Positional dependence of the CT number with use of a cone-beam CT scanner for an electron density phantom in particle beam therapy

被引:3
|
作者
Kusano Y. [1 ]
Uesaka S. [2 ]
Yajima K. [2 ]
Kumagai M. [3 ]
Mizuno H. [3 ]
Mori S. [3 ]
机构
[1] Promotion Division of Heavy Ion Facilities, Kanagawa Cancer Center, 1-1-2 Nakao, Asahi-ku, Yokohama
[2] Therapy-Level Dose Calibration Center, Association for Nuclear Technology in Medicine, 3-9-19 Kurosunadai, Inage-Ku
[3] Department of Accelerator Physics and Medical Physics, Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, 4-9-1 Anagawa, Inage-ku
关键词
Cone-beam CT; CT number; Electron density; Heel effect; QA;
D O I
10.1007/s12194-012-0194-x
中图分类号
学科分类号
摘要
We evaluated the CT number accuracy in determining a CT calibration method for treatment planning by use of a 256 multi-slice CT (MSCT) scanner. An electron density phantom, which extends full length in the longitudinal direction, was scanned by the 256 MSCT scanner in a single rotation. We inserted four types of samples (air, 100 % ethanol, 40 wt% aqueous K2HPO 4, and water) into the phantom. The regions of interest were set for all image slices, and the average CT number was calculated in the transverse and longitudinal sections. For the transverse image direction, the CT number became lower toward the center of the phantom with almost all samples. The causes of these phenomena might be attributed to the effects of scattered radiation and those of beam hardening due to the high-atomic-number material (40 wt% aqueous K2HPO4). However, the phenomena were too complicated to allow us to determine their causes only from the present studies. Meanwhile, for increasing the accuracy of the CT number calibration, a single sample should be inserted, and this step should be repeated with different samples. For the longitudinal direction, the CT number for a 40 wt % aqueous K 2HPO4 sample increased by 30 HU from the anode to the cathode side due to variations in the X-ray energy caused by the heel effect. The caveat is that the CT number varies in the longitudinal direction when a CT scanner with a wide beam width is used. © 2012 Japanese Society of Radiological Technology and Japan Society of Medical Physics.
引用
收藏
页码:241 / 247
页数:6
相关论文
共 50 条
  • [21] Simulating low-dose cone-beam CT: a phantom study
    Ferrero, Andrea
    Fetterly, Ken
    Yu, Lifeng
    Schueler, Beth
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [22] On the Determination of the Beam Output of a Kilovoltage Cone-Beam CT
    Ding, G.
    Coffey, C.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [23] Density calibrated cone-beam CT as a tool for adaptive radiotherapy
    Nyeng, T. B.
    Holm, A. I.
    Moller, D. S.
    Assenholt, M. S.
    Hansen, R.
    Nyvang, L.
    Ravkilde, T.
    Thomsen, M. S.
    Hoffmann, L.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1292 - S1294
  • [24] Crescent artifacts in cone-beam CT
    Giles, William
    Bowsher, James
    Li, Hao
    Yin, Fang-Fang
    MEDICAL PHYSICS, 2011, 38 (04) : 2116 - 2121
  • [25] Breathing motion in cone-beam CT
    Rit, S.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S119 - S120
  • [26] Towards cone-beam CT thermometry
    Li, Ming
    Abi-Jaoudeh, Nadine
    Kapoor, Ankur
    Kadoury, Samuel
    Xu, Sheng
    Noordhoek, Niels
    Radaelli, Alessandro
    Carelsen, Bart
    Wood, Bradford J.
    MEDICAL IMAGING 2013: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2013, 8671
  • [27] Clinical implementation of a cone-beam CT
    Betgen, A
    van Herk, M
    Remeijer, P
    Sonke, JJ
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S158 - S158
  • [28] Flat Detector Cone-Beam CT
    Kachelriess, Marc
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2013, 23 (02): : 91 - 91
  • [29] Cone-beam CT experience in Aarau
    Cossmann, PH
    Stuessi, A
    von Briel, C
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S9 - S9
  • [30] Cone-beam CT for radiotherapy applications
    Cho, P.S.
    Johnson, R.H.
    Griffin, T.W.
    1863, Institute of Physics Publishing Ltd, Bristol, Engl (40):