A non-enzymatic hydrogen peroxide sensor was developed using gold@platinum nanoparticles (Au@PtNPs) with core@shell structure fabricated on glassy carbon electrode (GCE) by electroless deposition method. Initially, gold nanoparticles (AuNPs) were deposited on GCE by reducing HAuCl4 in the presence of NH2OH and the deposited AuNPs on GCE act as the nucleation centre for the deposition of platinum nanoparticles (PtNPs) in the presence of H2PtCl6 and NH2OH. SEM and AFM studies demonstrated that the electroless deposition of Pt on Au was isotropic and uniform. Further, Au@PtNP-modified substrates were characterized by X-ray photoelectron spectroscopy (XPS), energy dispersive X-ray analysis (EDAX) and cyclic voltammetry (CV). XPS showed characteristic binding energies at 71.2 and 74.4 eV for PtNPs and, 83.6 and 87.3 eV for AuNPs indicating the zero-valent nature in both of them. The electrocatalytic activity of Au@PtNP-modified electrode was investigated towards hydrogen peroxide (HP) reduction. The modified electrode exhibited higher electrocatalytic activity towards HP by not only shifting its reduction potential by 370 mV towards less positive potential but also by enhancing the reduction current when compared to bare and AuNP-modified GCE. The present method shows better sensitivity compared to the reported methods in literature and the detection limit was found to be 60 nM.
引用
收藏
页码:331 / 338
页数:7
相关论文
共 50 条
  • [11] Non-enzymatic sensor based on a glassy carbon electrode modified with Ag nanoparticles/polyaniline/halloysite nanotube nanocomposites for hydrogen peroxide sensing
    Yang, Ziyin
    Zheng, Xiaohui
    Zheng, Jianbin
    [J]. RSC ADVANCES, 2016, 6 (63) : 58329 - 58335
  • [12] Gold nanoparticles directly modified glassy carbon electrode for non-enzymatic detection of glucose
    Chang, Gang
    Shu, Honghui
    Ji, Kai
    Oyama, Munetaka
    Liu, Xiong
    He, Yunbin
    [J]. APPLIED SURFACE SCIENCE, 2014, 288 : 524 - 529
  • [13] Electrosynthesis of gold nanoparticles/porous GaN electrode for non-enzymatic hydrogen peroxide detection
    Zhang, Miao-Rong
    Chen, Xue-Qing
    Pan, Ge-Bo
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2017, 240 : 142 - 147
  • [14] Non-enzymatic amperometric hydrogen peroxide sensor using a glassy carbon electrode modified with gold nanoparticles deposited on CVD-grown graphene
    Yawen Yuan
    Yiqun Zheng
    Jinglei Liu
    Hua Wang
    Shifeng Hou
    [J]. Microchimica Acta, 2017, 184 : 4723 - 4729
  • [15] Non-enzymatic amperometric hydrogen peroxide sensor using a glassy carbon electrode modified with gold nanoparticles deposited on CVD-grown graphene
    Yuan, Yawen
    Zheng, Yiqun
    Liu, Jinglei
    Wang, Hua
    Hou, Shifeng
    [J]. MICROCHIMICA ACTA, 2017, 184 (12) : 4723 - 4729
  • [16] A non-enzymatic hydrogen peroxide sensor based on poly(vinyl alcohol)-multiwalled carbon nanotubes-platinum nanoparticles hybrids modified glassy carbon electrode
    Fang, Yuxin
    Zhang, Di
    Qin, Xia
    Miao, Zhiying
    Takahashi, Shigehiro
    Anzai, Jun-ichi
    Chen, Qiang
    [J]. ELECTROCHIMICA ACTA, 2012, 70 : 266 - 271
  • [17] Copper oxide nanoparticles and ionic liquid modified carbon electrode for the non-enzymatic electrochemical sensing of hydrogen peroxide
    Ping, Jianfeng
    Ru, Shiping
    Fan, Kai
    Wu, Jian
    Ying, Yibin
    [J]. MICROCHIMICA ACTA, 2010, 171 (1-2) : 117 - 123
  • [18] Copper oxide nanoparticles and ionic liquid modified carbon electrode for the non-enzymatic electrochemical sensing of hydrogen peroxide
    Ping, Jianfeng
    Ru, Shiping
    Fan, Kai
    Wu, Jian
    Ying, Yibin
    [J]. Microchimica Acta, 2010, 171 (02) : 117 - 123
  • [19] Copper oxide nanoparticles and ionic liquid modified carbon electrode for the non-enzymatic electrochemical sensing of hydrogen peroxide
    Jianfeng Ping
    Shiping Ru
    Kai Fan
    Jian Wu
    Yibin Ying
    [J]. Microchimica Acta, 2010, 171 : 117 - 123
  • [20] Simple non-enzymatic electrochemical sensor for hydrogen peroxide based on nafion/platinum nanoparticles/reduced graphene oxide nanocomposite modified glassy carbon electrode
    Zhang, Cong
    Jiang, Haohai
    Ma, Rui
    Zhang, Yanyan
    Chen, Qiang
    [J]. IONICS, 2017, 23 (05) : 1309 - 1317