DISTANCE AND VOLUME MEASUREMENT USING 3-DIMENSIONAL ULTRASONOGRAPHY

被引:0
|
作者
RICCABONA, M [1 ]
NELSON, TR [1 ]
PRETORIUS, DH [1 ]
DAVIDSON, TE [1 ]
机构
[1] UNIV CALIF SAN DIEGO,DEPT RADIOL 0610,LA JOLLA,CA 92093
关键词
3-DIMENSIONAL ULTRASONOGRAPHY; VOLUME; MEASUREMENTS; 2-DIMENSIONAL ULTRASONOGRAPHY;
D O I
暂无
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The purpose of this study was to assess the accuracy of distance and volume measurements obtained by three-dimensional ultrasonography. A tissue-mimicking phantom was scanned using a prototype three-dimensional sonographic imaging system to verify distance measurements. Measurements were taken from the reconstructed three-dimensional sonographic data and compared to the real distances. Volume measurements were obtained by scanning 30 balloons of various shapes, sized 23 mi to 2400 mi. Each balloon was scanned twice in two orientations; three different masks were accomplished for each volume. Each volume measurement of 180 three-dimensional sonographic measurements was compared to conventional two-dimensional ultrasonographic volume estimates and to the actual, measured balloon size. Distance measurements had a mean error of 0.02 +/- 3.65% (range, -4.27 to 7.18%). Two-dimensional sonographic volume estimates using traditional scanner based methods had a mean error of 13.7 +/- 10.1%. Three-dimensional sonographic volume measurements had a mean error of 2.2 +/- 2.9% for regular and irregular objects over the entire range of volumes. The masking required 10 to 30 min. The field of view varied from 10 to 24 cm with a mean object depth of 9.8 cm. Three-dimensional ultrasonographic methods can provide accurate volume measurements of regular and irregular objects ana offer improved accuracy compared to traditional two-dimensional methods.
引用
收藏
页码:881 / 886
页数:6
相关论文
共 50 条
  • [31] Measurement of human muscle volume using ultrasonography
    Esformes, JI
    Narici, MV
    Maganaris, CN
    EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2002, 87 (01) : 90 - 92
  • [32] Measurement of human muscle volume using ultrasonography
    Joseph I. Esformes
    Marco V. Narici
    Constantinos N. Maganaris
    European Journal of Applied Physiology, 2002, 87 : 90 - 92
  • [33] 3-DIMENSIONAL DIAGNOSIS OF SUBMUCOSAL TUMOR (SMT) USING ENDOSCOPIC ULTRASONOGRAPHY (EUS)
    YASUDA, K
    KIYOTA, K
    FUJIMOTO, S
    NAKAJIMA, M
    KAWAI, K
    GASTROINTESTINAL ENDOSCOPY, 1987, 33 (02) : 180 - 180
  • [34] Measurement of orbital volume by a 3-dimensional software program: An experimental study - Discussion
    Marsh, JL
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2000, 58 (06) : 648 - 648
  • [35] Measurement of Carotid Plaque Volume with VOCAL™II Technique by 3-Dimensional Ultrasound
    Xing, Haiying
    Wang, Tingting
    Huang, Yining
    CEREBROVASCULAR DISEASES, 2013, 36 : 88 - 88
  • [36] VOLUME MEASUREMENTS IN 3-DIMENSIONAL MICROSCOPY
    VISSER, TD
    OUD, JL
    SCANNING, 1994, 16 (04) : 198 - 200
  • [37] An evaluation of three-dimensional ultrasonography for the measurement of gallbladder volume
    Hashimoto, S
    Goto, H
    Hirooka, Y
    Itoh, A
    Ishiguro, Y
    Kojima, S
    Hirai, T
    Hayakawa, T
    Naitoh, Y
    AMERICAN JOURNAL OF GASTROENTEROLOGY, 1999, 94 (12): : 3492 - 3496
  • [38] A New Diagnostic Method for Early Gastric Cancer: Volume Measurement by 3-Dimensional Endoscopic Ultrasonography in Early Gastric Cancer and Its Clinical Significance
    Futawatari, Nobue
    Kikuchi, Shiro
    Sakuramoto, Shinichi
    Kida, Mitsuhiro
    Watanabe, Masahiko
    ANTICANCER RESEARCH, 2008, 28 (5B) : 2907 - 2912
  • [39] Prenatal diagnosis of spondylocostal dysostosis with 3-dimensional ultrasonography
    Sallout, BL
    D'Agostini, DA
    Pretorius, DH
    JOURNAL OF ULTRASOUND IN MEDICINE, 2006, 25 (04) : 539 - 543
  • [40] Applications of 3-Dimensional Ultrasonography in the Neck, Excluding the Thyroid
    Frank, Susan J.
    Koenigsberg, Tova
    Gutman, David
    Koenigsberg, Mordecai
    JOURNAL OF ULTRASOUND IN MEDICINE, 2018, 37 (07) : 1791 - 1806