A new method with variable injection parameters in contrast-enhanced CT: a phantom study for evaluating an aortic peak enhancement

被引:1
|
作者
Terasawa K. [1 ,2 ]
Maruyama A. [3 ]
Tsukimata T. [1 ]
机构
[1] Central Radiology Department, Nagano Red Cross Hospital, 5-22-1, Wakasato, Nagano-shi, 380-8582, Nagano
[2] Graduate School of Health Science, Suzuka University of Medical Science, 1001-1, Kishioka-cho, Suzuka-shi, 510-0293, Mie
[3] Department of Radiology, Nagano Red Cross Hospital, Nagano
关键词
Contrast enhancement; Contrast material; Time-enhancement curve (TEC); Variable injection method (VIM); Variation factor (VF);
D O I
10.1007/s12194-015-0314-5
中图分类号
学科分类号
摘要
Contrast-enhanced CT employs a standard uniphasic single-injection method (SIM), wherein administration is based on two parameters: the iodine administration rate (mgI/s) and the injection duration (s). However, as the SIM uses a fixed iodine administration rate, only a uniform contrast enhancement can be achieved with this method. The iodine administration rate can be increased only by increasing the iodine dose or shortening the injection duration, and no arbitrary adjustments can be made to the peak enhancement characteristics of the time-enhancement curves (TECs) at the fixed injection parameters used in the SIM. To address this problem, we developed a variable injection method (VIM) with a new parameter, the variation factor (VF), to adjust the TECs. A phantom study with the VIM indicated that arbitrary adjustments to the iodine administration rate could be made without changing the injection duration or increasing the iodine load. In our study, VFs of 0.3 and 0.5, which showed earlier achievement of peak enhancements, showed better temporal separation between arterial vasculature and parenchyma or the venous vasculature than that obtained with the SIM. The higher peak enhancement provided by the VF of 0.3 was also considered to improve the contrast in qualitative diagnostic examinations. A VF of 0.5 increased the duration of the enhancement and was considered to produce stable enhancement of contrast in vascular investigations. The VF is now an essential parameter, and the VIM is useful as a reasonable contrast method that may contribute to both improved visualization and improvement in the accuracy of morphologic diagnosis. © 2015, Japanese Society of Radiological Technology and Japan Society of Medical Physics.
引用
收藏
页码:248 / 257
页数:9
相关论文
共 50 条
  • [1] Prediction of aortic peak enhancement in monophasic contrast injection protocols at multidetector CT: Phantom and patient studies
    Awai K.
    Nakayama Y.
    Nakaura T.
    Yanaga Y.
    Tamura Y.
    Hatemura M.
    Funama Y.
    Yamashita Y.
    Radiation Medicine, 2007, 25 (1): : 14 - 21
  • [2] Contrast-enhanced spiral CT of the liver: Effect of injection time on time to peak hepatic enhancement
    Department of Radiology, National Defense Medical College, 3-2, Namiki, Tokorozawa, Saitama 359, Japan
    J. COMPUT. ASSISTED TOMOGR., 4 (633-637):
  • [3] Contrast-enhanced spiral CT of the liver: Effect of injection time on time to peak hepatic enhancement
    Irie, T
    Kusano, S
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1996, 20 (04) : 633 - 637
  • [4] PREDICTION OF THE TIME TO PEAK HEPATIC ENHANCEMENT TO OPTIMIZE CONTRAST-ENHANCED SPIRAL CT
    IRIE, T
    SUZUKI, S
    YAMAUCHI, T
    KUSANO, S
    ACTA RADIOLOGICA, 1995, 36 (02) : 154 - 158
  • [5] Simulation of aortic peak enhancement on MDCT using a contrast material flow phantom: Feasibility study
    Awai, K
    Hatcho, A
    Nakayama, Y
    Kusunoki, S
    Liu, D
    Hatemura, M
    Funama, Y
    Denbo, M
    Sato, N
    Yamashita, Y
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (02) : 379 - 385
  • [6] CONTRAST-ENHANCED SPIRAL CT OF THE LIVER - EFFECT OF DIFFERENT AMOUNTS AND INJECTION RATES OF CONTRAST MATERIAL ON EARLY CONTRAST ENHANCEMENT
    SMALL, WC
    NELSON, RC
    BERNARDINO, ME
    BRUMMER, LT
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1994, 163 (01) : 87 - 92
  • [7] Effect of decreased contrast injection flow rate on aortic enhancement in 80-KV peak CT with contrast dose reduction
    Morisaka, Hiroyuki
    Matsuura, Koichiro
    Yamaguchi, Haruomi
    Ichikawa, Tomoaki
    Onishi, Hiroshi
    ACTA RADIOLOGICA, 2023, 64 (01) : 353 - 359
  • [8] Effect of varying injection rates of a saline chaser on aortic enhancement in CT angiography: phantom study
    Schindera, Sebastian T.
    Nelson, Rendon C.
    Howle, Laurens
    Nichols, Eli
    DeLong, David M.
    Merkle, Elmar M.
    EUROPEAN RADIOLOGY, 2008, 18 (08) : 1683 - 1689
  • [9] Effect of varying injection rates of a saline chaser on aortic enhancement in CT angiography: phantom study
    Sebastian T. Schindera
    Rendon C. Nelson
    Laurens Howle
    Eli Nichols
    David M. DeLong
    Elmar M. Merkle
    European Radiology, 2008, 18 : 1683 - 1689
  • [10] The Frequency of Insufficient Contrast Enhancement of the Pulmonary Artery in Routine Contrast-enhanced Chest CT and Its Improvement With an Increased Injection Rate A Prospective Study
    Ozawa, Yoshiyuki
    Hara, Masaki
    Shibamoto, Yuta
    JOURNAL OF THORACIC IMAGING, 2011, 26 (01) : 42 - 47