An open source automatic quality assurance (OSAQA) tool for the ACR MRI phantom

被引:0
|
作者
Jidi Sun
Michael Barnes
Jason Dowling
Fred Menk
Peter Stanwell
Peter B. Greer
机构
[1] University of Newcastle,
[2] Calvary Mater Newcastle,undefined
[3] CSIRO Australian E-Health Research Centre,undefined
关键词
Quality assurance; ACR MRI phantom; MR QA; Magnetic resonance imaging; Medical imaging;
D O I
暂无
中图分类号
学科分类号
摘要
Routine quality assurance (QA) is necessary and essential to ensure MR scanner performance. This includes geometric distortion, slice positioning and thickness accuracy, high contrast spatial resolution, intensity uniformity, ghosting artefact and low contrast object detectability. However, this manual process can be very time consuming. This paper describes the development and validation of an open source tool to automate the MR QA process, which aims to increase physicist efficiency, and improve the consistency of QA results by reducing human error. The OSAQA software was developed in Matlab and the source code is available for download from http://jidisun.wix.com/osaqa-project/. During program execution QA results are logged for immediate review and are also exported to a spreadsheet for long-term machine performance reporting. For the automatic contrast QA test, a user specific contrast evaluation was designed to improve accuracy for individuals on different display monitors. American College of Radiology QA images were acquired over a period of 2 months to compare manual QA and the results from the proposed OSAQA software. OSAQA was found to significantly reduce the QA time from approximately 45 to 2 min. Both the manual and OSAQA results were found to agree with regard to the recommended criteria and the differences were insignificant compared to the criteria. The intensity homogeneity filter is necessary to obtain an image with acceptable quality and at the same time keeps the high contrast spatial resolution within the recommended criterion. The OSAQA tool has been validated on scanners with different field strengths and manufacturers. A number of suggestions have been made to improve both the phantom design and QA protocol in the future.
引用
收藏
页码:39 / 46
页数:7
相关论文
共 50 条
  • [1] An open source automatic quality assurance (OSAQA) tool for the ACR MRI phantom
    Sun, Jidi
    Barnes, Michael
    Dowling, Jason
    Menk, Fred
    Stanwell, Peter
    Greer, Peter B.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2015, 38 (01) : 39 - 46
  • [2] AUTOMATED LOW CONTRAST DETECTION FOR ACR MRI PHANTOM QUALITY ASSURANCE
    Madamesila, Joseph
    Golestani, Ali
    RADIOTHERAPY AND ONCOLOGY, 2023, 186 : S136 - S137
  • [3] Quality assurance of clinical MRI scanners using ACR MRI phantom: Preliminary results
    Chen, CC
    Wan, YL
    Wai, YY
    Liu, HL
    JOURNAL OF DIGITAL IMAGING, 2004, 17 (04) : 279 - 284
  • [4] Quality Assurance of Clinical MRI Scanners Using ACR MRI Phantom: Preliminary Results
    Chien-Chuan Chen
    Yung-Liang Wan
    Yau-Yau Wai
    Ho-Ling Liu
    Journal of Digital Imaging, 2004, 17 : 279 - 284
  • [5] Technical Note: Validation of an automatic ACR phantom quality assurance tool for an MR-guided radiotherapy system
    Gao, Yu
    Lotey, Rajiv
    Low, Daniel A.
    Hu, Peng
    Yang, Yingli
    MEDICAL PHYSICS, 2021, 48 (04) : 1540 - 1545
  • [6] Validation of An Automatic ACR Phantom Quality Assurance Tool for the Low-Field MR Guided Radiotherapy System
    Gao, Y.
    Lotey, R.
    Hu, P.
    Yang, Y.
    MEDICAL PHYSICS, 2020, 47 (06) : E670 - E670
  • [7] Automatic Analysis of ACR Phantom Images in MRI
    Ben Alaya, Ines
    Mars, Mokhtar
    CURRENT MEDICAL IMAGING, 2020, 16 (07) : 892 - 901
  • [8] MRI quality assurance using the ACR phantom in a multi-unit imaging center
    Ihalainen, Toni M.
    Lonnroth, Nadja T.
    Peltonen, Juha I.
    Uusi-Simola, Jouni K.
    Timonen, Marjut H.
    Kuusela, Linda J.
    Savolainen, Sauli E.
    Sipila, Outi E.
    ACTA ONCOLOGICA, 2011, 50 (06) : 966 - 972
  • [9] Evaluation of the ACR MRI phantom for quality assurance tests of 1.5 T MRI scanners in Estonian hospitals
    Kaljuste, Doris
    Nigul, Mait
    PROCEEDINGS OF THE ESTONIAN ACADEMY OF SCIENCES, 2014, 63 (03) : 240 - 246
  • [10] Deep Learning Approaches for Automatic Quality Assurance of Magnetic Resonance Images Using ACR Phantom
    Torfeh, Tarraf
    Aouadi, Souha
    Yoganathan, S. A.
    Paloor, Satheesh
    Hammoud, Rabih
    Al-Hammadi, Noora
    BMC MEDICAL IMAGING, 2023, 23 (01)