Actinide molecular ion formation in collision/reaction cell of triple quadrupole ICP-MS/MS and its application to quantitative actinide analysis

被引:0
|
作者
T. Suzuki
T. Yamamura
C. Abe
K. Konashi
Y. Shikamori
机构
[1] Nagaoka University of Technology,Department of Nuclear System Safety Engineering
[2] Institute for Materials Research,Laboratory of Alpha
[3] Tohoku University,Ray Emitters
[4] Kyoto University,Institute for Integrated Radiation and Nuclear Science
[5] Tohoku University,Institute for Materials Research
[6] Agilent Technologies Japan,undefined
[7] Ltd.,undefined
关键词
Uranium; Neptunium; Plutonium; Americium; Curium; Oxygen reaction;
D O I
暂无
中图分类号
学科分类号
摘要
A study for the quantitative actinide analysis by using a triple quadrupole ICP-MS/MS, Agilent 8900, was carried out. We confirmed the formation percentages of ionic species of each actinide in several kinds of collision/reaction cell modes. It was found that the dominant ions in oxygen gas mode were oxide ions, and the ratio of monoxide ion/dioxide ion increases with atomic number in the case of same oxygen gas flow rate. The percentages of actinide ion and oxide ions depend on the oxygen gas flow rate, and the number of oxygen attached on actinides increases with the oxygen gas flow rate.
引用
收藏
页码:221 / 225
页数:4
相关论文
共 50 条
  • [31] Coupling between a calix[6]arene-based chromatography column and ICP-MS for on-line actinide analysis
    Bouvier-Capely, C. (celine.bouvier@irsn.fr), 1600, Royal Society of Chemistry (31):
  • [32] Comprehensive Analysis of Nanoparticles using Single and Triple Quadrupole ICP-MS and a Dedicated Data Evaluation Tool
    Kutscher, Daniel
    Ducos, Shona McShcehy
    Willis, Julian D.
    Siqueira, Bruno Menezes
    BRAZILIAN JOURNAL OF ANALYTICAL CHEMISTRY, 2018, 5 (20): : 60 - 67
  • [33] Single particle analysis of TiO2 in candy products using triple quadrupole ICP-MS
    Candas-Zapico, S.
    Kutscher, D. J.
    Montes-Bayon, M.
    Bettmer, J.
    TALANTA, 2018, 180 : 309 - 315
  • [34] Kinetic energy discrimination in collision/reaction cell ICP-MS: Theoretical review of principles and limitations
    Yamada, Noriyuki
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2015, 110 : 31 - 44
  • [35] Accurate, Low-Level Sulfur Analysis by ICP-MS Using MS/MS with Oxygen Reaction Cell Gas
    McCurdy, Ed
    Woods, Glenn
    Georg, Bastian
    Sugiyama, Naoki
    SPECTROSCOPY, 2020, 35 : 27 - 34
  • [36] Effects of deuterium in octopole reaction and collision cell ICP-MS on detection of selenium in extracellular fluids
    Ogra, Y
    Ishiwata, K
    Suzuki, KT
    ANALYTICA CHIMICA ACTA, 2005, 554 (1-2) : 123 - 129
  • [37] In-situ analysis of Os and Pb isotope ratios using laser ablation and collision-cell quadrupole ICP-MS
    Cox, RA
    Barnes, SJ
    GEOCHIMICA ET COSMOCHIMICA ACTA, 2005, 69 (10) : A375 - A375
  • [38] Elemental Analysis of Tree Nuts and Their Pressed Oils by Triple Quadrupole Inductively Coupled Plasma-Mass Spectrometry (ICP-MS/MS)
    Ni, Zhanglin
    Yang, Ling
    Qu, Minghua
    Li, Zhengxiang
    Tang, Fubin
    ANALYTICAL LETTERS, 2025, 58 (01) : 79 - 90
  • [39] Multielement analysis of human seminal plasma by octopole reaction cell ICP-MS
    20144300126370
    Živković, Tanja, 1600, Royal Society of Chemistry (29):
  • [40] Multielement analysis of human seminal plasma by octopole reaction cell ICP-MS
    Zivkovic, Tanja
    Tariba, Blanka
    Pizent, Alica
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2014, 29 (11) : 2114 - 2126