Reproducibility of measurements with a new slit lamp-mounted ocular blood flow tonograph

被引:0
|
作者
C. W. Spraul
Gabriele E. Lang
Marina Ronzani
Josef Högel
Gerhard K. Lang
机构
[1] Department of Ophthalmology,
[2] University of Ulm,undefined
[3] Eye Clinic and Hospital,undefined
[4] Prittwitzstrasse 43,undefined
[5] D-89075 Ulm,undefined
[6] Germany Fax +49-731-50-27549; e-mail Christoph.Spraul@Medizin.Uni-Ulm.de,undefined
[7] Department of Statistics,undefined
[8] University of Ulm,undefined
[9] Ulm,undefined
[10] Germany,undefined
关键词
Blood Flow; Linear Regression; Regression Analysis; Regression Coefficient; Linear Regression Analysis;
D O I
暂无
中图分类号
学科分类号
摘要
· Purpose: To assess the reliability of measurements of intraocular pressure (IOP), pulse amplitude (PA), and pulsatile ocular blood flow (POBF) and the validity of measurements of IOP with a new ocular blood flow tonograph. · Methods: Intraocular pressure pulse was assessed with pneumatic tonometry. In this study, we used the OBF Labs ocular blood flow tonograph with two different pneumatic probes, a modified Langham probe and a newly developed probe. One ophthalmologist compared two machines, first with the modified Langham probe and secondly with the newly developed probe (40 volunteers). Additionally, two ophthalmologists performed measurements on different days and in different sequence (34 volunteers). Furthermore, results of IOP measurements performed with this tonograph were compared to those of Goldmann applanation tonometry (213 volunteers). · Results: Using the modified Langham probe, POBF and PA were associated with the machine used for the test. However, using the newly developed pobe, no relevant or statistically significant differences were found for any variables, and reliability coefficients were between 0.70 and 0.90. Linear regression analysis of the Goldmann applanation tonometer on the measurements of IOP with the tonograph showed a regression coefficient of 0.765. · Conclusions: Results of measurements performed with this ocular blood flow tonograph and the newly developed probe are expected to be reliable and comparable.
引用
收藏
页码:274 / 279
页数:5
相关论文
共 50 条
  • [21] Reproducibility of a new technique to analyse retinal blood flow
    Iester, M
    Altier, M
    Vittone, P
    Zingirian, M
    Traverso, CE
    ACTA OPHTHALMOLOGICA SCANDINAVICA, 2002, 80 : 51 - 52
  • [22] Hemodynamic reproducibility during blood flow measurements of hemodialysis synthetic grafts
    DeSoto, DJ
    Ram, SJ
    Faiyaz, R
    Birk, CG
    Paulson, WD
    AMERICAN JOURNAL OF KIDNEY DISEASES, 2001, 37 (04) : 790 - 796
  • [23] Reproducibility of LDF blood flow measurements: Dynamical characterization versus averaging
    Stefanovska, A.
    Sheppard, L. W.
    Stankovski, T.
    McClintock, P. V. E.
    MICROVASCULAR RESEARCH, 2011, 82 (03) : 274 - 276
  • [24] Ocular blood flow measurements in healthy human myopic eyes
    Benavente-Perez, Alexandra
    Hosking, Sarah L.
    Logan, Nicola S.
    Broadway, David C.
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (11) : 1587 - 1594
  • [25] Ocular blood flow measurements in healthy human myopic eyes
    Alexandra Benavente-Pérez
    Sarah L. Hosking
    Nicola S. Logan
    David C. Broadway
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2010, 248 : 1587 - 1594
  • [26] A comparison of ocular blood flow measurements using two different techniques; Pulsatile ocular blood flow and color Doppler imaging.
    Chung, HS
    Harris, A
    McNulty, L
    Evans, DW
    Kagemann, L
    Kagemann, C
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1997, 38 (04) : 4891 - 4891
  • [27] Reproducibility of blood flow velocity measurements using colour decoded Doppler imaging
    Matthiessen, ET
    Zeitz, O
    Richard, G
    Klemm, M
    EYE, 2004, 18 (04) : 400 - 405
  • [28] Reproducibility of blood flow velocity measurements using colour decoded Doppler imaging
    E T Matthiessen
    O Zeitz
    G Richard
    M Klemm
    Eye, 2004, 18 : 400 - 405
  • [29] REPRODUCIBILITY OF REGIONAL MYOCARDIAL BLOOD-FLOW MEASUREMENTS WITH XE-133
    FLECK, E
    DIRSCHINGER, J
    REDL, A
    RUDOLPH, W
    ZEITSCHRIFT FUR KARDIOLOGIE, 1978, 67 (03): : 204 - 204
  • [30] Reproducibility of measurements of regional resting and hyperemic myocardial blood flow assessed with PET
    Nagamachi, S
    Czernin, J
    Kim, AS
    Sun, KT
    Bottcher, M
    Phelps, ME
    Schelbert, HR
    JOURNAL OF NUCLEAR MEDICINE, 1996, 37 (10) : 1626 - 1631