Limitations of current centrifugal blood pumps are related to heat generation of bearings and leakage of seals, to dead water zones, and to poor efficiency. A new concept is proposed in this paper to ameliorate these problems based on a miniaturized magnetic drive, and a prototype is introduced. The pump rotor is suspended and driven by a radial permanent magnetic field that stabilizes the impeller in 4 of the 6 spatial degrees of freedom and allows it to be top-spun on a single blood-flushed pivot bearing with minimal load and friction. A shrouded impeller with an open center and 4 logarithmically curved channels is run inside a cone-and-plate-type housing with a spiral volute chamber. In vitro testing was performed comparing this design with the BioMedicus, St. Jude, and Sarns pumps. The prototype is demonstrated to have the smallest internal volume (35 ml), surface (190 gcm), and passage time (0.5 s at 4 L/min), as well as the highest hydraulic efficiency (up to 0.4) of all devices studied.
引用
收藏
页码:620 / 624
页数:5
相关论文
共 50 条
  • [41] Estimation of changes in dynamic hydraulic force in a magnetically suspended centrifugal blood pump with transient computational fluid dynamics analysis
    Toru Masuzawa
    Akiko Ohta
    Nobuatu Tanaka
    Yi Qian
    Tomonori Tsukiya
    [J]. Journal of Artificial Organs, 2009, 12 : 150 - 159
  • [42] More than 1 year continuous operation of a centrifugal pump with a magnetically suspended impeller
    Nojiri, C
    Kijima, T
    Maekawa, J
    Horiuchi, K
    Kido, T
    Sugiyama, T
    Mori, T
    Sugiura, N
    Asada, T
    Shimane, H
    Nishimura, K
    Ban, T
    Akamatsu, T
    Ozaki, T
    Ito, H
    Suzuki, M
    Akutsu, T
    [J]. ASAIO JOURNAL, 1997, 43 (05) : M548 - M552
  • [43] In vivo testing of a magnetically suspended centrifugal pump designed for long-term use
    Yamada, T
    Nishimura, K
    Akamatsu, T
    Tsukiya, T
    Park, CH
    Kono, S
    Matsuda, K
    Ban, T
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1997, 20 (10): : 562 - 569
  • [44] Control of the pressure flow relationship with a magnetically suspended centrifugal pump in a chronic animal experiment
    Nishimura, K
    Tsukiya, T
    Akamatsu, T
    Kono, S
    Yamada, T
    Ban, T
    [J]. ASAIO JOURNAL, 1997, 43 (05) : M553 - M556
  • [45] A magnetically levitated centrifugal blood pump with a simple-structured disposable pump head
    Hijikata, Wataru
    Shinshi, Tadahiko
    Asama, Junichi
    Li, Lichuan
    Hoshi, Hideo
    Takatani, Setsuo
    Shimokohbe, Akira
    [J]. ARTIFICIAL ORGANS, 2008, 32 (07) : 531 - 540
  • [46] Compensation On-line of Errors Caused by Rotor Centrifugal Deformation for a Magnetically Suspended Sensitive Gyroscope
    Xin, Chao-Jun
    Cai, Yuan-Wen
    Ren, Yuan
    Fan, Ya-Hong
    [J]. JOURNAL OF ELECTRICAL ENGINEERING & TECHNOLOGY, 2018, 13 (02) : 1030 - 1041
  • [47] Disposable magnetically levitated centrifugal blood pump: Design and in vitro performance
    Hoshi, H
    Asama, J
    Shinshi, T
    Ohuchi, K
    Nakamura, M
    Mizuno, T
    Arai, H
    Shimokohbe, A
    Takatani, S
    [J]. ARTIFICIAL ORGANS, 2005, 29 (07) : 520 - 526
  • [48] Hydraulic Analysis and Optimization in impeller of Magnetically Levitated Centrifugal Blood Pump
    Zhang, Huimin
    Zhu, Qi
    [J]. FRONTIERS OF MANUFACTURING AND DESIGN SCIENCE II, PTS 1-6, 2012, 121-126 : 1204 - 1208
  • [49] HeartQuest ventricular assist device magnetically levitated centrifugal blood pump
    Bearnson, GB
    Jacobs, GB
    Kirk, J
    Khanwilkar, PS
    Nelson, KE
    Long, JW
    [J]. ARTIFICIAL ORGANS, 2006, 30 (05) : 339 - 346
  • [50] Less-invasive tools and technique for fully magnetically levitated centrifugal pump implantation
    Netuka, Ivan
    Konarik, Miroslav
    Pokorny, Martin
    Ivak, Peter
    Maly, Jiri
    Szarszoi, Ondrej
    [J]. ANNALS OF CARDIOTHORACIC SURGERY, 2021, 10 (02) : 289 - 291