Harmonisation of nanoparticle concentration measurements using GRIMM and TSI scanning mobility particle sizers

被引:0
|
作者
Manish Joshi
B. K. Sapra
Arshad Khan
S. N. Tripathi
P. M. Shamjad
Tarun Gupta
Y. S. Mayya
机构
[1] Bhabha Atomic Research Centre,Radiological Physics and Advisory Division
[2] Indian Institute of Technology,Department of Civil Engineering
来源
关键词
Nanoparticles; Harmonisation; SMPS; TSI; GRIMM; Sheath air dryer;
D O I
暂无
中图分类号
学科分类号
摘要
Regional studies focusing on the role of atmospheric nanoparticles in climate change have gained impetus in the last decade. Several multi-institutional studies involving measurement of nanoparticles with several kinds of instruments are on the rise. It is important to harmonize these measurements as the instruments may work on different techniques or principles and are developed by different manufacturers. Scanning mobility particle sizers (SMPS) are often used to measure size distribution of nanoparticles in the airborne phase. Two such commercially available instruments namely, GRIMM and TSI-SMPS have been compared for ambient and laboratory generated conditions. A stand-alone condensation particle counter (CPC) of TSI make was used as a reference for particle concentration measurements. The consistency of the results in terms of mean size and geometric standard deviation was seen to be excellent for both the SMPSs, with GRIMM always showing slightly (approximately 10 %) lower mean size. The integrated number concentration from GRIMM-SMPS was seen to be closer to stand-alone reference CPC compared to TSI-SMPS, for an ambient overnight comparison. However, a concentration-dependent response, i.e. the variations between the two instruments increasing with the concentration, was observed and possible reasons for this have been suggested. A separate experiment was performed for studying the modifying effect of diffusion dryer and sheath air dryer on the measured aerosol size spectra. A significant hygroscopic growth was noted when diffusion dryer was attached to one of the SMPS. The introduction of sheath air dryer in GRIMM-SMPS produced a significant shift towards lower mean size. These results have been compared and discussed with the recent inter-comparison results to strengthen and harmonize the measurement protocols.
引用
收藏
相关论文
共 50 条
  • [41] Quantitative microscopy-based measurements of circulating nanoparticle concentration using microliter blood volumes
    Tietjen, Gregory T.
    DiRito, Jenna
    Pober, Jordan S.
    Saltzman, W. Mark
    [J]. NANOMEDICINE-NANOTECHNOLOGY BIOLOGY AND MEDICINE, 2017, 13 (06) : 1863 - 1867
  • [42] Comparison of Particle Size Measurements with Laser-Induced Incandescence, Mass Spectroscopy, and Scanning Mobility Particle Sizing in a Laminar Premixed Ethylene/Air Flame
    Stirn, R.
    Baquet, T. Gonzalez
    Kanjarkar, S.
    Meier, W.
    Geigle, K. P.
    Grotheer, H. H.
    Wahl, C.
    Aigner, M.
    [J]. COMBUSTION SCIENCE AND TECHNOLOGY, 2009, 181 (02) : 329 - 349
  • [43] A concept of an automated function control for ambient aerosol measurements using mobility particle size spectrometers
    Schladitz, A.
    Merkel, M.
    Bastian, S.
    Birmili, W.
    Weinhold, K.
    Loeschau, G.
    Wiedensohler, A.
    [J]. ATMOSPHERIC MEASUREMENT TECHNIQUES, 2014, 7 (04) : 1065 - 1073
  • [45] Evaluation of chemical mechanical planarization slurry dispersion using a combined scanning mobility particle sizer-optical particle sizer system
    Kwak, Donggeon
    Kim, Juhwan
    Kim, Taesung
    [J]. AEROSOL SCIENCE AND TECHNOLOGY, 2023, 57 (09) : 833 - 841
  • [46] Nanoparticle measurements using single particle ICP-MS and capillary electrophoresis ICP-MS
    Olesik, John
    Jiao, Shi
    [J]. ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2017, 254
  • [47] Measurements of particle velocity and solid concentration near the wall of a cyclone using high-speed particle shadow velocimetry
    Cao, Ming-Qian
    Chen, Jian-Yi
    Hu, Bo
    Fan, Xiao-Qi
    Cui, Hao
    Wei, Yaodong
    [J]. POWDER TECHNOLOGY, 2023, 426
  • [48] Evaluation of Three Sample Introduction Systems for Impurity Analysis of an Ultrapure Reagent Using a Scanning Mobility Particle Sizer
    Chang, Alice Chinghsuan
    Hung, Shu-Han
    Pan, Yi-Hsuan
    Liu, Yi-Hung
    Fu, Wei-En
    Lin, Yu-Ju
    Lin, Fang-hsin
    [J]. ACS OMEGA, 2022, 7 (16): : 13622 - 13628
  • [49] Fast and precise measurement in the sub-20 nm size range using a Scanning Mobility Particle Sizer
    Troestl, Jasmin
    Tritscher, Torsten
    Bischof, Oliver F.
    Horn, Hans-Georg
    Krinke, Thomas
    Baltensperger, Urs
    Gysel, Martin
    [J]. JOURNAL OF AEROSOL SCIENCE, 2015, 87 : 75 - 87
  • [50] Real-time measurements of suspended sediment concentration and particle size using five techniques
    Felix, D.
    Albayrak, I.
    Abgottspon, A.
    Boes, R. M.
    [J]. 28TH IAHR SYMPOSIUM ON HYDRAULIC MACHINERY AND SYSTEMS (IAHR2016), PTS 1-12, 2016, 49