Impact of 6MV photon beam attenuation by carbon fiber couch and immobilization devices in IMRT planning and dose delivery

被引:26
|
作者
Munjal, R. K. [1 ]
Negi, P. S. [1 ]
Babu, A. G. [1 ]
Sinha, S. N. [1 ]
Anand, A. K. [1 ]
Kataria, T. [1 ]
机构
[1] Rajiv Gandhi Canc Inst & Res Ctr, Delhi, India
关键词
Attenuation; base plate; carbon fiber; dosimetry; intensity modulated radiation therapy; treatment couch;
D O I
10.4103/0971-6203.26690
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Multiple fields in IMRT and optimization allow conformal dose to the target and reduced dose to the surroundings and the regions of interest. Thus we can escalate the dose to the target to achieve better tumor control with low morbidity. Orientation of multiple beams can be achieved by i) different gantry angles, ii) rotating patient's couch isocentrically. In doing so, one or more beam may pass through different materials like the treatment couch, immobilization cast fixation plate, head and neck rest or any other supportive device. Our observations for 6MV photon beam on PRIMUS-KXE2 with MED-TEC carbon fiber tabletop and 10 x 10 cm(2) field size reveals that the maximum dose attenuation by the couch was of the order of 2.96% from gantry angle 120-160 degrees. Attenuation due to cast fixation base plate of PMMA alone was of the order of 5.8-10.55% at gantry angle between 0 and 90 degrees. Attenuation due to carbon fiber base plate alone was 3.8-7.98%. Attenuation coefficient of carbon fiber and PMMA was evaluated and was of the order of 0.082 cm(-1) and 0.064 cm(-1) respectively. Most of the TPS are configured for direct beam incidence attenuation correction factors only. Whereas when the beam is obliquely incident on the couch, base plate, headrest and any other immobilization device get attenuated more than the direct beam incidence. The correction factors for oblique incidence beam attenuation are not configured in most of the commercially available treatment planning systems. Therefore, such high variations in dose delivery could lead to under-dosage to the target volume for treatments requiring multiple fields in IMRT and 3D-CRT and need to be corrected for monitor unit calculations.
引用
收藏
页码:67 / 71
页数:5
相关论文
共 30 条
  • [21] Surface dose for five telecobalt machines, 6MV photon beam from four linear accelerators and a Hi-Art TomoTherapy
    Kinhikar, Rajesh A.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2008, 7 (05) : 381 - 384
  • [22] Surface dose measurements in chest wall postmastectomy radiotherapy to achieve optimal dose delivery with 6 MV photon beam
    Lobo, Dilson
    Banerjee, Sourjya
    Srinivas, Challapalli
    Athiyamaan, M. S.
    Reddy, Shreyas
    Sunny, Johan
    Ravichandran, Ramamoorthy
    Kotian, Himani
    Prakash Saxena, P. U.
    JOURNAL OF MEDICAL PHYSICS, 2021, 46 (04) : 324 - 333
  • [23] The effect of carbon fibre treatment couch with and without immobilisation devices on radiotherapy dose calculation using three different planning algorithms and photon beam energies
    Oulhouq, Y.
    Rrhioua, A.
    Bakari, D.
    Zerfaoui, M.
    Krim, D.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2022, 21 (04) : 501 - 505
  • [24] Determination of the primary dose component and primary linear attenuation coefficient in a 6 MV photon beam using a small attenuator
    Trauernicht, Christoph Jan
    RADIATION MEASUREMENTS, 2010, 45 (10) : 1465 - 1468
  • [25] SKIN DOSE IMPACT FROM VACUUM IMMOBILIZATION DEVICE AND CARBON FIBER COUCH IN INTENSITY MODULATED RADIATION THERAPY FOR PROSTATE CANCER
    Lee, Kuo-Wei
    Wu, Jian-Kuen
    Jeng, Shiu-Chen
    Liu, Yen-Wan Hsueh
    Cheng, Jason Chia-Hsien
    MEDICAL DOSIMETRY, 2009, 34 (03) : 228 - 232
  • [26] Dosimetry and Implementation of High-Dose Rate MLC-Based GRID Therapy Using a 6MV Flattened Filter Free (FFF) Photon Beam
    Dou, K.
    Lerma, F.
    Jacobs, M.
    Li, B.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [27] Dose attenuation effect of hip prostheses in a 9-MV photon beam: Commercial treatment planning system versus Monte Carlo calculations
    Mesbahi A.
    Nejad F.S.
    Radiation Medicine, 2007, 25 (10): : 529 - 535
  • [28] Investigation of AAA dose calculation algorithm accuracy in surface and buildup region for 6MV photon beam using markus parallel-plate ion chamber
    Kesen, Nazmiye Donmez
    Akbas, Ugur
    Koksal, Canan
    Bilge, Hatice
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2019, 27 (02) : 361 - 369
  • [29] Over-Response Correction of SNC 350p ROOS and PTW Markus Chambers in the Surface Dose and Build-Up Region Dosimetry of 6MV Photon Beam
    Zope, Mukesh Kumar
    Patil, Deepali Bhaskar
    Saroj, Dinesh Kumar
    Aslam, P.A.
    George, Basil
    Iranian Journal of Medical Physics, 2024, 21 (04) : 272 - 279
  • [30] Impact of treatment planning with deformable image registration on dose distribution for carbon-ion beam lung treatment using a fixed irradiation port and rotating couch
    Kumagai, M.
    Mori, S.
    Yamamoto, N.
    BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1050):