A magnetically levitated centrifugal blood pump with a simple-structured disposable pump head

被引:41
|
作者
Hijikata, Wataru [2 ]
Shinshi, Tadahiko [1 ]
Asama, Junichi [1 ]
Li, Lichuan [3 ]
Hoshi, Hideo [4 ]
Takatani, Setsuo [4 ]
Shimokohbe, Akira [1 ]
机构
[1] Tokyo Inst Technol, Precis & Intelligence Lab, Midori Ku, Yokohama, Kanagawa 2268503, Japan
[2] Interdisciplinary Grad Sch Sci & Engn, Yokohama, Kanagawa, Japan
[3] Xi An Jiao Tong Univ, Sch Elect Engn, Xian 710049, Peoples R China
[4] Tokyo Med & Dent Univ, Inst Biomat & Bioengn, Tokyo, Japan
关键词
centrifugal blood pump; disposable pump head; magnetic bearing; magnetic coupling; heat generation;
D O I
10.1111/j.1525-1594.2008.00576.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A magnetically levitated centrifugal blood pump (MedTech Dispo) has been developed for use in a disposable extracorporeal system. The design of the pump is intended to eliminate mechanical contact with the impeller, to facilitate a simple disposable mechanism, and to reduce the blood-heating effects that are caused by motors and magnetic bearings. The bearing rotor attached to the impeller is suspended by a two degrees-of-freedom controlled radial magnetic bearing stator, which is situated outside the rotor. In the space inside the ringlike rotor, a magnetic coupling disk is placed to rotate the rotor and to ensure that the pump head is thermally isolated from the motor. In this system, the rotor can exhibit high passive stiffness due to the novel design of the closed magnetic circuits. The disposable pump head, which has a priming volume of 23 mL, consists of top and bottom housings, an impeller, and a rotor with a diameter of 50 mm. The pump can provide a head pressure of more than 300 mm Hg against a flow of 5 L/min. The normalized index of hemolysis of the MedTech Dispo is 0.0025 +/- 0.0005 g/100 L at 5 L/min against 250 mm Hg. This is one-seventh of the equivalent figure for a Bio Pump BPX-80 (Medtronic, Inc., Minneapolis, MN, USA), which has a value of 0.0170 +/- 0.0096 g/100 L. These results show that the MedTech Dispo offers high pumping performance and low blood trauma.
引用
收藏
页码:531 / 540
页数:10
相关论文
共 50 条
  • [1] Disposable magnetically levitated centrifugal blood pump: Design and in vitro performance
    Hoshi, H
    Asama, J
    Shinshi, T
    Ohuchi, K
    Nakamura, M
    Mizuno, T
    Arai, H
    Shimokohbe, A
    Takatani, S
    [J]. ARTIFICIAL ORGANS, 2005, 29 (07) : 520 - 526
  • [2] Dynamic characteristics of a magnetically levitated impeller in a centrifugal blood pump
    Asama, Junichi
    Shinshi, Tadahiko
    Hoshi, Hideo
    Takatani, Setsuo
    Shimokohbe, Akira
    [J]. ARTIFICIAL ORGANS, 2007, 31 (04) : 301 - 311
  • [3] Magnetically suspended centrifugal blood pump with an axially levitated motor
    Masuzawa, T
    Ezoe, S
    Kato, T
    Okada, Y
    [J]. ARTIFICIAL ORGANS, 2003, 27 (07) : 631 - 638
  • [4] HeartQuest ventricular assist device magnetically levitated centrifugal blood pump
    Bearnson, GB
    Jacobs, GB
    Kirk, J
    Khanwilkar, PS
    Nelson, KE
    Long, JW
    [J]. ARTIFICIAL ORGANS, 2006, 30 (05) : 339 - 346
  • [5] Hydraulic Analysis and Optimization in impeller of Magnetically Levitated Centrifugal Blood Pump
    Zhang, Huimin
    Zhu, Qi
    [J]. FRONTIERS OF MANUFACTURING AND DESIGN SCIENCE II, PTS 1-6, 2012, 121-126 : 1204 - 1208
  • [6] A new design for a compact centrifugal blood pump with a magnetically levitated rotor
    Asama, J
    Shinshi, T
    Hoshi, H
    Takatani, S
    Shimokohbe, A
    [J]. ASAIO JOURNAL, 2004, 50 (06) : 550 - 556
  • [7] Prevention of thrombus formation in blood pump by mechanical circular orbital excitation of impeller in magnetically levitated centrifugal pump
    Hatakenaka, Kohei
    Hijikata, Wataru
    Fujiwara, Tatsuki
    Ohuchi, Katsuhiro
    Inoue, Yusuke
    [J]. ARTIFICIAL ORGANS, 2023, 47 (02) : 425 - 431
  • [8] A cost-effective extracorporeal magnetically-levitated centrifugal blood pump employing a disposable magnet-free impeller
    Hijikata, W.
    Mamiya, T.
    Shinshi, T.
    Takatani, S.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2011, 225 (H12) : 1149 - 1157
  • [9] Peripheral component malfunction of a fully magnetically levitated centrifugal pump masquerading as pump thrombosis
    Mastoris, Ioannis
    El Khashab, Mohamed
    Abicht, Travis
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2019, 38 (09): : 1000 - 1002
  • [10] A compact highly efficient and low hemolytic centrifugal blood pump with a magnetically levitated impeller
    Asama, J
    Shinshi, T
    Hoshi, H
    Takatani, S
    Shimokohbe, A
    [J]. ARTIFICIAL ORGANS, 2006, 30 (03) : 160 - 167