Risk of cancer with metal-on-metal hip replacements: population based study

被引:45
|
作者
Makela, Keijo T. [1 ]
Visuri, Tuomo [2 ]
Pulkkinen, Pekka [2 ]
Eskelinen, Antti [3 ]
Remes, Ville [4 ]
Virolainen, Petri [1 ]
Junnila, Mika [1 ]
Pukkala, Eero [5 ,6 ]
机构
[1] Turku Univ Hosp, Surg Hosp, Dept Orthopaed & Traumatol, FI-20701 Turku, Finland
[2] Univ Helsinki, Hjelt Inst, FI-00014 Helsinki, Finland
[3] Coxa Hosp Joint Replacement, FI-33101 Tampere, Finland
[4] Univ Helsinki, Cent Hosp, Peijas Hosp, Dept Orthopaed & Traumatol, FI-00029 Helsinki, Finland
[5] Univ Tampere, Sch Hlth Sci, FI-33100 Tampere, Finland
[6] Finnish Canc Registry, Inst Stat & Epidemiol Canc Res, FI-00130 Helsinki, Finland
来源
关键词
KNEE ARTHROPLASTY; WEAR DEBRIS; FOLLOW-UP; OSTEOARTHRITIS; DISSEMINATION; POLYETHYLENE; NATIONWIDE; FAILURE; COHORT;
D O I
10.1136/bmj.e4646
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective To assess the risk of cancer associated with modern primary metal-on-metal hip replacements. Design Population based study. Setting Nationwide retrospective comparative register. Participants 10 728 patients who underwent metal-on-metal total hip arthroplasty and 18 235 patients who underwent conventional metal-on-polyethylene, ceramic-on-polyethylene, and ceramic-on-ceramic total hip arthroplasty (the non-metal-on-metal cohort) in the Finnish Arthroplasty Register 2001-10. Data on cancer cases up to 2010 for these cohorts were extracted from the Finnish Cancer Registry. Main outcome measures The relative risk of cancer was expressed as the ratio of observed to expected number of cases from the Finnish population-that is, the standardised incidence ratio. The relative risk of cancer in the metal-on-metal cohort compared with the non-metal-on-metal cohort was estimated with analyses of these ratios and Poisson regression. Results The overall risk of cancer in patients with metal-on-metal hip implants was similar to that in the Finnish population (378 observed v 400 expected, standardised incidence ratio 0.95, 95% confidence interval 0.85 to 1.04). The overall risk of cancer in patients with metal-on-metal hip implants was also no higher than in patients who had received non-metal-on-metal hip implants (relative risk 0.92, 0.81 to 1.05). Conclusions Metal-on-metal hip replacements are not associated with an increased overall risk of cancer during a mean follow-up of four years.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Midterm risk of cancer with metal-on-metal hip replacements not increased in a Finnish population
    Ekman, Elina
    Laaksonen, Inari
    Eskelinen, Antti
    Pulkkinen, Pekka
    Pukkala, Eero
    Makela, Keijo
    ACTA ORTHOPAEDICA, 2018, 89 (05) : 575 - 579
  • [2] CoCrMo metal-on-metal hip replacements
    Liao, Yifeng
    Hoffman, Emily
    Wimmer, Markus
    Fischer, Alfons
    Jacobs, Joshua
    Marks, Laurence
    PHYSICAL CHEMISTRY CHEMICAL PHYSICS, 2013, 15 (03) : 746 - 756
  • [4] Total hip replacements with metal-on-metal bearings
    Milosev, Ingrid
    Kovac, Simon
    Topolovec, Matevz
    Pisot, Venceslav
    Trebse, Rihard
    ZDRAVNISKI VESTNIK-SLOVENIAN MEDICAL JOURNAL, 2009, 78 : 34 - 40
  • [5] Biological effects of metal-on-metal hip replacements
    Bhamra, MS
    Case, CP
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) : 379 - 384
  • [6] The tribology of metal-on-metal total hip replacements
    Scholes, SC
    Unsworth, A
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) : 183 - 194
  • [7] Failure rates of metal-on-metal hip replacements
    Kelsall, Diane
    CANADIAN MEDICAL ASSOCIATION JOURNAL, 2012, 184 (07) : 790 - 790
  • [8] Metal-on-metal hip implants and the risk of cancer
    Rosengren, Bjorn Erik
    BMJ-BRITISH MEDICAL JOURNAL, 2012, 345
  • [9] Metal-on-metal hip replacements: implications for general practice
    Matharu, Victoria K.
    Matharu, Gulraj S.
    BRITISH JOURNAL OF GENERAL PRACTICE, 2017, 67 (665): : 544 - 545
  • [10] Failure rates of stemmed metal-on-metal hip replacements
    Amstutz, Harlan C.
    Le Duff, Michel J.
    LANCET, 2012, 380 (9837): : 105 - 105