Determination of Low Wear Rates in Metal-On-Metal Hip Joint Replacements Based on Ultra Trace Element Analysis in Simulator Studies

被引:0
|
作者
J. P. Kretzer
M. Krachler
J. Reinders
E. Jakubowitz
M. Thomsen
C. Heisel
机构
[1] University of Heidelberg,Laboratory of Biomechanics and Implant Research, Department of Orthopaedics
[2] University of Heidelberg,Institute of Earth Sciences
[3] German Red Cross Hospital,undefined
[4] ARCUS Sportklinik,undefined
来源
Tribology Letters | 2010年 / 37卷
关键词
Chemical analytical techniques; Human joints; Replacements; Biotribology; Wear/failure testing devices;
D O I
暂无
中图分类号
学科分类号
摘要
Metal-on-metal joint bearings are being increasingly used for total hip arthroplasty. Previous results from simulator tests to determine wear of metal-on-metal bearings exhibited high fluctuations. Consequently, wear tests had to be performed for a long period of time to achieve stable values. The aim of this study was to establish a method for the precise measurement of wear of metal-on-metal bearings. Wear was determined by analyzing the concentrations of particles and selected elements in the test medium (serum) using high resolution-inductively coupled plasma-mass spectrometry (HR-ICP-MS). The procedure was first validated and compared to gravimetric measurements on two different implant designs and then applied in wear tests on four total hip resurfacing implants. The validation showed a significant reliability of the HR-ICP-MS method (p ≤ 0.02). The final wear test revealed that the HR-ICP-MS method can precisely detect very low wear rates and accurately characterize alterations in wear progression. The duration of simulation can be reduced due to the high detection power and low fluctuation in HR-ICP-MS wear determination analyses. This approach is suitable for test implants with extremely low wear rates.
引用
收藏
页码:23 / 29
页数:6
相关论文
共 50 条
  • [21] Simulator wear of some commercially available metal-on-metal hip implants
    Medley, John B.
    Dowling, Judith M.
    Poggie, Robert A.
    Krygier, Jan J.
    Bobyn, J. Dennis
    ASTM Special Technical Publication, 1999, (1346): : 92 - 110
  • [22] Simulator wear of some commercially available metal-on-metal hip implants
    Medley, JB
    Dowling, JM
    Poggie, RA
    Krygier, JJ
    Bobyn, JD
    ALTERNATIVE BEARING SURFACES IN TOTAL JOINT REPLACEMENT, 1998, 1346 : 92 - 110
  • [23] On the role of the lambda parameter in simulator wear of metal-on-metal hip implants
    Paré, P
    Medley, JB
    Chan, FW
    Young, SK
    TRIBOLOGICAL RESEARCH AND DESIGN FOR ENGINEERING SYSTEMS, 2003, : 281 - 290
  • [24] Understanding the differences between the wear of metal-on-metal and ceramic-on-metal total hip replacements
    Figueiredo-Pina, C. G.
    Yan, Y.
    Neville, A.
    Fisher, J.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2008, 222 (H3) : 285 - 296
  • [25] A predictive analysis of long-term friction and wear characteristics of metal-on-metal total hip replacements
    Dowson, D.
    Wang, F. C.
    Wang, W. Z.
    Jin, Z. M.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART J-JOURNAL OF ENGINEERING TRIBOLOGY, 2007, 221 (J3) : 460 - 460
  • [26] Biotribological study of multilayer coated metal-on-metal hip prostheses in a hip joint simulator
    Ortega-Saenz, J. A.
    Alvarez-Vera, M.
    Hernandez-Rodriguez, M. A. L.
    WEAR, 2013, 301 (1-2) : 234 - 242
  • [27] A Meta-Analysis of Design- and Manufacturing-Related Parameters Influencing the Wear Behavior of Metal-on-Metal Hip Joint Replacements
    Kretzer, Jan Philippe
    Kleinhans, Jennifer A.
    Jakubowitz, Eike
    Thomsen, Marc
    Heisel, Christian
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2009, 27 (11) : 1473 - 1480
  • [28] Femoral head diameter and carbon composition effect on wear of metal-on-metal hip replacements
    Ihaddadene, R.
    Affatato, S.
    Zavalloni, M.
    Bouzid, S.
    Viceconti, M.
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2011, 14 : 31 - 32
  • [29] Risk of cancer with metal-on-metal hip replacements: population based study
    Makela, Keijo T.
    Visuri, Tuomo
    Pulkkinen, Pekka
    Eskelinen, Antti
    Remes, Ville
    Virolainen, Petri
    Junnila, Mika
    Pukkala, Eero
    BMJ-BRITISH MEDICAL JOURNAL, 2012, 345
  • [30] Surface Engineering: A Low Wearing Solution for Metal-On-Metal Hip Surface Replacements
    Leslie, Ian J.
    Williams, Sophie
    Brown, Chris
    Anderson, James
    Isaac, Graham
    Hatto, Peter
    Ingham, Eileen
    Fisher, John
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2009, 90B (02) : 558 - 565