Control of the Low-energy X-rays by Using MCNP5 and Numerical Analysis for a New Concept Intra-oral X-ray Imaging System

被引:0
|
作者
Jangyong Huh
Yunseo Ji
Rena Lee
机构
[1] Ewha Womans University,Department of Science Education
[2] University of Ulsan College of Medicine,Department of Biomedical Engineering
[3] Ewha Womans University,Department of Radiation Oncology, School of Medicine
来源
关键词
X-ray transport; MCNP5; Flattening filter;
D O I
暂无
中图分类号
学科分类号
摘要
An X-ray control algorithm to modulate the X-ray intensity distribution over the FOV (field of view) has been developed by using numerical analysis and MCNP5, a particle transport simulation code on the basis of the Monte Carlo method. X-rays, which are widely used in medical diagnostic imaging, should be controlled in order to maximize the performance of the X-ray imaging system. However, transporting X-rays, like a liquid or a gas is conveyed through a physical form such as pipes, is not possible. In the present study, an X-ray control algorithm and technique to uniformize the Xray intensity projected on the image sensor were developed using a flattening filter and a collimator in order to alleviate the anisotropy of the distribution of X-rays due to intrinsic features of the X-ray generator. The proposed method, which is combined with MCNP5 modeling and numerical analysis, aimed to optimize a flattening filter and a collimator for a uniform distribution of X-rays. Their size and shape were estimated from the method. The simulation and the experimental results both showed that the method yielded an intensity distribution over an X-ray field of 6×4 cm2 at SID (source to image-receptor distance) of 5 cm with a uniformity of more than 90% when the flattening filter and the collimator were mounted on the system. The proposed algorithm and technique are not only confined to flattening filter development but can also be applied for other X-ray related research and development efforts.
引用
收藏
页码:1078 / 1085
页数:7
相关论文
共 43 条
  • [1] Control of the Low-energy X-rays by Using MCNP5 and Numerical Analysis for a New Concept Intra-oral X-ray Imaging System
    Huh, Jangyong
    Ji, Yunseo
    Lee, Rena
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 72 (09) : 1078 - 1085
  • [2] Measurement of scattered and transmitted x-rays from intra-oral and panoramic dental x-ray equipment
    Holroyd, John
    JOURNAL OF RADIOLOGICAL PROTECTION, 2018, 38 (02) : 793 - 806
  • [3] MCNP5 evaluation of dose dissipation in tissue-like media exposed to low-energy monoenergetic X-ray microbeam
    Shaun D. Clarke
    Tatjana Jevremovic
    Radiation and Environmental Biophysics, 2005, 44 : 225 - 233
  • [4] MCNP5 evaluation of dose dissipation in tissue-like media exposed to low-energy monoenergetic X-ray microbeam
    Clarke, SD
    Jevremovic, T
    RADIATION AND ENVIRONMENTAL BIOPHYSICS, 2005, 44 (03) : 225 - 233
  • [5] Uncertainty Analysis in the Simulation of X-ray Spectra in the Diagnostic Range using the MCNP5 code
    Gallardo, S.
    Querol, A.
    Rodenas, J.
    Verdu, G.
    2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2011, : 389 - 392
  • [6] Clinical relative biological effectiveness of low-energy x-rays emitted by miniature x-ray devices
    Brenner, DJ
    Leu, CS
    Beatty, JF
    Shefer, RE
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (02): : 323 - 333
  • [7] RESPONSE DEPENDENCE OF FILM BADGES FOR X-RAYS AND GAMMA-RAYS ON X-RAY QUALITY IN THE LOW-ENERGY REGION
    SAKIHARA, K
    MATSUMOTO, T
    NAOI, J
    YAMACHI, I
    KATO, A
    FUKUDA, K
    MATSUMOTO, S
    HEALTH PHYSICS, 1985, 49 (02): : 377 - 377
  • [8] Possibilities and limitations of X-ray diffraction using high energy X-rays on a laboratory system
    te Nijenhuis, Hans
    Gateshki, Milen
    Fransen, Martijn
    ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES, 2008, 64 : C199 - C199
  • [9] Possibilities and limitations of X-ray diffraction using high-energy X-rays on a laboratory system
    Nijenhuis, J. Te
    Gateshki, M.
    Fransen, M. J.
    ZEITSCHRIFT FUR KRISTALLOGRAPHIE, 2009, : 163 - 169
  • [10] Low-energy shelf response in thin energy-dispersive X-ray detectors from Compton scattering of hard X-rays
    Michel-Hart, N.
    Elam, W. T.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 863 : 1 - 6