Early migration characteristics of a hydroxyapatite-coated femoral stem: an RSA study

被引:0
|
作者
David Campbell
Graham Mercer
Kjell G. Nilsson
Vanessa Wells
John R. Field
Stuart A. Callary
机构
[1] Repatriation General Hospital,Orthopaedic Unit
[2] Umeå University Hospital,Department of Orthopaedics
[3] Flinders University,CORe: Comparative Orthopaedic Research Surgical Facility
[4] University of Adelaide,Discipline of Orthopaedics and Trauma
[5] Royal Adelaide Hospital,Department of Orthopaedics and Trauma
来源
关键词
Stem Rotation; Early Migration; Stem Subsidence; Cementless Stem; Radiostereometric Analysis;
D O I
暂无
中图分类号
学科分类号
摘要
Measurement of early stem subsidence can be used to predict the likelihood of long-term femoral component loosening and clinical failure. Data that examines the early migration pattern of clinically proven stems will provide clinicians with useful baseline data with which to compare new stem designs. This study was performed to evaluate the early migration pattern of a hydroxyapatite-coated press-fit femoral component that has been in use for over ten years. We enrolled 30 patients who underwent THA for osteoarthritis. The median age was 70 years (range, 55–80 years). Patients were clinically assessed using the Harris hip score. Radiostereometric analysis was used to evaluate stem migration at three to four days, six months, one year and two years. We observed a mean subsidence of 0.73 mm at six months, 0.62 mm at one year and 0.58 mm at two years and a mean retroversion of 1.82° at six months, 1.90° at one year and 1.59° at two years. This data suggests that subsidence is confined to the first six months after which there was no further subsidence. The results from this study can be compared with those from novel cementless stem designs to help predict the long-term outcome one may expect from new cementless stem designs.
引用
收藏
页码:483 / 488
页数:5
相关论文
共 50 条
  • [41] Hydroxyapatite-coated versus grit-blasted femoral stems
    Amstutz, HC
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (08): : 1206 - 1207
  • [42] Tissue response to the components of a hydroxyapatite-coated composite femoral implant
    Hacking, S. A.
    Pauyo, T.
    Lim, L.
    Legoux, J. G.
    Bureau, M. N.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2010, 94A (03) : 953 - 960
  • [43] Femoral revision surgery using a fully hydroxyapatite-coated stem: a cohort study of twenty two to twenty seven years
    Reikeras, Olav
    INTERNATIONAL ORTHOPAEDICS, 2017, 41 (02) : 271 - 275
  • [44] HYDROXYAPATITE-COATED PROXIMAL INGROWTH FEMORAL STEMS - A MATCHED-PAIR CONTROL STUDY
    MCPHERSON, EJ
    DORR, LD
    GRUEN, TA
    SABERI, MT
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1995, (315) : 223 - 230
  • [45] X-RAY-DIFFRACTION OF NEWLY FORMED BONE CLOSE TO ALUMINA-COATED OR HYDROXYAPATITE-COATED FEMORAL STEM
    SAVARINO, L
    CENNI, E
    STEA, S
    DONATI, ME
    PAGANETTO, G
    MORONI, A
    TONI, A
    PIZZOFERRATO, A
    BIOMATERIALS, 1993, 14 (12) : 900 - 905
  • [46] Femoral Component Revision Using an Extensively Hydroxyapatite-Coated Stem (vol 19, pg 8, 2004)
    Crawford, Charles H., III
    JOURNAL OF ARTHROPLASTY, 2009, 24 (04): : 659 - 659
  • [47] Fracture of fully hydroxyapatite-coated titanium femoral stem of a total hip replacement - a report of 3 cases
    Sharma, DK
    Brooks, S
    Creasey, S
    Lewis, B
    ACTA ORTHOPAEDICA SCANDINAVICA, 2004, 75 (06): : 768 - 771
  • [48] Early failure of hemispheric hydroxyapatite-coated acetabular cups
    Kim, Shin-Yoon
    Kim, Do-Heon
    Kim, Yong-Goo
    Oh, Chang-Wug
    Ihn, Joo-Chul
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (446) : 233 - 238
  • [49] Surface characteristics of a novel hydroxyapatite-coated dental implant
    Jung, Ui-Won
    Hwang, Ji-Wan
    Choi, Da-Yae
    Hu, Kyung-Seok
    Kwon, Mi-Kyung
    Choi, Seong-Ho
    Kim, Hee-Jin
    JOURNAL OF PERIODONTAL AND IMPLANT SCIENCE, 2012, 42 (02): : 59 - 63
  • [50] Hydroxyapatite-coated femoral stems - A matched-pair analysis of coated and uncoated implants
    Rothman, RH
    Hozack, WJ
    Ranawat, A
    Moriarty, L
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1996, 78A (03): : 319 - 324