Localization of the ventricular pacing site from BSPM and standard 12-lead ECG: a comparison study

被引:0
|
作者
Ksenia A. Sedova
Peter M. van Dam
Marie Blahova
Lucie Necasova
Josef Kautzner
机构
[1] Czech Technical University in Prague,Department of Biomedical Technology, Faculty of Biomedical Engineering
[2] University Medical Center Utrecht,Department of Cardiology
[3] Institute for Clinical and Experimental Medicine,Department of Cardiology
来源
关键词
D O I
暂无
中图分类号
学科分类号
摘要
Inverse ECG imaging methods typically require 32–250 leads to create body surface potential maps (BSPM), limiting their routine clinical use. This study evaluated the accuracy of PaceView inverse ECG method to localize the left or right ventricular (LV and RV, respectively) pacing leads using either a 99-lead BSPM or the 12-lead ECG. A 99-lead BSPM was recorded in patients with cardiac resynchronization therapy (CRT) during sinus rhythm and sequential LV/RV pacing. The non-contrast CT was performed to localize precisely both ECG electrodes and CRT leads. From a BSPM, nine signals were selected to obtain the 12-lead ECG. Both BSPM and 12-lead ECG were used to localize the RV and LV lead, and the localization error was calculated. Consecutive patients with dilated cardiomyopathy, previously implanted with a CRT device, were enrolled (n = 19). The localization error for the RV/LV lead was 9.0 [IQR 4.8–13.6] / 7.7 [IQR 0.0–10.3] mm using the 12-lead ECG and 9.1 [IQR 5.4–15.7] / 9.8 [IQR 8.6–13.1] mm for the BSPM. Thus, the noninvasive lead localization using the 12-lead ECG was accurate enough and comparable to 99-lead BSPM, potentially increasing the capability of 12-lead ECG for the optimization of the LV/RV pacing sites during CRT implant or for the most favorable programming.
引用
收藏
相关论文
共 50 条
  • [21] Simulating Arbitrary Electrode Reversals in Standard 12-Lead ECG
    Krasteva, Vessela
    Jekova, Irena
    Schmid, Ramun
    [J]. SENSORS, 2019, 19 (13)
  • [22] Map representation and diagnostic performance of the standard 12-lead ECG
    Kornreich, F
    Lux, RL
    MacLeod, RS
    [J]. JOURNAL OF ELECTROCARDIOLOGY, 1995, 28 : 121 - 123
  • [23] USE OF THE 12-LEAD ECG IN PREDICTING VENTRICULAR-TACHYCARDIA SITE OF ORIGIN - SURGICAL IMPLICATIONS
    MILLER, JM
    CASSIDY, DM
    VASSALLO, JA
    MARCHLINSKI, FE
    HARGROVE, WC
    JOSEPHSON, ME
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1986, 7 (02) : A141 - A141
  • [24] Evaluation of the Accuracy of ECG Captured by CardioChip through Comparison of Lead I Recording to a Standard 12-Lead ECG Recording Device
    Lo, Chi-In
    Chang, Sheng-Shiung
    Tsai, Jui-Peng
    Kuo, Jen-Yuan
    Chen, Ying-Ju
    Huang, Ming-Yuan
    Lee, Chao-Hsiung
    Sung, Kuo-Tzu
    Hung, Chung-Lieh
    Hou, Charles Jia-Yin
    Lai, Edward
    Yeh, Hung-I
    Chang, Wen-Ling
    Chang, Wen-Han
    [J]. ACTA CARDIOLOGICA SINICA, 2018, 34 (02) : 144 - 151
  • [25] Comparison of electrocardiograms (ECG) waveforms and centralized ECG measurements between a simple 6-lead mobile ECG device and a standard 12-lead ECG
    Kleiman, Robert
    Darpo, Borje
    Brown, Randy
    Rudo, Todd
    Chamoun, Svetlana
    Albert, David E.
    Bos, Johan Martijn
    Ackerman, Michael J.
    [J]. ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2021, 26 (06)
  • [26] Study on the use of standard 12-lead ECG data for rhythm-type ECG classification problems
    Park, Junsang
    An, Junho
    Kim, Jinkook
    Jung, Sunghoon
    Gil, Yeongjoon
    Jang, Yoojin
    Lee, Kwanglo
    Oh, Il-young
    [J]. COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2022, 214
  • [27] Comparison between a 6-lead smartphone ECG and 12-lead ECG in athletes
    Orchard, Jessica J.
    Orchard, John W.
    Raju, Hariharan
    La Gerche, Andre
    Puranik, Rajesh
    Semsarian, Chris
    [J]. JOURNAL OF ELECTROCARDIOLOGY, 2021, 66 : 95 - 97
  • [28] PREDICTION OF LEFT-VENTRICULAR DYSSYNERGY FROM STANDARD 12-LEAD ECG AFTER ACUTE MYOCARDIAL-INFARCTION
    NYGART, J
    GRANDE, P
    SAUNAMAKI, K
    HINDMAN, N
    WAGNER, G
    [J]. ACTA MEDICA SCANDINAVICA, 1986, : 121 - 121
  • [29] Feasibility of His bundle pacing facilitated by EASI derived 12-lead ECG
    Langanke, Alexander
    Andreas, Klaus
    [J]. JOURNAL OF ELECTROCARDIOLOGY, 2023, 81 : 272 - 276
  • [30] Comparison of a cardiac mapping device with standard 12-lead ECG in the diagnosis of acute coronary syndrome
    Fermann, GJ
    Lindsell, CJ
    Joseph, KJ
    Gibler, WB
    [J]. ANNALS OF EMERGENCY MEDICINE, 2004, 44 (04) : S73 - S73