A Novel Amperometric Glutamate Biosensor Based on Glutamate Oxidase Adsorbed on Silicalite

被引:0
|
作者
O. V. Soldatkina
O. O. Soldatkin
B. Ozansoy Kasap
D. Yu. Kucherenko
I. S. Kucherenko
B. Akata Kurc
S. V. Dzyadevych
机构
[1] Taras Shevchenko National University of Kyiv,Institute of High Technologies
[2] Institute of Molecular Biology and Genetics of NAS of Ukraine,Micro and Nanotechnology Department
[3] Middle East Technical University,Central Laboratory
[4] Middle East Technical University,undefined
来源
关键词
Silicalite; Enzyme; Biosensor; Glutamate oxidase; Amperometry;
D O I
暂无
中图分类号
学科分类号
摘要
In this work, we developed a new amperometric biosensor for glutamate detection using a typical method of glutamate oxidase (GlOx) immobilization via adsorption on silicalite particles. The disc platinum electrode (d = 0.4 mm) was used as the amperometric sensor. The procedure of biosensor preparation was optimized. The main parameters of modifying amperometric transducers with a silicalite layer were determined along with the procedure of GlOx adsorption on this layer. The biosensors based on GlOx adsorbed on silicalite demonstrated high sensitivity to glutamate. The linear range of detection was from 2.5 to 450 μM, and the limit of glutamate detection was 1 μM. It was shown that the proposed biosensors were characterized by good response reproducibility during hours of continuous work and operational stability for several days. The developed biosensors could be applied for determination of glutamate in real samples.
引用
收藏
相关论文
共 50 条
  • [21] A biosensor based on co-immobilized L-glutamate oxidase and L-glutamate dehydrogenase for analysis of monosodium glutamate in food
    Basu, AK
    Chattopadhyay, P
    Roychudhuri, U
    Chakraborty, R
    BIOSENSORS & BIOELECTRONICS, 2006, 21 (10): : 1968 - 1972
  • [22] GLUTAMATE OXIDASE BASED BIOSENSORS
    WOLLENBERGER, U
    SCHELLER, F
    PAWLOWA, M
    MULLER, HG
    RISINGER, L
    GORTON, L
    BIOSENSORS : APPLICATIONS IN MEDICINE, ENVIRONMENTAL PROTECTION AND PROCESS CONTROL, 1989, 13 : 33 - 36
  • [23] L-GLUTAMATE BIOSENSOR USING A NOVEL L-GLUTAMATE OXIDASE AND ITS APPLICATION TO FLOW-INJECTION ANALYSIS SYSTEM
    YE, BC
    LI, QS
    LI, YR
    LI, XB
    YU, JT
    JOURNAL OF BIOTECHNOLOGY, 1995, 42 (01) : 45 - 52
  • [24] AMPEROMETRIC L-GLUTAMATE SENSOR USING A NOVEL L-GLUTAMATE OXIDASE FROM STREPTOMYCES-PLATENSIS NTU-3304
    CHEN, CY
    SU, YC
    ANALYTICA CHIMICA ACTA, 1991, 243 (01) : 9 - 15
  • [25] DETERMINATION OF L-GLUTAMATE AND L-GLUTAMINE IN PHARMACEUTICAL FORMULATIONS BY AMPEROMETRIC L-GLUTAMATE OXIDASE BASED ENZYME SENSORS
    BOTRE, F
    BOTRE, C
    LORENTI, G
    MAZZEI, F
    PORCELLI, F
    SCIBONA, G
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1993, 11 (08) : 679 - 686
  • [26] DETERMINATION OF GLUTAMATE PYRUVATE TRANSAMINASE AND PYRUVATE WITH AN AMPEROMETRIC PYRUVATE OXIDASE SENSOR
    MIZUTANI, F
    TSUDA, K
    KARUBE, I
    SUZUKI, S
    MATSUMOTO, K
    ANALYTICA CHIMICA ACTA, 1980, 118 (01) : 65 - 71
  • [27] Amperometric glutamate biosensor based on self-assembling glutamate dehydrogenase and dendrimer-encapsulated platinum. nanoparticles onto carbon nanotubes
    Tang, Longhua
    Zhu, Yihua
    Xu, Lihuan
    Yang, Xiaoling
    Li, Chunzhong
    TALANTA, 2007, 73 (03) : 438 - 443
  • [28] Monitoring of the velocity of high-affinity glutamate uptake by isolated brain nerve terminals using amperometric glutamate biosensor
    Soldatkin, O.
    Nazarova, A.
    Krisanova, N.
    Borysov, A.
    Kucherenko, D.
    Kucherenko, I.
    Pozdnyakova, N.
    Soldatkin, A.
    Borisova, T.
    TALANTA, 2015, 135 : 67 - 74
  • [29] Characterization of a thermophilic L-glutamate dehydrogenase biosensor for amperometric determination of L-glutamate by flow injection analysis
    Pasco, N.
    Jeffries, C.
    Davies, Q.
    Downard, A.J.
    Roddick-Lanzilotta, A.D.
    Gorton, L.
    Biosensors and Bioelectronics, 1999, 14 (02): : 171 - 178
  • [30] Characterisation of a thermophilic L-glutamate dehydrogenase biosensor for amperometric determination of L-glutamate by flow injection analysis
    Pasco, N
    Jeffries, C
    Davies, Q
    Downard, AJ
    Roddick-Lanzilotta, AD
    Gorton, L