A Method for Assessing the Effect of Metal Artifact Reduction Algorithms on Quantitative Characteristics of CT Images

被引:0
|
作者
Yu. A. Vasilev
D. S. Semenov
E. S. Akhmad
O. Yu. Panina
K. A. Sergunova
A. V. Petraikin
机构
[1] Research and Practical Clinical Center for Diagnostics and Telemedicine Technologies of the Moscow Healthcare Department,
来源
Biomedical Engineering | 2020年 / 54卷
关键词
D O I
暂无
中图分类号
学科分类号
摘要
The effect of metal artifact reduction algorithms on CT image quality was considered. The mean signal intensities and noise levels were compared for objects with different X-ray densities and spatial positions using phantom modeling. It was shown that, despite an improvement in the visual characteristics of the images, their quantitative characteristics may deteriorate: the change in the mean density reached in some cases 15%; the increase in noise, 30%.
引用
收藏
页码:285 / 288
页数:3
相关论文
共 50 条
  • [41] DICDNet: Deep Interpretable Convolutional Dictionary Network for Metal Artifact Reduction in CT Images
    Wang, Hong
    Li, Yuexiang
    He, Nanjun
    Ma, Kai
    Meng, Deyu
    Zheng, Yefeng
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2022, 41 (04) : 869 - 880
  • [42] An Image Inpainting Based Surrogate Data Strategy for Metal Artifact Reduction in CT Images
    Oehler, M.
    Buzug, T. M.
    4TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, 2009, 22 (1-3): : 651 - 654
  • [43] Evaluation of the Impact of Using TOF Technique on Metal Artifact Reduction in PET/CT Images
    Sharifpour, R.
    Ghafarian, P.
    Ay, M. R.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 : S363 - S364
  • [44] Quantitative evaluation of metal artifact reduction techniques
    Kolind, SH
    MacKay, AL
    Munk, PL
    Xiang, QS
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 20 (03) : 487 - 495
  • [45] Effect of metal CT artifact with model-based dose calculation algorithms
    Reniers, B.
    Verhaegen, F.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S147 - S148
  • [46] Prior-based Metal Artifact Reduction in CT using Statistical Metal Segmentation on Projection Images
    Hegazy, M. A.
    Cho, M. H.
    Lee, S. Y.
    Kim, Kisoo
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
  • [47] Dual-domain metal trace inpainting network for metal artifact reduction in baggage CT images
    Hai, Chao
    He, Jingze
    Li, Baolei
    He, Penghui
    Sun, Liang
    Wu, Yapeng
    Yang, Min
    MEASUREMENT, 2023, 207
  • [48] Iterative deblurring for CT metal artifact reduction
    Wang, G
    Snyder, DL
    OSullivan, JA
    Vannier, MW
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 1996, 15 (05) : 657 - 664
  • [49] Superiorized method for metal artifact reduction
    Humphries, Thomas
    Wang, Boyang
    MEDICAL PHYSICS, 2020, 47 (09) : 3984 - 3995
  • [50] CT metal artifact reduction method correcting for beam hardening and missing projections
    Verburg, Joost M.
    Seco, Joao
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (09):