Submillisievert imaging protocol using full reconstruction and advanced patient motion correction in 320-row area detector coronary CT angiography

被引:0
|
作者
Yuko Kawaguchi
Shinichiro Fujimoto
Kazuhisa Takamura
Etsuro Kato
Shoko Suda
Rie Matsumori
Makoto Hiki
Kanako K. Kumamaru
Hiroyuki Daida
机构
[1] Juntendo University Graduate School of Medicine,Department of Cardiovascular Medicine
[2] Juntendo University Graduate School of Medicine,Department of Radiology
关键词
320-row area detector coronary computed tomography angiography; Full reconstruction; Advanced patient motion correction; Radiation dosing; Submillisievert;
D O I
暂无
中图分类号
学科分类号
摘要
Radiation exposure remains a concern in the use of coronary CT angiography (CCTA). Full reconstruction (Full) and reconstruction using advanced patient motion correction (APMC) could obtain a lower radiation dose using low tube current scanning in a 320-row Area Detector CT (320-ADCT). The radiation dose for an imaging protocol using Full and APMC in daily practice was estimated. A total of 209 patients who underwent CCTA in 1 rotation scanning with 100 kv and adaptive iterative dose reduction 3D in 320-ADCT were enrolled. Imaging protocols were classified into 3 groups based on estimated slow filling time: (1) slow filling time ≥ 275 msec, Full with 30% of usual tube current (N = 43)(Full30%mA) (2) 206.3 msec ≤ slow filling time < 275 msec, APMC with 50% of usual tube current (N = 48)(APMC50%mA); and (3) 137.5 msec ≤ slow filling time < 206.3 msec, Half reconstruction with usual tube current (N = 118)(Half100%mA). Radiation dose was estimated by the effective dose. The diagnostic accuracy of CCTA was compared with that of invasive coronary angiography in 28 patients. The effective doses of Full30%mA, APMC50%mA, and Half100%mA were 0.77 ± 0.31, 1.30 ± 0.85, and 1.98 ± 0.68, respectively. Of 28 patients, the sensitivity, specificity, accuracy, positive predictive value, and negative predictive value in vessel-based analyses were: Full30%mA, 66.7, 82.4, 80.0, 40.0, and 93.3%; APMC50%mA, 100.0, 80.0, 83.3, 50.05, and 100.0%; and Half100%mA, 90.9, 83.0, 86.3, 78.95, and 92.9%, respectively. An imaging protocol using Full30%mA and APMC50%mA was one of the methods how radiation dose could be reduced radiation dose maintained diagnostic accuracy compared to imaging using conventional Half100%mA.
引用
收藏
页码:465 / 474
页数:9
相关论文
共 50 条
  • [11] Coronary computed tomography angiography and endocardial leads - Image quality in 320-row CT using iterative reconstruction
    Garmer, Marietta
    Bonsels, Marc
    Metz, Frauke
    Klein-Wiele, Oliver
    Brandts, Bodo
    Groenemeyer, Dietrich
    CLINICAL IMAGING, 2018, 50 : 157 - 163
  • [12] Feasibility of 320-row area detector CT coronary angiography using 40 mL of contrast material: assessment of image quality and diagnostic accuracy
    Rihyeon Kim
    Eun-Ah Park
    Whal Lee
    Jin Wook Chung
    European Radiology, 2016, 26 : 3802 - 3810
  • [13] Feasibility of 320-row area detector CT coronary angiography using 40 mL of contrast material: assessment of image quality and diagnostic accuracy
    Kim, Rihyeon
    Park, Eun-Ah
    Lee, Whal
    Chung, Jin Wook
    EUROPEAN RADIOLOGY, 2016, 26 (11) : 3802 - 3810
  • [14] The usefulness of subtraction coronary computed tomography angiography for in-stent restenosis assessment of patients with CoCr stent using 320-row area detector CT
    Li, Jian
    Guo, Man-Tao
    Yang, Xiao
    Gao, Fang
    Li, Na
    Huang, Ming-Gang
    MEDICINE, 2021, 100 (51) : E28345
  • [15] Continuous imaging of esophagogastric junction in patients with reflux esophagitis using 320-row area detector CT: A feasibility study
    Fukazawa, Kousuke
    Furuta, Kenji
    Adachi, Kyoichi
    Shimura, Shino
    Kamiyama, Kazutoshi
    Aimi, Masahito
    Ohara, Shunji
    Kajitani, Takafumi
    Tsurusaki, Masakatsu
    Kitagaki, Hajime
    Kinoshita, Yoshikazu
    JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2013, 28 (10) : 1600 - 1607
  • [16] Effect of saline flush on enhancement of proximal and distal segments using 320-row coronary CT angiography
    Tomizawa, Nobuo
    Suzuki, Fumio
    Akahane, Masaaki
    Torigoe, Rumiko
    Kiryu, Shigeru
    Ohtomo, Kuni
    EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (08) : 1255 - 1259
  • [17] Coronary computed tomography angiography with 320-row detector and using the AIDR-3D: initial experience
    Neto, Roberto Sasdelli
    Nomura, Cesar Higa
    Sandoval Macedo, Ana Carolina
    Bianco, Danilo Perussi
    Kay, Fernando Uliana
    Szarf, Gilberto
    da Silva Teles, Gustavo Borges
    Shoji, Hamilton
    Santana Netto, Pedro Vieira
    Duarte Passos, Rodrigo Bastos
    Chate, Rodrigo Caruso
    Ishikawa, Walther Yoshiharu
    Bacellar Costa Lima, Joao Paulo
    Rocha, Marcelo Assis
    Marcos, Vinicius Neves
    Failla, Bruna Bonaventura
    de Gusmao Funari, Marcelo Buarque
    EINSTEIN-SAO PAULO, 2013, 11 (03): : 400 - 404
  • [18] Can sufficient preoperative information of intracranial aneurysms be obtained by using 320-row detector CT angiography alone?
    Eri Hayashida
    Akira Sasao
    Toshinori Hirai
    Kiyotoshi Hamasaki
    Toru Nishi
    Daisuke Utsunomiya
    Seitaro Oda
    Yasuhiko Iryo
    Joji Urata
    Yasuyuki Yamashita
    Japanese Journal of Radiology, 2013, 31 : 600 - 607
  • [19] Can sufficient preoperative information of intracranial aneurysms be obtained by using 320-row detector CT angiography alone?
    Hayashida, Eri
    Sasao, Akira
    Hirai, Toshinori
    Hamasaki, Kiyotoshi
    Nishi, Toru
    Utsunomiya, Daisuke
    Oda, Seitaro
    Iryo, Yasuhiko
    Urata, Joji
    Yamashita, Yasuyuki
    JAPANESE JOURNAL OF RADIOLOGY, 2013, 31 (09) : 600 - 607
  • [20] Subtraction coronary CT angiography using second-generation 320-detector row CT
    Kunihiro Yoshioka
    Ryoichi Tanaka
    Kenta Muranaka
    Tadashi Sasaki
    Takanori Ueda
    Takuya Chiba
    Kouta Takeda
    Tsuyoshi Sugawara
    The International Journal of Cardiovascular Imaging, 2015, 31 : 51 - 58