A cryosystem for cooling samples in a secondary-ion mass spectrometer with a static analyzer

被引:0
|
作者
A. N. Pustovit
A. M. Ionov
机构
[1] Russian Academy of Sciences,Institute of Microelectronics Technology and High Purity Materials
[2] Russian Academy of Sciences,Institute of Solid State Physics
关键词
Cryogenic Temperature; Cooling Sample; Cryogenic System; GaAs Sample; Transport Line;
D O I
暂无
中图分类号
学科分类号
摘要
A cryogenic system for cooling samples in an IMS-4F secondary-ion mass spectrometer with a static analyzer is described. The cryogenic system allows maintenance of the sample temperature with an accuracy of ±1 K in a temperature range of 60–420 K by combining heating and cooling (by using liquid helium or nitrogen). The effect of the sample temperature (300 and 110 K) on the secondary-ion mass spectra of Si and GaAs samples is considered.
引用
收藏
页码:91 / 94
页数:3
相关论文
共 50 条
  • [31] IMAGE POTENTIAL AND ION TRAJECTORIES IN SECONDARY-ION MASS-SPECTROMETRY
    GIBBS, RA
    HOLLAND, SP
    FOLEY, KE
    GARRISON, BJ
    WINOGRAD, N
    PHYSICAL REVIEW B, 1981, 24 (10): : 6178 - 6181
  • [32] CHARGE COMPENSATION FOR IMAGING LARGE INSULATING SAMPLES BY USING SECONDARY-ION TANDEM MASS-SPECTROMETRY
    SHORT, RT
    MCMAHON, JM
    HOLLAND, WM
    TODD, PJ
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1994, 5 (01) : 37 - 43
  • [33] IONIC MICROSCOPY (SECONDARY-ION MASS-SPECTROMETRY) - EXAMPLES OF ITS APPLICATION TO PLANT-SAMPLES
    PARIOT, C
    JAUNEAU, A
    VERDUS, MC
    POBLA, C
    LEFEBVRE, F
    DEMARTY, M
    THELLIER, M
    RIPOLL, C
    JOURNAL OF TRACE AND MICROPROBE TECHNIQUES, 1995, 13 (03): : 307 - 316
  • [34] Determination of the ion transmission coefficient for a mass spectrometer with a quadrupole ion analyzer
    Wojcik, L
    Bederski, K
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1996, 153 (2-3): : 139 - 144
  • [35] A secondary ion microprobe ion trap mass spectrometer
    Todd, PJ
    Schaaff, TG
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2002, 13 (09) : 1099 - 1107
  • [36] SPATIALLY MULTIDIMENSIONAL SECONDARY-ION MASS-SPECTROMETRY ANALYSIS
    RUDENAUER, FG
    ANALYTICA CHIMICA ACTA, 1994, 297 (1-2) : 197 - 230
  • [37] MASS SPECTROMETER ION SOURCE FOR SOLID SAMPLES
    KENDALL, BRF
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1958, 29 (12): : 1089 - 1092
  • [38] A secondary ion microprobe ion trap mass spectrometer
    Todd, Peter J.
    Schaaff, T.Gregory
    1600, Elsevier Inc. (13):
  • [39] DETERMINATION OF HYDROGEN IN SILICATES BY SECONDARY-ION MASS-SPECTROMETRY
    OTTOLINI, L
    BOTTAZZI, P
    ZANETTI, A
    VANNUCCI, R
    ANALYST, 1995, 120 (05) : 1309 - 1313
  • [40] DETERMINATION OF CALCIUM IN STEEL BY SECONDARY-ION MASS-SPECTROMETRY
    FUJIWARA, H
    MURAO, N
    ICHISE, E
    TETSU TO HAGANE-JOURNAL OF THE IRON AND STEEL INSTITUTE OF JAPAN, 1994, 80 (12): : 902 - 907