Control of the Low-energy X-rays by Using MCNP5 and Numerical Analysis for a New Concept Intra-oral X-ray Imaging System

被引:0
|
作者
Jangyong Huh
Yunseo Ji
Rena Lee
机构
[1] Ewha Womans University,Department of Science Education
[2] University of Ulsan College of Medicine,Department of Biomedical Engineering
[3] Ewha Womans University,Department of Radiation Oncology, School of Medicine
来源
关键词
X-ray transport; MCNP5; Flattening filter;
D O I
暂无
中图分类号
学科分类号
摘要
An X-ray control algorithm to modulate the X-ray intensity distribution over the FOV (field of view) has been developed by using numerical analysis and MCNP5, a particle transport simulation code on the basis of the Monte Carlo method. X-rays, which are widely used in medical diagnostic imaging, should be controlled in order to maximize the performance of the X-ray imaging system. However, transporting X-rays, like a liquid or a gas is conveyed through a physical form such as pipes, is not possible. In the present study, an X-ray control algorithm and technique to uniformize the Xray intensity projected on the image sensor were developed using a flattening filter and a collimator in order to alleviate the anisotropy of the distribution of X-rays due to intrinsic features of the X-ray generator. The proposed method, which is combined with MCNP5 modeling and numerical analysis, aimed to optimize a flattening filter and a collimator for a uniform distribution of X-rays. Their size and shape were estimated from the method. The simulation and the experimental results both showed that the method yielded an intensity distribution over an X-ray field of 6×4 cm2 at SID (source to image-receptor distance) of 5 cm with a uniformity of more than 90% when the flattening filter and the collimator were mounted on the system. The proposed algorithm and technique are not only confined to flattening filter development but can also be applied for other X-ray related research and development efforts.
引用
收藏
页码:1078 / 1085
页数:7
相关论文
共 43 条
  • [21] Low-dose X-ray structure analysis of cytochrome c oxidase utilizing high-energy X-rays
    Ueno, Go
    Shimada, Atsuhiro
    Yamashita, Eiki
    Hasegawa, Kazuya
    Kumasaka, Takashi
    Shinzawa-Itoh, Kyoko
    Yoshikawa, Shinya
    Tsukihara, Tomitake
    Yamamoto, Masaki
    JOURNAL OF SYNCHROTRON RADIATION, 2019, 26 : 912 - 921
  • [22] A portable organic plastic scintillator dosimetry system for low energy X-rays: A feasibility study using an intraoperative X-ray unit as the radiation source
    Williams, Kerry
    Robinson, Neil
    Trapp, Jamie
    Ackerly, Trevor
    Das, Ram
    Kemp, Penny
    Geso, Moshi
    JOURNAL OF MEDICAL PHYSICS, 2007, 32 (02) : 73 - 76
  • [23] Development of low-energy x-ray fluorescence micro-distribution analysis using a laser plasma x-ray source and multilayer optics?
    Stuik, R
    Shmaenok, LA
    Fledderus, H
    Andreev, SS
    Shamov, EA
    Zuev, SY
    Salashchenko, NN
    Bijkerk, F
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (03) : 387 - 390
  • [24] CT reconstruction by using the MLS-ART technique and the KCD imaging system - I: Low-energy X-ray studies
    Guan, HQ
    Gaber, MW
    DiBianca, FA
    Zhu, YP
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (04) : 355 - 358
  • [25] New applications of LEEIXS (low-energy electron-induced X-ray spectrometry) in surface analysis
    Romand, M
    Charbonnier, M
    Baborowski, J
    JOURNAL DE PHYSIQUE IV, 1996, 6 (C4): : 467 - 474
  • [26] Rapid, Non-destructive Inspection and Classification of Inhalation Blisters Using Low-Energy X-ray Imaging
    Rao, Nagaraja
    Ament, Brian
    Parmee, Richard
    Cameron, Jonathan
    Mayo, Martin
    JOURNAL OF PHARMACEUTICAL INNOVATION, 2018, 13 (03) : 270 - 282
  • [27] Rapid, Non-destructive Inspection and Classification of Inhalation Blisters Using Low-Energy X-ray Imaging
    Nagaraja Rao
    Brian Ament
    Richard Parmee
    Jonathan Cameron
    Martin Mayo
    Journal of Pharmaceutical Innovation, 2018, 13 : 270 - 282
  • [28] In-house-built single-crystal anomalous X-ray diffraction measurement system using white X-rays and an energy dispersive detector
    Simura, Rayko
    Yamane, Hisanori
    Sakakura, Terutoshi
    Sugiyama, Kazumasa
    JOURNAL OF THE CERAMIC SOCIETY OF JAPAN, 2021, 129 (09) : 594 - 600
  • [29] Effects of beam energy and filtration on the signal-to-noise ratio achieved with the Dexis intra-oral charge-coupled device x-ray system
    Kitagawa, H
    Farman, AG
    Kelly, MS
    Kuroyanagi, K
    JOURNAL OF DENTAL RESEARCH, 2000, 79 : 376 - 376
  • [30] Depth-Resolved Structure Analysis of Cylindrical Microdomain in Block Copolymer Thin Film by Grazing-Incidence Small-Angle X-ray Scattering Utilizing Low-Energy X-rays
    Saito, Itsuki
    Miyazaki, Tsukasa
    Yamamoto, Katsuhiro
    MACROMOLECULES, 2015, 48 (22) : 8190 - 8196