHETIC HEART-VALVES FOR THE ELDERLY
    BEALL, AC
    MEDICAL INSTRUMENTATION, 1982, 16 (04): : 205 - 206
  • [23] THROMBOEMBOLISM AND PROSTHETIC HEART-VALVES
    CRISCITIELLO, MG
    LEVINE, HJ
    HOSPITAL PRACTICE, 1992, 27 (12): : 69 - &
  • [24] PROSTHETIC HEART-VALVES - A REVIEW
    MCCLUNG, JA
    STEIN, JH
    AMBROSE, JA
    HERMAN, MV
    REED, GE
    PROGRESS IN CARDIOVASCULAR DISEASES, 1983, 26 (03) : 237 - 270
  • [25] PRESSURE DROPS ACROSS PROSTHETIC AORTIC HEART-VALVES UNDER STEADY AND PULSATILE FLOW - INVITRO MEASUREMENTS
    YOGANATHAN, AP
    CORCORAN, WH
    HARRISON, EC
    JOURNAL OF BIOMECHANICS, 1979, 12 (02) : 153 - 164
  • [26] PROSTHETIC HEART-VALVES - WAVE-FORM COMPARISONS AND AVERAGE VALUE DISPOSITION FOR PULSATILE FLOW INVITRO
    SHE, JW
    LI, MY
    WANG, KQ
    YU, YH
    JOURNAL OF BIOMEDICAL ENGINEERING, 1990, 12 (05): : 375 - 378
  • [27] INVITRO EVALUATION OF PROSTHETIC HEART-VALVES - TOWARDS COMPARABLE TESTING
    BARBARO, V
    DANIELE, C
    GRIGIONI, M
    PALOMBO, A
    SARGENTINI, A
    JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1992, 16 (01) : 10 - 14
  • [28] MALFUNCTIONS OF PROSTHETIC HEART-VALVES - FREQUENCY, DIAGNOSIS, REOPERATION RESULTS
    HORSTKOTTE, D
    SCHULTE, HD
    BIRCKS, W
    LOOGEN, F
    ZEITSCHRIFT FUR KARDIOLOGIE, 1985, 74 : 79 - 79
  • [29] VENTRICULAR FLOW DYNAMIC PAST BILEAFLET PROSTHETIC HEART-VALVES
    GARITEY, V
    GANDELHEID, T
    FUSERI, J
    PELISSIER, R
    RIEU, R
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1995, 18 (07): : 380 - 391
  • [30] FUNCTIONAL GEOMETRICAL PARAMETERS OF THE FLOW THROUGH PROSTHETIC HEART-VALVES
    KOHLER, J
    THOLE, S
    LIFE SUPPORT SYSTEMS, 1986, 4 (04): : 345 - 348