Reproducibility of measurements with a new slit lamp-mounted ocular blood flow tonograph

被引:0
|
作者
C. W. Spraul
Gabriele E. Lang
Marina Ronzani
Josef Högel
Gerhard K. Lang
机构
[1] Department of Ophthalmology,
[2] University of Ulm,undefined
[3] Eye Clinic and Hospital,undefined
[4] Prittwitzstrasse 43,undefined
[5] D-89075 Ulm,undefined
[6] Germany Fax +49-731-50-27549; e-mail Christoph.Spraul@Medizin.Uni-Ulm.de,undefined
[7] Department of Statistics,undefined
[8] University of Ulm,undefined
[9] Ulm,undefined
[10] Germany,undefined
关键词
Blood Flow; Linear Regression; Regression Analysis; Regression Coefficient; Linear Regression Analysis;
D O I
暂无
中图分类号
学科分类号
摘要
· Purpose: To assess the reliability of measurements of intraocular pressure (IOP), pulse amplitude (PA), and pulsatile ocular blood flow (POBF) and the validity of measurements of IOP with a new ocular blood flow tonograph. · Methods: Intraocular pressure pulse was assessed with pneumatic tonometry. In this study, we used the OBF Labs ocular blood flow tonograph with two different pneumatic probes, a modified Langham probe and a newly developed probe. One ophthalmologist compared two machines, first with the modified Langham probe and secondly with the newly developed probe (40 volunteers). Additionally, two ophthalmologists performed measurements on different days and in different sequence (34 volunteers). Furthermore, results of IOP measurements performed with this tonograph were compared to those of Goldmann applanation tonometry (213 volunteers). · Results: Using the modified Langham probe, POBF and PA were associated with the machine used for the test. However, using the newly developed pobe, no relevant or statistically significant differences were found for any variables, and reliability coefficients were between 0.70 and 0.90. Linear regression analysis of the Goldmann applanation tonometer on the measurements of IOP with the tonograph showed a regression coefficient of 0.765. · Conclusions: Results of measurements performed with this ocular blood flow tonograph and the newly developed probe are expected to be reliable and comparable.
引用
收藏
页码:274 / 279
页数:5
相关论文
共 50 条
  • [1] Reproducibility of measurements with a new slit lamp-mounted ocular blood flow tonograph
    Spraul, CW
    Lang, GE
    Ronzani, M
    Hogel, J
    Lang, GK
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1998, 236 (04) : 274 - 279
  • [2] RELIABILITY OF MEASUREMENTS WITH A NEW SLIT LAMP MOUNTED OCULAR BLOOD-FLOW TONOGRAPH
    SPRAUL, CW
    LANG, GE
    RONZANI, M
    PILLUNAT, LE
    HOGEL, I
    LANG, GK
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1995, 36 (04) : S436 - S436
  • [3] Applanation tonometry: a comparison of the Perkins handheld and Goldmann slit lamp-mounted methods
    Arora, R.
    Bellamy, H.
    Austin, M. W.
    CLINICAL OPHTHALMOLOGY, 2014, 8 : 605 - 610
  • [4] The Ocular Blood Flow Tonograph: A new instrument for the measurement of intraocular pressure in rabbits
    Chidlow, G
    Nash, MS
    Crowhurst, C
    Bron, AJ
    Osborne, NN
    EXPERIMENTAL EYE RESEARCH, 1996, 63 (04) : 463 - 469
  • [5] Clinical comparison of the ocular blood flow tonograph and the Goldmann applanation tonometer
    Esgin, H
    Alimgil, ML
    Erda, S
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 1998, 8 (03) : 162 - 166
  • [6] Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer
    Gunvant, P
    Baskaran, M
    Vijaya, L
    Joseph, IS
    Watkins, RJ
    Nallapothula, M
    Broadway, DC
    O'Leary, DJ
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2004, 88 (04) : 518 - 522
  • [7] Ocular biometric measurements with a slit-lamp method (Orbscan®).
    Allouch, C
    Touzeau, O
    Borderie, V
    Bourcier, T
    Scheer, S
    Ameline, B
    Monin, C
    Morel, C
    Larricart, P
    Laroche, L
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2001, 42 (04) : S894 - S894
  • [8] Ocular biometric measurements with a slit-lamp method (Orbscan®)
    Allouch, C
    Touzeau, O
    Borderie, V
    Puech, M
    Scheer, S
    Laroche, L
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2001, 24 (07): : 710 - 715
  • [9] Pulsatile ocular blood flow measurements in healthy eyes: Reproducibility and reference values
    Yang, YC
    Hulbert, MFG
    Batterbury, M
    Clearkin, LG
    JOURNAL OF GLAUCOMA, 1997, 6 (03) : 175 - 179
  • [10] Repeatability and Reproducibility of Slit Lamp, Optical Coherence Tomography, and Scheimpflug Measurements of Corneal Scars
    Das, Manoranjan
    Menda, Shivali A.
    Panigrahi, Arun K.
    Prajna, N. Venkatesh
    Yen, Michael
    Tsang, Betty
    Kumar, Alisha
    Rose-Nussbaumer, Jennifer
    Acharya, Nisha R.
    McCulloch, Charles E.
    Lietman, Thomas M.
    McLeod, Stephen D.
    Keenan, Jeremy D.
    OPHTHALMIC EPIDEMIOLOGY, 2019, 26 (04) : 251 - 256