In Vitro and In Vivo Performance Evaluation of the Second Developmental Version of the PediaFlow Pediatric Ventricular Assist Device

被引:0
|
作者
Maul T.M. [1 ,2 ,3 ]
Kocyildirim E. [2 ]
Johnson Jr. C.A. [1 ,2 ]
Daly A.R. [1 ,2 ]
Olia S.E. [1 ,2 ]
Woolley J.R. [1 ,2 ]
Snyder S. [5 ]
Bengston S.G. [2 ]
Kameneva M.V. [1 ,2 ]
Antaki J.F. [2 ,4 ]
Wagner W.R. [1 ,2 ]
Borovetz H.S. [1 ,2 ]
Wearden P.D. [2 ,3 ,6 ]
机构
[1] Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA
[2] McGowan Institute for Regenerative Medicine, University of Pittsburgh, Pittsburgh, PA
[3] Department of Cardiothoracic Surgery, Children's Hospital of Pittsburgh of UPMC, Pittsburgh, PA
[4] Department of Bioengineering, Carnegie Mellon University, Pittsburgh, PA
[5] LaunchPoint Technologies, Inc., Goleta, CA
[6] Children's Hospital of Pittsburgh, Pittsburgh, PA 15201, Faculty Pavilion 05-143
基金
美国国家卫生研究院; 美国国家科学基金会;
关键词
In vivo; Mechanical circulatory support; Pediatric; Pre-clinical study; Ventricular assist device;
D O I
10.1007/s13239-011-0061-7
中图分类号
学科分类号
摘要
Ventricular assist devices (VADs) have significantly impacted the treatment of adult cardiac failure, but few options exist for pediatric patients. This has motivated our group to develop an implantable magnetically levitated rotodynamic VAD (PediaFlow®) for 3-20 kg patients. The second prototype design of the PediaFlow (PF2) is 56% smaller than earlier prototypes, and achieves 0.5-1.5 L/min blood flow rates. In vitro hemodynamic performance and hemolysis testing were performed with analog blood and whole ovine blood, respectively. In vivo evaluation was performed in an ovine model to evaluate hemocompatibility and end-organ function. The in vitro normalized index of hemolysis was 0.05-0.14 g/L over the specified operating range. In vivo performance was satisfactory for two of the three implanted animals. A mechanical defect caused early termination at 17 days of the first in vivo study, but two subsequent implants proceeded without complication and electively terminated at 30 and 70 days. Serum chemistries and plasma free hemoglobin were within normal limits. Gross necropsy revealed small, subclinical infarctions in the kidneys of the 30 and 70 day animals (confirmed by histopathology). The results of these experiments, particularly the biocompatibility demonstrated in vivo encourage further development of a miniature magnetically levitated VAD for the pediatric population. Ongoing work including further reduction of size will lead to a design freeze in preparation for of clinical trials. © 2011 Biomedical Engineering Society.
引用
收藏
页码:253 / 262
页数:9
相关论文
共 50 条
  • [21] In Vivo Assessment of Thrombogenicity of the Penn State Pediatric Ventricular Assist Device
    Weiss, W. J.
    Carney, E. L.
    Clark, J. B.
    Cooper, T.
    Peterson, R.
    Nifong, T. P.
    Siedlecli, C.
    Lukic, B.
    Rosenberg, G.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2010, 29 (02): : S120 - S121
  • [22] Hemodynamic performance in-vivo of a new ventricular assist device
    Corona, F.
    Barragan, R.
    Calderon, M.
    Martinez, H.
    Infante, O.
    Molina, J.
    Lespron, C.
    Hernandez, A.
    Escobedo, C.
    Catrip, J.
    Gorzelewski, A.
    Tena, C.
    Graullera, V.
    Sacristan, E.
    2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 394 - 397
  • [23] Initial In Vivo Evaluation of a Novel Left Ventricular Assist Device
    Wu, Guanghui
    Lin, Changyan
    Li, Haiyang
    Hou, Xiaotong
    Chen, Chen
    Liu, Xiujian
    Xu, Chuangye
    Wang, Jing
    Yang, Peng
    Qu, Wenbo
    BIOMED RESEARCH INTERNATIONAL, 2015, 2015
  • [24] In vivo evaluation of skin integration with ventricular assist device drivelines
    Cavalcanti, Amanda S.
    Diaz, Raquel Sanchez
    Bolle, Eleonore C. L.
    Bartnikowski, Nicole
    Fraser, John F.
    McGiffin, David
    Savi, Flavia Medeiros
    Shafiee, Abbas
    Dargaville, Tim R.
    Gregory, Shaun D.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2022, 41 (08): : 1032 - 1043
  • [25] In vivo evaluation of investigation the HeartWare centrifugal ventricular assist device
    Tuzun, Egemen
    Roberts, Katy
    Cohn, William E.
    Sargin, Murat
    Gemmato, Courtney J.
    Radovancevic, Branislav
    Frazier, O. H.
    TEXAS HEART INSTITUTE JOURNAL, 2007, 34 (04) : 406 - 411
  • [26] Initial in vivo evaluation of the DexAide right ventricular assist device
    Ootaki, Y
    Kamohara, K
    Akiyama, M
    Zahr, F
    Kopcak, MW
    Dessoffy, R
    Massiello, A
    Horvath, D
    Chen, JF
    Benefit, S
    Golding, LAR
    Fukamachi, K
    ASAIO JOURNAL, 2005, 51 (06) : 739 - 742
  • [27] In Vitro and in Vivo Hemolysis Tests of a Maglev Implantable Ventricular Assist Device
    Cai, Keqiang
    Pan, Lianqiang
    Liu, Yuqian
    Wu, Guanghui
    Lin, Changyan
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 3, 2019, 68 (03): : 131 - 134
  • [28] Device Thrombogenicity Emulation: An In Silico Predictor of In Vitro and In Vivo Ventricular Assist Device Thrombogenicity
    Wei Che Chiu
    Phat L. Tran
    Zain Khalpey
    Eric Lee
    Yi-Ren Woo
    Marvin J. Slepian
    Danny Bluestein
    Scientific Reports, 9
  • [29] Device Thrombogenicity Emulation: An In Silico Predictor of In Vitro and In Vivo Ventricular Assist Device Thrombogenicity
    Chiu, Wei Che
    Tran, Phat L.
    Khalpey, Zain
    Lee, Eric
    Woo, Yi-Ren
    Slepian, Marvin J.
    Bluestein, Danny
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [30] Ventricular assist device in pediatric patients
    Gravino, R. Rita
    Limongelli, G.
    Petraio, A.
    Verrengia, M.
    Rea, A.
    Salerno, G.
    Masarone, D.
    Maiello, C.
    Russo, M. G.
    Pacileo, G.
    EUROPEAN JOURNAL OF HEART FAILURE, 2016, 18 : 349 - 350