A comparative study of mid-infrared diffuse reflection (DR) and attenuated total reflection (ATR) spectroscopy for the detection of fungal infection on RWA2-corn

被引:0
|
作者
Gregor Kos
Rudolf Krska
Hans Lohninger
Peter R. Griffiths
机构
[1] Center for Analytical Chemistry,Institute for Agrobiotechnology (IFA
[2] Vienna University of Technology,Tulln)
[3] University of Idaho,Institute for Chemical Technologies and Analytics, Analytical Chemistry Division
来源
关键词
Attenuated total reflection; Diffuse reflection; Mid-infrared spectroscopy; Comparison study; Cereals; Mycotoxins;
D O I
暂无
中图分类号
学科分类号
摘要
An investigation into the rapid detection of mycotoxin-producing fungi on corn by two mid-infrared spectroscopic techniques was undertaken. Corn samples from a single genotype (RWA2, blanks, and contaminated with Fusarium graminearum) were ground, sieved and, after appropriate sample preparation, subjected to mid-infrared spectroscopy using two different accessories (diffuse reflection and attenuated total reflection). The measured spectra were evaluated with principal component analysis (PCA) and the blank and contaminated samples were classified by cluster analysis. Reference data for fungal metabolites were obtained with conventional methods. After extraction and clean-up, each sample was analyzed for the toxin deoxynivalenol (DON) by gas chromatography with electron capture detection (GC-ECD) and ergosterol (a parameter for the total fungal biomass) by high-performance liquid chromatography with diode array detection (HPLC-DAD). The concentration ranges for contaminated samples were 880–3600 μg/kg for ergosterol and 300–2600 μg/kg for DON. Classification efficiency was 100% for ATR spectra. DR spectra did not show as obvious a clustering of contaminated and blank samples. Results and trends were also observed in single spectra plots. Quantification using a PLS1 regression algorithm showed good correlation with DON reference data, but a rather high standard error of prediction (SEP) with 600 μg/kg (DR) and 490 μg/kg (ATR), respectively, for ergosterol. Comparing measurement procedures and results showed advantages for the ATR technique, mainly owing to its ease of use and the easier interpretation of results that were better with respect to classification and quantification.
引用
收藏
页码:159 / 166
页数:7
相关论文
共 50 条
  • [31] Detection of Ni2+ by a Dimethylglyoxime Probe Using Attenuated Total-Reflection Infrared Spectroscopy
    Thomas Ponnuswamy
    Oliver Chyan
    Analytical Sciences, 2002, 18 : 449 - 453
  • [32] Applying Attenuated Total Reflection-Mid-Infrared (ATR-MIR) Spectroscopy to Detect Hairtail Surimi in Mixed Surimi and Their Surimi Products
    You Zhao-hong
    Liu Zi-hao
    Gong Chao-yong
    Yang Xiao-ling
    Cheng Fang
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2015, 35 (10) : 2930 - 2939
  • [33] New tool for epidermal and cosmetic formulation studies by attenuated total-reflection spectroscopy using a flexible mid-infrared fiber probe
    H. Heise
    L. Küpper
    W. Pittermann
    L. Butvina
    Fresenius’ Journal of Analytical Chemistry, 2001, 371 : 753 - 757
  • [34] Epidermal in vivo and in vitro studies by attenuated total reflection mid-infrared spectroscopy using flexible silver halide fibre-probes
    Heise, HM
    Küpper, L
    Pittermann, W
    Stücker, M
    JOURNAL OF MOLECULAR STRUCTURE, 2003, 651 : 127 - 132
  • [35] New tool for epidermal and cosmetic formulation studies by attenuated total-reflection spectroscopy using a flexible mid-infrared fiber probe
    Heise, HM
    Küpper, L
    Pittermann, W
    Butvina, LN
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 371 (06): : 753 - 757
  • [36] Attenuated total reflection-Fourier transform infrared spectroscopy (ATR-FTIR) detection as a rapid and convenient screening test for cystinuria
    Primiano, Aniello
    Persichilli, Silvia
    Di Giacinto, Flavio
    Ciasca, Gabriele
    Baroni, Silvia
    Ferraro, Pietro Manuel
    De Spirito, Marco
    Urbani, Andrea
    Gervasoni, Jacopo
    CLINICA CHIMICA ACTA, 2021, 518 : 128 - 133
  • [37] Non-invasive evaluation of embryos using mid-infrared attenuated total reflection spectrometry of incubation medium: a preliminary study
    Aslih, Nardin
    Dekel, Ben Zion
    Malonek, Dov
    Michaeli, Medeia
    Polotov, Diana
    Shalom-Paz, Einat
    REPRODUCTIVE BIOMEDICINE ONLINE, 2023, 46 (05) : 793 - 801
  • [38] Rapid Diagnosis of Type II Diabetes Using Fourier Transform Mid-Infrared Attenuated Total Reflection Spectroscopy Combined with Support Vector Machine
    Fang, Tao
    Li, Yuanpeng
    Li, Fucui
    Huang, Furong
    ANALYTICAL LETTERS, 2018, 51 (09) : 1400 - 1416
  • [39] Detection Limits for Blood on Fabrics Using Attenuated Total Reflection Fourier Transform Infrared (ATR FT-IR) Spectroscopy and Derivative Processing
    Lu, Zhenyu
    DeJong, Stephanie A.
    Cassidy, Brianna M.
    Belliveau, Raymond G.
    Myrick, Michael L.
    Morgan, Stephen L.
    APPLIED SPECTROSCOPY, 2017, 71 (05) : 839 - 846
  • [40] Stand-off detection of solid targets with diffuse reflection spectroscopy using a high-power mid-infrared supercontinuum source
    Kumar, Malay
    Islam, Mohammed N.
    Terry, Fred L., Jr.
    Freeman, Michael J.
    Chan, Allan
    Neelakandan, Manickam
    Manzur, Tariq
    APPLIED OPTICS, 2012, 51 (15) : 2794 - 2807