In Vitro and in Vivo Hemolysis Tests of a Maglev Implantable Ventricular Assist Device

被引:0
|
作者
Cai, Keqiang [1 ]
Pan, Lianqiang [1 ]
Liu, Yuqian [1 ]
Wu, Guanghui [1 ,2 ]
Lin, Changyan [1 ,2 ]
机构
[1] Capital Med Univ, Beijing Anzhen Hosp, Beijing 100029, Peoples R China
[2] Beijing Inst Heard Lung & Blood Vessel Dis, Beijing 100029, Peoples R China
基金
美国国家科学基金会;
关键词
CH-VAD; Maglev blood pump; Hemolysis performance; In vivo test; PUMP;
D O I
10.1007/978-981-10-9023-3_23
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: Implantation of a ventricular assist device (VAD) is a seminal therapeutic option for patients with terminal cardiac failure. A growing number of VAD patients are successfully bridged to transplantation, or can even live permanently with the device. However, the success is restricted by frequent severe complications. Haemolysis is a relevant adverse effect of several VAD types, which is the result of destruction of red blood cells, reduced by wall shear stress, flow acceleration and interaction with artificial surfaces. The CH-VAD, a small implantable continuous-flow blood pump, featuring a magnetically levitated impeller and enough hydrodynamic performance, was under development and completed for a 60-days animal implantation experiment in 6 sheep. The goal of this study is to validate the hemolysis of the pump through in vitro and in vivo studies. Methods: A series of in vitro tests was quantified experimentally by using in vitro circulation loop system according to ASTM F1841, the standard practice for the assessment of hemolysis in continuous-flow blood pumps. The hemolysis test in vivo was performed during a 60-days ovine model implantation, which was being conducted under the Institutional Animal Care and Use Committee (IACUC) protocol 05-0600 1. Results in vitro tests showed that the average normalized index of hemolysis (NIH) value of the VAD was 0.007 mg/l. The hemolysis in vivo was evaluated based on the amount of free hemoglobin in the plasma, and which showed that the free hemoglobin level in plasma peaked at 0.95 mg/l on the fifth postoperative day and then returned to an acceptable range of 6.0 mg/dL. Conclusion The magnetic levitation left ventricular assist device has good hemolytic performance. These acceptable performance results supported proceeding initial clinical trail conditions.
引用
收藏
页码:131 / 134
页数:4
相关论文
共 50 条
  • [1] HYDRAULICS AND IN VITRO HEMOLYSIS TEST OF A MAGLEV IMPLANTABLE VENTRICULAR ASSIST DEVICE
    Wu, Guanghui
    Xu, Chuangye
    Liu, Xiujian
    Lin, Changyan
    Yang, Lin
    Li, Haiyang
    Hou, Xiaotong
    Chen, Chen
    Yang, Peng
    [J]. JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2017, 17 (07)
  • [2] In vitro and in vivo testing of an implantable motor-driven left ventricular assist device
    Murakami, T
    Kikugawa, D
    Fukuhiro, Y
    Kanazawa, S
    Fujiwara, T
    Katsumura, T
    Kukunaga, S
    Matsuura, Y
    [J]. ARTIFICIAL ORGANS, 1996, 20 (02) : 152 - 155
  • [3] In vivo evaluation of an implantable magnetic suspending left ventricular assist device
    Lin, Changyan
    Wu, Guanghui
    Liu, Xiujian
    Xu, Chuangye
    Hou, Xiaotong
    Li, Haiyang
    Chen, Chen
    Yang, Peng
    Wang, Jing
    Liu, Yuyang
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2015, 38 (03): : 138 - 145
  • [4] 30-day in vivo study of a fully maglev extracorporeal ventricular assist device
    Li, Ping
    Wu, Tingting
    Hsu, Po-Lin
    Wei, Xufeng
    Dong, Nianguo
    [J]. ARTIFICIAL ORGANS, 2022, 46 (11) : 2171 - 2178
  • [5] The HeartMate III: Design and in vivo studies of a Maglev centrifugal left ventricular assist device
    Loree, HM
    Bourque, K
    Gernes, DB
    Richardson, JS
    Poirier, VL
    Barletta, N
    Fleischli, A
    Foiera, G
    Gempp, TM
    Schoeb, R
    Litwak, KN
    Akimoto, T
    Kameneva, M
    Watach, MJ
    Litwak, P
    [J]. ARTIFICIAL ORGANS, 2001, 25 (05) : 386 - 391
  • [6] THE IMPLANTABLE BIVENTRICULAR DLR ASSIST DEVICE: FIRST IN VITRO AND IN VIVO RESULTS
    Schmid, T.
    Schiller, W.
    Vodermayer, B.
    Liepsch, D.
    Welz, A.
    Hirzinger, G.
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2009, 32 (07): : 425 - 425
  • [7] Circulation System Modelling with National Implantable Ventricular Assist Device (NIVAD): Hydrodynamic Characters, Hemolysis Test in Vitro and Animal Experiments
    Li, Tieyan
    Ye, Liang
    Lu, Rong
    Fan, Huimin
    Liu, Zhongmin
    [J]. LIFE SYSTEM MODELING AND SIMULATION, 2014, 461 : 234 - 245
  • [8] In vivo evaluation of the biocompatibility of the totally implantable ventricular assist device (HeartSaver VAD)
    Mussivand, T
    Harasaki, H
    Litwak, K
    Slaughter, MS
    Gray, LA
    Dowling, RD
    Mueller, R
    Masters, RG
    Hendry, PJ
    Beck-Da-Silva, L
    Davies, R
    Haddad, H
    Mesana, TG
    Keon, WJ
    [J]. ASAIO JOURNAL, 2003, 49 (04) : 459 - 462
  • [9] In vivo evaluation of the miniaturized Gyro centrifugal pump as an implantable ventricular assist device
    Takami, Y
    Otsuka, G
    Mueller, J
    Sugita, Y
    Nakata, K
    Tayama, E
    Ohashi, Y
    Schima, H
    Schmallegger, H
    Wolner, E
    Nose, Y
    [J]. ARTIFICIAL ORGANS, 1998, 22 (08) : 713 - 720
  • [10] Design, development, and first in vivo results of an implantable ventricular assist device, MicroVad
    Kerkhoffs, W
    Schumacher, O
    Meyns, B
    Verbeken, E
    Leunens, V
    Bollen, H
    Reul, H
    [J]. ARTIFICIAL ORGANS, 2004, 28 (10) : 904 - 910