Feasibility of His bundle pacing facilitated by EASI derived 12-lead ECG

被引:0
|
作者
Langanke, Alexander [1 ]
Andreas, Klaus [1 ]
机构
[1] Med Klin 1, Varisano Klinikum Bad Soden, Kronberger Str 36, D-65812 Bad Soden, Germany
关键词
Electrocardiogram; Vector-ECG; His bundle pacing; Conduction system pacing; EASI; ELECTROCARDIOGRAM; IMPLANTATION; STANDARD; SOCIETY; SYSTEM;
D O I
10.1016/j.jelectrocard.2023.10.010
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: His bundle pacing (HBP) has become popular in recent years as a more physiological alternative to conventional right ventricular pacing. Implantation requires 12-lead ECG during surgery, which is not readily available in a standard operating room. Often but not always HBP is performed in an electrophysiology lab. EASI is a reduced lead system which enables derived 12-lead ECG. EASI derived 12-lead ECGs on modern tablet computers offer a more mobile and lightweight ECG solution which does not obstruct fluoroscopy during implantation. This case series aims to compare standard 12-lead ECG to EASI derived 12-lead ECG in patients undergoing HBP implantation. Methods and results: A total of 11 patients received permanent HBP guided only by fluoroscopy, a pacing system analyzer (Medtronic CareLink SmartSync Device Manager) and EASI derived 12-lead ECG (CardioSecur Pro). During the first postoperative device interrogation HBP criteria, as defined in the EHRA consensus paper on conduction system pacing, were evaluated with the EASI derived system as well as a standard 12-lead ECG and compared to each other. There was perfect agreement with regards to these criteria which lead to identical conclusions in all cases. Conclusion: HBP implantation can be performed with EASI derived 12-lead ECG instead of conventional 12-lead ECG. Criteria for discriminating between selective His bundle, non-selective His bundle or myocardial capture alone are clearly visible in the EASI derived ECG leading to the same conclusion when compared to standard 12-lead ECG. Compared to a conventional 12-lead ECG the EASI system offers a leaner setup with less visual obstruction on fluoroscopy.
引用
收藏
页码:272 / 276
页数:5
相关论文
共 50 条
  • [1] Accuracy of QT measurement in the EASI-derived 12-lead ECG
    Martinez, Juan Pablo
    Laguna, Pablo
    Olmos, Salvador
    Pahlm, Olle
    Petersson, Jonas
    Sornmo, Leif
    [J]. 2006 28TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-15, 2006, : 4192 - +
  • [2] Sensitivity to noise and artifacts of the Mason-Likar 12-lead ECG and the EASI-derived 12-lead ECG for monitoring applications
    Thulin, A
    Pettersson, J
    Pahlm, O
    [J]. CRITICAL CARE MEDICINE, 2002, 30 (12) : A76 - A76
  • [3] A new 5 electrode 12-lead ECG (EASI) is equivalent to standard 12-lead ECG for myocardial infarction
    Wehr, G
    Peters, R
    Khalifé, K
    Banning, A
    Kuehlkamp, V
    Rickards, A
    Duller, B
    Sechtem, U
    [J]. INTENSIVE CARE MEDICINE, 2002, 28 : S78 - S78
  • [4] Investigation Of A Transfer Function Between Standard 12-Lead ECG And EASI ECG
    Oleksy, Wojciech
    Tkacz, Ewaryst
    [J]. ANALYSIS OF BIOMEDICAL SIGNALS AND IMAGES, 2010, : 322 - 327
  • [5] Derived 12-lead ECG systems
    Nelwan, SP
    Meij, SH
    [J]. JOURNAL OF ELECTROCARDIOLOGY, 2006, 39 (01) : 29 - 30
  • [6] A vector-based 5 electrode 12-lead ECG (EASI®) is equivalent to the conventional 12-lead ECG for diagnosis of myocardial ischemia
    Wehr, G
    Peters, R
    Khalife, K
    Banning, A
    Kuehlkamp, V
    Rickards, A
    Adams, R
    Sechtem, U
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (05) : 122A - 122A
  • [7] Diagnostic conclusions from the EASI-derived 12-lead electrocardiogram as compared with the standard 12-lead electrocardiogram in children
    Welinder, A
    Feild, DQ
    Liebman, J
    Maynard, C
    Wagner, GS
    Wettrell, G
    Pahlm, O
    [J]. AMERICAN HEART JOURNAL, 2006, 151 (05) : 1059 - 1064
  • [8] A vector-based, 5-electrode, 12-lead monitoring ECG (EASI) is equivalent to conventional 12-lead ECG for diagnosis of acute coronary syndromes
    Wehr, G
    Peters, RJ
    Khalifé, K
    Banning, AP
    Kuehlkamp, V
    Rickards, AF
    Sechtem, U
    [J]. JOURNAL OF ELECTROCARDIOLOGY, 2006, 39 (01) : 22 - 28
  • [9] EASI™ 12-lead ECG with a handheld computer refines cardiovascular diagnosis in general practice
    Pollet, Anne-Gaelle
    Guenancia, Charles
    Garcia, Rodrigue
    Viart, Guillaume
    Dubois, Daphnee
    Bourgois, Marc-antoine
    Chapelet, Francois
    Loyez, Thibault
    Dautriche, Benoit
    Guyomar, Yves
    Graux, Pierre
    Marechaux, Sylvestre
    Menet, Aymeric
    [J]. JOURNAL OF ELECTROCARDIOLOGY, 2022, 73 : 96 - 102
  • [10] Atrial tachycardias arising from the region of the his bundle: 12-lead ecg and intracardiac electrogram characteristics
    Grosse, A.
    Raffa, S.
    Brunelli, M.
    Wauters, K.
    Geller, C.
    [J]. EUROPEAN HEART JOURNAL, 2009, 30 : 701 - 701