Evaluation of a Passive Air Sampler for Measuring Indoor Formaldehyde

被引:0
|
作者
Sun-Tae Kim
Bongbeen Yim
Jaeho Jeong
机构
[1] Daejeon University,Department of Environmental Engineering, College of Engineering
[2] R & D Center,undefined
[3] Envors Co.,undefined
[4] Ltd.,undefined
来源
Analytical Sciences | 2007年 / 23卷
关键词
D O I
暂无
中图分类号
学科分类号
摘要
A passive air sampler, using 4-amino-3-hydrazino-5-mercapto-1,2,4-triazole, was evaluated for the determination of formaldehyde in indoor environments. Chromatography paper cleaned using a 3% hydrogen peroxide solution was experimentally determined as being the optimum absorption filter for the collection of formaldehyde (0.05 μg cm-2 formaldehyde). From a linear-regression analysis between the mass of formaldehyde time-collected on a passive air sampler and the formaldehyde concentration measured by an active sampler, the sampling rate of the passive air sampler was 1.52 L h-1. The sampling rate, determined for the passive air sampler in relation to the temperature (19–28°C) and the relative humidity (30–90%), were 1.56 ± 0.04 and 1.58 ± 0.07 L h-1, respectively. The relationship between the sampling rate and the air velocity was a linear-regression within the observed range. In the case of exposed samplers, the stability of the collected formaldehyde decreased with increasing storage time (decrease of ca. 25% after 22 days); but with the unexposed samplers the stability of the blank remained relatively unchanged for 7 days (decrease of ca. 37% after 22 days). The detection limits for the passive air sampler with an exposure time of 1 day and 7 days were 10.4 and 1.48 μg m-3, respectively.
引用
收藏
页码:497 / 500
页数:3
相关论文
共 50 条
  • [31] Evidence of indoor sinks for formaldehyde through the field measurements using passive flux sampler and mass balance
    Plaisance, Herve
    Blondel, Alodie
    Desauziers, Valerie
    Mocho, Pierre
    ENVIRONMENTAL SCIENCE AND POLLUTION RESEARCH, 2019, 26 (29) : 29679 - 29686
  • [32] Characteristics of formaldehyde emissions from indoor materials assessed by a method using passive flux sampler measurements
    Plaisance, H.
    Blondel, A.
    Desauziers, V.
    Mocho, P.
    BUILDING AND ENVIRONMENT, 2014, 73 : 249 - 255
  • [33] Evidence of indoor sinks for formaldehyde through the field measurements using passive flux sampler and mass balance
    Herve Plaisance
    Alodie Blondel
    Valerie Desauziers
    Pierre Mocho
    Environmental Science and Pollution Research, 2019, 26 : 29679 - 29686
  • [34] Development of a Passive Sampler using 4-amino-3-hydrazino-5-mercapto- 1, 2, 4-triazole for Measuring Indoor Formaldehyde
    Kim, Sun-Tae
    Vim, Bongbeen
    Jeong, Jaeho
    JOURNAL OF KOREAN SOCIETY FOR ATMOSPHERIC ENVIRONMENT, 2005, 21 (06) : 593 - 603
  • [35] MODIFICATION OF A HIGH-EFFICIENCY PASSIVE SAMPLER TO DETERMINE NITROGEN-DIOXIDE OR FORMALDEHYDE IN AIR
    MULIK, JD
    LEWIS, RG
    MCCLENNY, WA
    WILLIAMS, DD
    ANALYTICAL CHEMISTRY, 1989, 61 (02) : 187 - 189
  • [36] The applicability of DNPH active collectors as passive collectors for the determination of formaldehyde in indoor air
    Witthauer, J
    Heilemann, KJ
    Bischof, W
    GEFAHRSTOFFE REINHALTUNG DER LUFT, 1999, 59 (10): : 369 - 373
  • [37] Development of colorimetric type passive flux sampler for formaldehyde
    Shinohara, N
    Kidani, A
    Kobayashi, T
    Uchi, T
    Ochiai, S
    Fujii, M
    Nakano, N
    Yanagisawa, Y
    INDOOR AIR 2005: PROCEEDINGS OF THE 10TH INTERNATIONAL CONFERENCE ON INDOOR AIR QUALITY AND CLIMATE, VOLS 1-5, 2005, : 2665 - 2668
  • [38] Evaluation and application of a passive air sampler for polycylic aromatic hydrocarbons (PAHs)
    Esen, Fatma
    Evci, Yildiz M.
    Tasdemir, Yucel
    JOURNAL OF ENVIRONMENTAL SCIENCE AND HEALTH PART A-TOXIC/HAZARDOUS SUBSTANCES & ENVIRONMENTAL ENGINEERING, 2017, 52 (10): : 1022 - 1029
  • [39] Passive flux sampler for measurement of formaldehyde emission rates
    Shinohara, Naohide
    Fujii, Minoru
    Yamasaki, Akihiro
    Yanagisawa, Yukio
    ATMOSPHERIC ENVIRONMENT, 2007, 41 (19) : 4018 - 4028
  • [40] Evaluation and Application of a Passive Air Sampler for Atmospheric Volatile Organic Compounds
    Huang, Cenyan
    Tong, Lei
    Dai, Xiaorong
    Xiao, Hang
    AEROSOL AND AIR QUALITY RESEARCH, 2018, 18 (12) : 3047 - 3055