Measurement of total CO2 in microliter samples of urine and other biological fluids using infrared detection of CO2

被引:0
|
作者
Francesco Trepiccione
Francesco Maria Iena
Laura Catalini
Francesco Martino Carpi
Mogens Koed
Sebastian Frische
机构
[1] University of Aarhus,Department of Biomedicine
[2] University of Campania “Luigi Vanvitelli”,Department of Cardio
[3] University of Camerino,Thoracic and Respiratory Science
关键词
Acid/base status; Bicarbonate; NDIR; Urine; Total CO;
D O I
暂无
中图分类号
学科分类号
摘要
The purpose of this study is to describe a low-cost and simply made instrument capable of measuring the total CO2 content of microliter volumes of biological fluids utilizing a commercially available CO2 sensor based on a NDIR detector. The described instrument is based on transformation of dissolved HCO3− to CO2 by acidification and subsequent measurement of the produced CO2. The instrument has a linear response in the range 0.025–10 μmol HCO3−, which enables measurements in fresh urine and plasma samples down to 5 μl. The values from plasma were compared to measurements made on 65 μl whole blood in an automatic blood gas analyzer and found not to differ significantly. Compared to currently commercially available instruments applying the same principles to measure total CO2, this study provides a simple and robust alternative which even can be used on smaller sample volumes.
引用
收藏
页码:1267 / 1275
页数:8
相关论文
共 50 条
  • [21] DISSOLUTION OF CO2 IN GEOTHERMAL FLUIDS
    HARADEN, J
    ENERGY, 1988, 13 (09) : 689 - 690
  • [22] INTERPRETATION OF URINE CO2 TENSION
    KENNEDY, TJ
    EDEN, M
    BERLINER, RW
    FEDERATION PROCEEDINGS, 1957, 16 (01) : 72 - 72
  • [23] To CO2 or not to CO2 ,that is the question!
    Hohnadel, DC
    Gruttadauria, M
    Murray, K
    Levin, M
    D'Souza, J
    CLINICAL CHEMISTRY, 2003, 49 (06) : A90 - A90
  • [24] The biological CO2 fixation using Chlorella sp. with high capability in fixing CO2
    Murakami, M
    Yamada, F
    Nishide, T
    Muranaka, T
    Yamaguchi, N
    Takimoto, Y
    ADVANCES IN CHEMICAL CONVERSIONS FOR MITIGATING CARBON DIOXIDE, 1998, 114 : 315 - 320
  • [25] Measurement of CO2 in the Atmosphere
    Bolf, N.
    KEMIJA U INDUSTRIJI-JOURNAL OF CHEMISTS AND CHEMICAL ENGINEERS, 2019, 68 (5-6): : 266 - 266
  • [26] MEASUREMENT OF CO2 EXCHANGE OF PLANTS ON AN INFRARED GAS ANALYZER
    KORZH, BV
    SOVIET PLANT PHYSIOLOGY, 1976, 23 (02): : 355 - 361
  • [27] OSMOMETRY OF CO2 IN GAS SAMPLES
    OH, MS
    KEE, CS
    URIBARRI, J
    CARROLL, HJ
    CLINICAL CHEMISTRY, 1983, 29 (05) : 884 - 886
  • [28] O2/CO2: Biological Detection to Homeostatic Control
    Fitzgerald, Robert S.
    ARTERIAL CHEMORECEPTORS: NEW DIRECTIONS AND TRANSLATIONAL PERSPECTIVES, 2018, 1071 : 1 - 12
  • [29] Photothermal detection of CO2
    Oehler, O
    PROGRESS IN NATURAL SCIENCE, 1996, 6 : S610 - S613
  • [30] Photothermal detection of CO2
    Oehler, O
    9TH INTERNATIONAL CONFERENCE ON PHOTOACOUSTIC AND PHOTOTHERMAL PHENOMENA, CONFERENCE DIGEST, 1996, : 522 - 523