Cost-benefit analysis of intraoperative 3D imaging [Kosten- und nutzenanalyse der intraoperativen 3D-bildgebung]

被引:0
|
作者
Hüfner T. [1 ,2 ]
Stübig T. [1 ]
Gösling T. [1 ]
Kendoff D. [1 ]
Geerling J. [1 ]
Krettek C. [1 ]
机构
[1] Unfallchirurgische Klinik, Medizinische Hochschule, Hannover
[2] Unfallchirurgische Klinik, Medizinische Hochschule, 30625 Hannover
来源
Der Unfallchirurg | 2007年 / 110卷 / 1期
关键词
Cost analysis; Costs per minute; Intraoperative 3D imaging; Iso-C[!sup]3D[!/sup;
D O I
10.1007/s00113-006-1202-6
中图分类号
学科分类号
摘要
Background. With intraoperative 3D imaging, inevitable corrections may be done already during the operation, and a second procedure can be avoided. The purpose of this study was to perform a cost-benefit analysis during the first year of intraoperative 3D application in order to provide a cost transparency for the surgeon within the current DRG system. Methods. On the basis of internal data and the literature, the annual operating costs of the ISO-C3D were calculated at 27,940 euros (purchase price, depreciation, maintenance, repair), the costs of an average revision as a secondary operation at 2,385 euros (costs avoided with the ISO-C3D, Siremobil, Siemens, Erlangen, Germany), and the dynamic costs of an intraoperative 3D scan were averaged to 131.08 euros (draping, additional time, personnel). Results. In the year 2003 intraoperative 3D scanning was done in 126 patients, and intraoperative revision was performed in 24 (19%) due to the additional intraoperative 3D information provided by the Siremobil. In 11 (8.7%) patients the implant position was corrected and in 13 (10.3%) patients the reduction was improved. Taking only fixed costs into consideration, 29,311.52 euros could be saved, and when fixed and dynamic costs are taken into account 12,795.44 euros could be saved. Since the parameters for each hospital are different, the following formula for an individual computation is suitable. For the calculation of the cost the following mathematical relationship results: (annual fixed costs) + (costs per scan × number of cases) - (revision costs × revision rate [p]) × number of cases [N]) = 0. Discussion. Although the costs of an ISO-C3D are considerably high, an economic benefit can also accrue with frequent application and high rates of avoided revision. However, if the rate of avoided revision adds up to only 5%, a substantial deficit may result. © 2006 Springer Medizin Verlag.
引用
收藏
页码:14 / 21
页数:7
相关论文
共 50 条
  • [1] Cost-benefit analysis of intraoperative 3D imaging
    Huefner, T.
    Stuebig, T.
    Goesling, T.
    Kendoff, D.
    Geerling, J.
    Krettek, C.
    [J]. UNFALLCHIRURG, 2007, 110 (01): : 14 - 21
  • [2] Intraoperative 3‑D-Bildgebung in der WirbelsäulenchirurgieIntraoperative 3D imaging in spinal surgery
    O. Gonschorek
    S. Hauck
    V. Bühren
    [J]. Der Unfallchirurg, 2016, 119 (10): : 817 - 824
  • [3] Intraoperative imaging with the ISO C 3D [Intraoperative bildgebung mit dem ISO C 3D]
    Wich M.
    Spranger N.
    Ekkernkamp A.
    [J]. Der Chirurg, 2004, 75 (10): : 982 - 987
  • [4] Möglichkeiten und Grenzen der intraoperativen 2D-Bildgebung in der UnfallchirurgiePossibilities and limits of intraoperative 2D imaging in trauma surgery
    Julius Watrinet
    Lisa Wenzel
    Julian Fürmetz
    Peter Augat
    Philipp Blum
    Claas Neidlein
    Markus Bormann
    Fabian Stuby
    Christian von Rüden
    [J]. Die Unfallchirurgie, 2023, 126 (12) : 935 - 941
  • [5] Cost-benefit analysis of 3D conformal radiation therapyKosten-Nutzen-Analyse der 3D-Konformationsbestrahlung
    S. H. Levitt
    F. M. Khan
    P. D. Higgins
    M. B. Nierengarten
    [J]. Strahlentherapie und Onkologie, 1997, 173 (9) : 441 - 443
  • [6] Wirtschaftlichkeit der intraoperativen 3D-Bildgebung mit einem mobilen chirurgischen C-BogenCost effectiveness of intraoperative three-dimensional imaging with a mobile surgical C-arm
    M. Bischoff
    A. Hebecker
    E. Hartwig
    F. Gebhard
    [J]. Der Unfallchirurg, 2004, 107 (8): : 712 - 715
  • [7] Intraoperative 3‑D-Bildgebung als Ergänzung oder Ersatz der intraoperativen Navigation?Qualitätssicherung in der computerassistierten ChirurgieIntraoperative 3D imaging as a complement or an alternative to intraoperative navigation?Quality assurance in computer-assisted surgery
    M. E. H. Wagner
    H. Essig
    [J]. Der MKG-Chirurg, 2017, 10 (3): : 190 - 196
  • [8] Application fields of intraoperative transesophageal 3D echocardiography [Anwendungsgebiete der intraoperativen transösophagealen 3D-Echokardiographie]
    Magunia H.
    Rosenberger P.
    [J]. Der Anaesthesist, 2014, 63 (8-9) : 662 - 677
  • [9] Acetabular defect classification in times of 3D imaging and patient-specific treatment protocols; [Periazetabuläre Defektklassifikation in Zeiten von 3D-Bildgebung und patientenspezifischer Therapieprotokolle]
    Horas K.
    Arnholdt J.
    Steinert A.F.
    Hoberg M.
    Rudert M.
    Holzapfel B.M.
    [J]. Der Orthopäde, 2017, 46 (2): : 168 - 178
  • [10] Intraoperative three-dimensional imaging to monitor selective decompression in lumbar spinal stenosis [Intraoperative 3D-bildgebung zur kontrolle der selektiven dekompression bei lumbaler wirbelkanalstenose]
    Mauer U.M.
    Kunz U.
    [J]. Der Orthopäde, 2006, 35 (12): : 1258 - 1260