A comparative study of mid-infrared diffuse reflection (DR) and attenuated total reflection (ATR) spectroscopy for the detection of fungal infection on RWA2-corn

被引:0
|
作者
Gregor Kos
Rudolf Krska
Hans Lohninger
Peter R. Griffiths
机构
[1] Center for Analytical Chemistry,Institute for Agrobiotechnology (IFA
[2] Vienna University of Technology,Tulln)
[3] University of Idaho,Institute for Chemical Technologies and Analytics, Analytical Chemistry Division
来源
关键词
Attenuated total reflection; Diffuse reflection; Mid-infrared spectroscopy; Comparison study; Cereals; Mycotoxins;
D O I
暂无
中图分类号
学科分类号
摘要
An investigation into the rapid detection of mycotoxin-producing fungi on corn by two mid-infrared spectroscopic techniques was undertaken. Corn samples from a single genotype (RWA2, blanks, and contaminated with Fusarium graminearum) were ground, sieved and, after appropriate sample preparation, subjected to mid-infrared spectroscopy using two different accessories (diffuse reflection and attenuated total reflection). The measured spectra were evaluated with principal component analysis (PCA) and the blank and contaminated samples were classified by cluster analysis. Reference data for fungal metabolites were obtained with conventional methods. After extraction and clean-up, each sample was analyzed for the toxin deoxynivalenol (DON) by gas chromatography with electron capture detection (GC-ECD) and ergosterol (a parameter for the total fungal biomass) by high-performance liquid chromatography with diode array detection (HPLC-DAD). The concentration ranges for contaminated samples were 880–3600 μg/kg for ergosterol and 300–2600 μg/kg for DON. Classification efficiency was 100% for ATR spectra. DR spectra did not show as obvious a clustering of contaminated and blank samples. Results and trends were also observed in single spectra plots. Quantification using a PLS1 regression algorithm showed good correlation with DON reference data, but a rather high standard error of prediction (SEP) with 600 μg/kg (DR) and 490 μg/kg (ATR), respectively, for ergosterol. Comparing measurement procedures and results showed advantages for the ATR technique, mainly owing to its ease of use and the easier interpretation of results that were better with respect to classification and quantification.
引用
收藏
页码:159 / 166
页数:7
相关论文
共 50 条
  • [21] Mid-infrared fiber-optic attenuated total reflection spectroscopy of the solid-liquid phase transition of water
    Millo, A
    Raichlin, Y
    Katzir, A
    APPLIED SPECTROSCOPY, 2005, 59 (04) : 460 - 466
  • [22] The influence of wetting and drying cycles on mid-infrared attenuated total reflection spectra of quartz: Understanding spectroscopy of disturbed soil
    Karlowatz, M
    Aleksandrov, A
    Orlando, T
    Cathcart, JM
    Mizaikoff, B
    DETECTION AND REMEDIATION TECHNOLOGIES FOR MINES AND MINELIKE TARGETS IX, PTS 1 AND 2, 2004, 5415 : 629 - 637
  • [23] A novel, direct, reagent-free method for the detection of beeswax adulteration by single-reflection attenuated total reflectance mid-infrared spectroscopy
    Maia, Miguel
    Barros, Ana I. R. N. A.
    Nunes, Fernando M.
    TALANTA, 2013, 107 : 74 - 80
  • [24] Comparative Study of Fourier Transform Infrared Spectroscopy in Transmission, Attenuated Total Reflection, and Total Reflection Modes for the Analysis of Plastics in the Cultural Heritage Field
    Picollo, Marcello
    Bartolozzi, Giovanni
    Cucci, Costanza
    Galeotti, Monica
    Marchiafava, Veronica
    Pizzo, Benedetto
    APPLIED SPECTROSCOPY, 2014, 68 (04) : 389 - 397
  • [25] Determination of the electric field intensities in a mid-infrared spectroelectrochemical cell using attenuated total reflection spectroscopy with the Otto optical configuration
    Brooksby, PA
    Fawcett, WR
    ANALYTICAL CHEMISTRY, 2001, 73 (06) : 1155 - 1160
  • [26] Ultrasound-Enhanced Attenuated Total Reflection Mid-infrared Spectroscopy In-Line Probe: Acquisition of Cell Spectra in a Bioreactor
    Koch, Cosima
    Brandstetter, Markus
    Wechselberger, Patrick
    Lorantfy, Bettina
    Plata, Maria Reyes
    Radel, Stefan
    Herwig, Christoph
    Lendl, Bernhard
    ANALYTICAL CHEMISTRY, 2015, 87 (04) : 2314 - 2320
  • [27] Rapid detection of gelatin in dental materials using attenuated total reflection fourier transform infrared spectroscopy (ATR-FTIR)
    Irfanita, N.
    Jaswir, I.
    Mirghani, M. E. S.
    Sukmasari, S.
    Ardini, Y. D.
    Lestari, W.
    1ST PHYSICS AND TECHNOLOGIES IN MEDICINE AND DENTISTRY SYMPOSIUM, 2017, 884
  • [28] Determination of Water-soluble Protein and Total Sugar in Royal Jelly Based on Attenuated Total Reflection Mid-infrared Spectroscopy and CARS Algorithm
    Guo C.
    Wu W.
    Chen D.
    Ye Q.
    Lu W.
    Guan R.
    Xiao C.
    Journal of Chinese Institute of Food Science and Technology, 2021, 21 (11) : 183 - 190
  • [29] Characterization of Silver Halide Fiber Optics and Hollow Silica Waveguides for Use in the Construction of a Mid-Infrared Attenuated Total Reflection Fourier Transform Infrared (ATR FT-IR) Spectroscopy Probe
    Damin, Craig A.
    Sommer, Andre J.
    APPLIED SPECTROSCOPY, 2013, 67 (11) : 1252 - 1263
  • [30] Detection of Ni2+ by a dimethylglyoxime probe using attenuated total-reflection infrared spectroscopy
    Ponnuswamy, T
    Chyan, O
    ANALYTICAL SCIENCES, 2002, 18 (04) : 449 - 453