Development of functional chest imaging with a dynamic flat-panel detector (FPD).

被引:8
|
作者
Tanaka R. [1 ]
Sanada S. [1 ]
Fujimura M. [1 ]
Yasui M. [1 ]
Nakayama K. [1 ]
Matsui T. [1 ]
Hayashi N. [1 ]
Matsui O. [1 ]
机构
[1] Department of Radiological Technology, Graduate School of Medical Science, Kanazawa University, Kodatsuno, Kanazawa
关键词
Functional imaging; Dynamic chest radiography; Flat-panel detector; Blood flow; Visualization; Image subtraction;
D O I
10.1007/s12194-008-0020-7
中图分类号
学科分类号
摘要
Dynamic FPD permits the acquisition of distortion-free radiographs with a large field of view and high image quality. In the present study, we investigated the feasibility of functional imaging for evaluating the pulmonary sequential blood distribution with an FPD, based on changes in pixel values during cardiac pumping. Dynamic chest radiographs of seven normal subjects were obtained in the expiratory phase by use of an FPD system. We measured the average pixel value in each region of interest that was located manually in the heart and lung areas. Subsequently, inter-frame differences and differences from a minimum-intensity projection image, which was created from one cardiac cycle, were calculated. These difference values were then superimposed on dynamic chest radiographs in the form of a color display, and sequential blood distribution images and a blood distribution map were created. The results were compared to typical data on normal cardiac physiology. The clinical effectiveness of our method was evaluated in a patient who had abnormal pulmonary blood flow. In normal cases, there was a strong correlation between the cardiac cycle and changes in pixel value. Sequential blood distribution images showed a normal pattern at determined by the physiology of pulmonary blood flow, with a symmetric distribution and no blood flow defects throughout the entire lung region. These findings indicated that pulmonary blood flow was reflected on dynamic chest radiographs. In an abnormal case, a defect in blood flow was shown as defective in color in a blood distribution map. The present method has the potential for evaluation of local blood flow as an optional application in general chest radiography.
引用
下载
收藏
页码:137 / 143
页数:6
相关论文
共 50 条
  • [21] Radiation dose reduction in chest radiography using a flat-panel amorphous silicon detector
    Hosch, WP
    Fink, C
    Radeleff, B
    Kampschulte, A
    Kauffmann, GW
    Hansmann, J
    CLINICAL RADIOLOGY, 2002, 57 (10) : 902 - 907
  • [22] Chest radiography with a digital flat-panel detector: Experimental receiver operating characteristic analysis
    Metz, S
    Damoser, P
    Hollweck, R
    Roggel, R
    Engelke, C
    Woertler, K
    Renger, B
    Rummeny, EJ
    Link, TM
    RADIOLOGY, 2005, 234 (03) : 776 - 784
  • [23] Indirect flat-panel detector with avalanche gain
    Zhao, W
    Hunt, DC
    Tanioka, K
    Rowlands, JA
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 150 - 161
  • [24] Development of a selenium-based flat-panel detector for radiography and fluoroscopy
    Yamada, S
    Tsukamoto, A
    Tanaka, M
    Asahina, H
    Nishiki, M
    Suzuki, K
    Uematsu, Y
    CAR '98 - COMPUTER ASSISTED RADIOLOGY AND SURGERY, 1998, 1165 : 542 - 547
  • [25] Development and characterization of a flat-panel detector-based microtomography system
    Lee, SC
    Kim, HK
    Chun, IK
    Cho, MH
    Cho, MH
    Lee, SY
    ADVANCES IN NONDESTRUCTIVE EVALUATION, PT 1-3, 2004, 270-273 : 245 - 251
  • [26] Temporal-spatial characteristic evaluation in a dynamic flat-panel detector system
    Kawashima, H.
    Tanaka, R.
    Matsubara, K.
    Ichikawa, K.
    Sakuta, K.
    Minami, S.
    Hayashi, N.
    Sanada, S.
    Kawamura, M.
    Yamamoto, T.
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [27] Sequential dual-energy subtraction imaging technique using a dynamic flat-panel detector during respiration
    Tanaka, R
    Sanada, S
    Kobayashi, T
    Suzuki, M
    Matsui, T
    Inoue, H
    Hayashi, N
    Nanbu, Y
    CARS 2004: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2004, 1268 : 1251 - 1251
  • [28] Development of patient collation system by kinetic analysis for chest dynamic radiogram with flat panel detector
    Tsuchiya, Yuichiro
    Kodera, Yoshie
    MEDICAL IMAGING 2006: PACS AND IMAGING INFORMATICS, 2006, 6145
  • [29] Chest radiography with a flat-panel detector:: Image quality with dose reduction after copper filtration
    Hamer, OW
    Sirlin, CB
    Strotzer, M
    Borisch, I
    Zorger, N
    Feuerbach, S
    Völk, M
    RADIOLOGY, 2005, 237 (02) : 691 - 700
  • [30] Routine chest radiography with a flat-panel detector: Image quality with standard dose and dose reduction
    Strotzer, M
    Voelk, M
    Fruend, R
    Hamer, O
    Zorger, N
    Feuerbach, SG
    RADIOLOGY, 2001, 221 : 354 - 354