Comparison of the McGrath MAC video laryngoscope and the Pentax Airwayscope during chest compression: a manikin study

被引:0
|
作者
Atsushi Kotera
Hiroki Irie
Shinsuke Iwashita
Junichi Taniguchi
Shunji Kasaoka
Yoshihiro Kinoshita
机构
[1] Kumamoto University Hospital,Department of Emergency and General Medicines
[2] Kumamoto University Hospital,Department of Intensive Care Unit
关键词
McGrath MAC; Pentax AirwayScope; Chest compression;
D O I
10.1186/2052-0492-2-18
中图分类号
学科分类号
摘要
We tested the utility of the McGrath MAC® (McG) video laryngoscope during chest compression compared with the Pentax Airwayscope® (AWS). We recruited 59 participants into the simulation study. The difference in the time to intubation (TTI [sec]) between without and with chest compression was significant for the AWS attempts (median 13, range 6–28 vs. median 15, range 6–72, p = 0.0247) but not significant for the McG attempts (median 16, range 6–75 vs. median 16, range 6–71); however, the difference of the TTIs is not serious clinically. The utility of the two devices during chest compressions is almost similar although their characteristics are different.
引用
收藏
相关论文
共 50 条
  • [41] A Randomized Comparison Simulating Face to Face Endotracheal Intubation of Pentax Airway Scope, C-MAC Video Laryngoscope, Glidescope Video Laryngoscope, and Macintosh Laryngoscope
    Choi, Hyun Young
    Oh, Young Min
    Kang, Gu Hyun
    Kang, Hyunggoo
    Jang, Yong Soo
    Kim, Wonhee
    Kim, Euichung
    Cho, Young Soon
    Choi, Hyukjoong
    Kim, Hyunjong
    Kim, Gyoung Yong
    BIOMED RESEARCH INTERNATIONAL, 2015, 2015
  • [42] Comparison of self-assembled video laryngoscope versus McGrath MAC®: A randomised controlled trial
    Sukmono, Besthadi
    Manggala, Sidharta K.
    Auerkari, Aino N.
    Christina, Budiani
    INDIAN JOURNAL OF ANAESTHESIA, 2022, 66 (05) : 350 - 357
  • [43] Comparison of Pentax-AWS® with Macintosh laryngoscope for tracheal intubation without interruption of chest compression
    Xue, F. S.
    Xiong, J.
    Wang, Q.
    Yuan, Y. J.
    Liao, X.
    RESUSCITATION, 2010, 81 (07) : 912 - 913
  • [44] Comparison of three types of laryngoscope for tracheal intubation during rhythmic chest compressions: A manikin study
    Koyama, J.
    Iwashita, T.
    Okamoto, K.
    RESUSCITATION, 2010, 81 (09) : 1172 - 1174
  • [45] Can The McGrath® MAC Video Laryngoscope be Used in MRI Zone 4?
    Chee, Won
    Peng, Qi
    ANESTHESIA AND ANALGESIA, 2017, 124 : 1159 - 1160
  • [46] A manikin study comparing McGrath Mac® and Airtraq® with Macintosh laryngoscope in tracheal intubation by intensive care unit doctors
    Chan, H. K. W.
    Wong, O. F.
    Kwan, G. W. M.
    HONG KONG JOURNAL OF EMERGENCY MEDICINE, 2015, 22 (06) : 337 - 344
  • [47] Randomized controlled trial comparing the McGrath MAC video laryngoscope with the King Vision video laryngoscope in adult patients
    Alvis, Bret D.
    Hester, Douglas
    Watson, Dusty
    Higgins, Michael
    St Jacques, Paul
    MINERVA ANESTESIOLOGICA, 2016, 82 (01) : 30 - 35
  • [48] Pentax-AWS is More Useful Than Macintosh Laryngoscope and Airtraq to Establish Tracheal Intubation while Uninterrupted Chest Compression: a Comparative Study with Manikin
    Koyama, Jun-ichl
    CIRCULATION, 2008, 118 (18) : S1481 - S1481
  • [49] A COMPARISON BETWEEN NOVICE AND TRAINED PERSONNEL WHEN USING THE MACINTOSH LARYNGOSCOPE, THE PENTAX AWS®, THE C-MAC™ AND THE BONFILS INTUBATION FIBERSCOPE: A MANIKIN STUDY
    Liaw, C.
    Lye, S.
    Seet, E.
    Koh, K.
    ANESTHESIA AND ANALGESIA, 2012, 114
  • [50] Comparison of results from novice and trained personnel using the Macintosh laryngoscope, Pentax AWS®, C-MAC™ and Bonfils intubation fibrescope: a manikin study
    Lye, Soh Teng
    Liaw, Chen Mei
    Seet, Edwin
    Koh, Kwong Fah
    SINGAPORE MEDICAL JOURNAL, 2013, 54 (02) : 64 - 68